1. Introduction
The number of prosthetic heart valve (PHV) implantation procedures continues to increase worldwide. 1 Traditionally, surgeons replaced valves surgically, but in the last 15 years transcatheter implantation techniques have become firmly established in routine clinical care. 1–3 PHV dysfunction includes obstruction, (para)valvular leakage, structural valve deterioration and endocarditis. 4 Echocardiography remains an excellent first-line imaging tool for PHV assessment, providing both anatomical and functional information. 4 However, cardiac computed tomography (CT) may provide complementary information to assess PHV functioning and, in case of dysfunction, helps determine its cause and plan a potential re-operation or intervention. 4,5
This multi-society consensus document is written by a team of experts and provides a brief overview of heart valve replacement techniques, and available imaging techniques for PHV assessment. It then discusses in depth the role of cardiac CT, including scan acquisition, image postprocessing, imaging findings, interpretation, and reporting in light of the various types of PHV dysfunction. Given the relatively limited amount of available data and literature on the application of CT to evaluate PHV function, the writing group has included consensus statements. These are based not on the available published literature complemented by expert consensus but do not represent a formal guideline with levels of evidence. In addition, we provide an overview table summarizing the different techniques for PHV assessment (Table 1).
This copy is for personal use only. To order printed copies, contact reprints@rsna.org
Abbreviations
18F-FDG = 18F-fluorodeoxyglucose, AVR = aortic valve replacement, BVD = bioprosthetic valve degeneration, $\mathrm{CT}=$ computed tomography, ECG $=$ electrocardiography, HALT $=$ hypoattenuated leaflet thickening, MVR $=$ mitral valve replacement, PET $=$ positron emission tomography, $\mathrm{PHV}=$ prosthetic heart valve, PPM = patient-prosthesis mismatch, PVE = prosthetic valve endocarditis, PVL $=$ paravalvular leakage, SAVR $=$ surgical aortic valve replacement, SMVR = surgical mitral valve replacement, SVD = structural valve dysfunction, TAVI = transcatheter aortic valve implantation, TAVR $=$ transcatheter aortic valve replacement, THV = transcatheter heart valve, TMVR $=$ transcatheter mitral valve replacement, TPV $=$ transcatheter pulmonary valve, TTE $=$ transthoracic echocardiography, $\mathrm{ViV}=$ valve-in-valve
Keywords
Computed Tomography, Valve, Cardiac, Imaging, Dysfunction, Prosthetic
2. Heart valve replacement technique
PHVs include mechanical and biologic valves placed via surgical or transcatheter (biological valves only) techniques into an aortic, mitral, pulmonary, or tricuspid valve position. Common nomenclature includes surgical aortic valve replacement (SAVR), transcatheter aortic valve replacement (TAVR), transcatheter aortic valve implantation (TAVI), surgical mitral valve replacement (SMVR), transcatheter mitral valve replacement (TMVR), and transcatheter pulmonary valve (TPV).
Contemporary mechanical valves typically feature bileaflet designs (two semilunar shaped discs in a ring); however, patients may still present with complications of older type PHVs such as tilting disc or ball and cage valves. 6 Biological valves include stented valves (bovine pericardium or porcine valve tissue), and less commonly, stentless bioprosthetic valves (i.e., Medtronic Freestyle), homografts from human cadaveric tissue, or autografts
||Table 1: Overview of the different imaging modalities for prosthetic heart valve assessment.|||| |---|---|---|---|---| ||Echocardiography|CT|Cine fHuoroscopy| Nuclea imaging| | Normal functioning PHV ||||| |Leaflet motion|++||+++ (mechanical PHV only) n/a|| ||+++| +++ (mechanical PHV only) ++ (biological PHV) |n/a|| | Flow and pressure gradients ||n/a||n/a| | Determination PHV type and size | + (mechanical vs. biological) |+++||n/a| | PHV dysfunction||||| || +(to detect ↑ pressure +++ gradient) |||n/a| | Pannus||| ++ (indirect by detecting leaflet motion restriction in mechanical PHV) || |Thrombus|++|+++| ++ (indirect by detecting leafet motion restriction in mechanical PHV) |n/a| |||||| |Paravalvular leak|+++|+ (location and size)|+ (contrast needed)|n/a| | Patient prosthesis mismatch |++||n/a|n/a| |Endocarditis|Endocarditis|Endocarditis|Endocarditis|| |Endocarditis in|+++|+++||++ (18F-FDG PET-CT)| |general||||| |Vegetations|+++||n/a|| |Valve dehiscence|+++|++|+++|n/a| | Abscesses and my- cotic aneurysms ||+++|n/a|n/a| |||||| | Biological PHV dys-||||| |function||||| |HALT|++|+++|n/a|n/a| |RELM|+|++|n/a|n/a| | Structural valve deterioration/BVD |++|+++|n/a| +++ (18F-Fluoride PET)* |
Comparison of the information provided by different imaging modalities regarding PHV (dys)function. $\mathrm{{N}/a=}$ not applicable; HALT $=$ hypoattenuated leaflet thickening; RELM $=$ reduced leaflet motion; BVD $=$ bioprosthetic valve dysfunction; 18F-FDG PET-CT = 18F-fluorodeoxyglucose positron emission tomography-computed tomography; 18F-fluoride PET $=18\mathrm{F}$ -fluoride positron emission tomography; \* mainly in research setting. Range: – to $+++$ .
with a patient’s pulmonary valve (i.e., Ross procedure). 6 The stented valves are the most commonly used PHVs. Transcatheter valves consist of biological tissue mounted on an expandable frame. Fig. 1 provides examples of the various valve types.
The choice of valve replacement type and technique depends on the patient’s demographics, co-morbidities, clinical presentation, and the surgeon’s and institutional practice pattern. Mechanical valves require lifetime anticoagulation to prevent valve thrombosis but can conversely demonstrate lifetime durability. Biologic valves can be managed with aspirin only but are limited in durability.
During surgical valve replacement the native valve is excised, valve sutures are placed around the annulus, and the PHV is seated and secured with sutures in the orthotopic position. Pledgets (small synthetic patches to reduce the risk of sutures tearing through tissue) may be placed in the supra- or sub-annular position, depending on the surgeon’s preference. Sutures are secured with surgeon hand-knots or an automated suture fastening device (i.e., Cor-Knot), which will appear as metal targets adjacent to the cuff of the prosthetic valve. In the last decade, stented sutureless aortic bioprostheses that are implanted have also been introduced clinically. 7 After opening of the aorta and excising the leaflets the collapsed prosthetic valve is re-expanded under direct vision.
Transcatheter valves are mounted in a collapsed state on the delivery system, passed through the blood vessels, and then re-expanded in the orthotopic valve position. Importantly, the native valve remains in place, and the leaflets are left in the open position adjacent to the transcatheter valve. In patients with an existing valve prosthesis, a transcatheter valve can be placed into the existing frame to overcome the structural deterioration of the bioprosthetic valve, called valve-in-valve (ViV).
3. General principles of PHV imaging assessment The first step of PHV evaluation involves reviewing the operative report which contains information on the prosthesis brand, model type, device number labelling, implantation technique and date of implant. If the operative report is unavailable, familiarity with the CT or fluoroscopic appearance of PHVs enables a streamlined approach to confirm the specific device. Knowledge of the used PHV allows device-specific assessments, while minimizing potential confusion regarding its form and function. 4,8 Second, reviewing previous imaging studies, when available, adds to the narrative of the current imaging findings. Serial imaging allows for longitudinal comparison reads for early detection and serial evaluation of signs of valve degeneration. Lastly, all available imaging data from different modalities should be included in the interpretation and synthesis of available imaging sets, as different imaging modalities provide valuable information on different aspects of PHV (dys)functioning. 4
4. Transthoracic and transesophageal echocardiography for PHV assessment
Transthoracic echocardiography (TTE) assesses the hemodynamic and anatomical function of PHVs. 4 It is non-invasive, free of radiation and iodinated contrast agents and easily accessible, while providing both anatomical and functional information. Transesophageal echocardiography (TEE) is the preferred imaging modality to identify the mechanism of PHV dysfunction or failure. TEE offers higher spatial resolution than TTE and views unobstructed by ribs and lungs but is semi-invasive and requires sedation (which may affect general haemodynamics and hence valvular flow) prior to performing the examination. Both TTE and TEE have a higher temporal resolution than CT.
 Figure 1: Photographs of different prosthetic heart valve types. (A) Mechanical bileaflet valve prosthesis. Two semilunar disks are suspended in the valve ring at the hinge points. Most contemporary mechanical valves have this design. (B) Tilting disk mechanical prosthesis. The single disk is held in place by fixed struts that allow the disk to move between the open and closed position. (C) Mechanical ball in cage valve. This valve type is no longer implanted but may be encountered in specific patients. (D, E) Biological stented frame valves which consist of a frame in which the valve leaflets of either porcine valve tissue $\{\pmb{\mathfrak{D}}\}$ or bovine pericardium ( E ) are suspended. All valve types in A – E have a fabric suture ring (white ring in the images) through which the sutures are placed to insert the valve. (F) Biological non stented valve and aorta prosthesis. The valve and aorta are explanted porcine tissue. (G, H) Transcatheter biological prostheses that consist of a metal stent in which leaflets made of bovine ( G ) or porcine ( H ) tissue are suspended. Reproduced with permission from Pibarot et al. 6
European and American guidelines recommend a baseline echocardiogram to establish baseline measurements for future comparison within 30 days or 6 to 12 weeks after a PHV is placed. 9,10 Routine echocardiographic evaluation can then be performed at 1-year and annually thereafter, or at 5-year, 10-year, and then annual intervals to evaluate for valvular degeneration depending on the PHV type that was placed. 9,10 In patients with PHV, echocardiography is the first line of investigation for new onset symptoms of heart failure. 4,9,10 Imaging for PHVs can be divided into two primary categories of study: detection of valve degeneration or dysfunction and subsequent identification of their underlying mechanisms.
Standard of care echocardiographic assessment for PHV follows an established search profile pattern for hemodynamic evaluation. Established calculated echocardiographic parameters suggesting PHV deterioration have been defined by PHV-specific velocities and gradients and Doppler velocity indices. 4 Echocardiographic evaluation of PHV should be a comprehensive, integrative approach, utilizing multiple echo views and angulations, including off-axis angles to acquire 2-dimensional, Doppler (color, pulse, and continuous wave), M-mode, and 3-dimensional images with and without color. 4 Using these tools, echo can often visualize the PHV leaflet motion, presence or absence of bioprosthetic leaflet thickening, calcification, or masses (thrombus and vegetation), stability of the PHV sewing ring, and flow through and around the valve. 4,11 Color Doppler interrogation of the PHV is engaged for assessment of presence or absence of PHV regurgitation. Color Doppler imaging helps identify the presence, location (central vs. paravalvular), and severity of regurgitant jets. A complete echo assessment of PHV must also include chamber size and function, wall thickness, intracardiac pressure estimation, and assessment of adjacent valve function. 4,11,12 Echocardiographic template reporting should include the PHV type, size, date of implant, indication for the examination, patient symptoms, height, weight, vitals (heart rate and blood pressure), and results of any prior studies.
The presence of sewing ring and stent frame shadowing artifacts limits adequate visualization of surgical and transcatheter PHV by TTE. 13 Acoustic shadowing from the PHV may impair adequate Doppler assessment. TTE Doppler flow malalignment may lead to under-estimation of PHV velocity/gradients, and the presence of eccentric jets of regurgitation may not be appreciated by TTE.
TEE may overcome some of the limitations of TTE and can be a useful adjunct for PHV assessment, particularly with 3D and multiplanar reconstruction. TEE allows for optimal visualization
Key Points
■ After PHV implantation, a baseline echocardiography is recommended by both European and American guidelines, to be followed by repeated echocardiography examinations at a variable time interval depending on the used device. ■ Transthoracic echocardiography is the first-line imaging test in case of suspected PHV dysfunction and provides both anatomical and functional information. ■ Transesophageal echocardiography may overcome some of the limitations of TTE and can be a useful adjunct for PHV assessment but is semi-invasive.
of posterior structures, and transgastric views of the PHV enable thorough evaluation of PHVs for paravalvular leaks, valve dehiscence, PHV leaflet motion, and feasibility of next-generation repair/replacement strategies. The American Society of Echocardiography has described detailed information on echocardiographic assessment of PHV. 4
5. X-ray cine fluoroscopy for leaflet motion assessment of mechanical valves
Cine fluoroscopy is an easy, accessible, and non-invasive examination that provides high temporal and spatial resolution. Non-contrast exams can be used to determine the general type of prosthesis (based on appearance), relative stability of the device, and mechanical PHV leaflet mobility. Using iodinated contrast adds qualitative assessment of regurgitation of the valve, though it does not always identify the exact mechanism of regurgitation (central vs paravalvular). Furthermore, intra-arterial contrast delivery is required and hence it is only used in very selected cases, with TTE and TEE being available as alternatives.
Cine fluoroscopy can evaluate adequate leaflet opening and closing of mechanical PHV leaflets, often complementing echocardiography in the setting of elevated gradients or significant regurgitation. 4,9,14 Cine fluoroscopy examinations are performed with the patient in a supine position, with the operator adjusting the C-arm to align the valve in a side view “in profile” as well as “en face,” if feasible. Due to differing anatomy and post-surgical orientation, the cranial/caudal and right/left angulations are often different for each patient, and care must be taken to fully align the annular frame and leaflets each in a single side view plane to accurately measure the opening and closing angles. 14 These angles are then compared to expected normal values as available from the manufacturer’s documentation. Limitations of cine fluoroscopy include lack of any hemodynamic or advanced anatomical assessment to determine the severity of PHV dysfunction or mechanism of failure. Thus, cine fluoroscopy is best when paired with clinical and hemodynamic data. 4,14
Key Points
■ Cine fluoroscopy can be used to assess mechanical PHV leaflet motion and determine opening and closing angles. ■ Depending on the PHV orientation it may be difficult to obtain a view precisely perpendicular to the valve leaflets.
6. Positron emission tomography CT
Molecular cardiovascular imaging provides complementary information on the metabolic activity of pathological processes in PHV tissue and surrounding structures, which is particularly useful in challenging cases with a diagnostic dilemma. In principle, molecular imaging is performed using a systemic injection of a ligand-specific radiotracer which colocalizes with a target of interest (GLUT1,3 receptor, nanocrystalline hydroxyapatite, glycoprotein IIb/IIIa receptor) and provides a metric of emitted photon energy detection using either a positron emission tomography (PET) or single-photon emission computerized tomography scanner. Recent advances in whole-body PET scanners enable rapid, low radiation exposure acquisitions
which have a spatial resolution of less than $3\,\mathrm{\mm}$ . Infection and inflammation imaging are the main clinical applications for molecular cardiovascular imaging. 18F-fluorodeoxyglucose (18F-FDG) 15,16 and 99mTc-Hexa-methyl-propyl-amineoxime radiolabelled white blood cells 17 have high accuracy in detecting prosthetic valve endocarditis (PVE), as well as highlighting remote sites of systemic infection by directly visualizing activated macrophages in situ. Discrete focal 18F-FDG uptake which colocalizes with low CT attenuation tissue provides supportive evidence to confirm a diagnosis of prosthetic valve endocarditis. 15,16 However, one should be aware of false positive interpretation of normal perivalvular 18F-FDG uptake and other confounding factors such as the use of Bioglue. 15,18 More recently, an expanding number of disease specific radiotracers have facilitated the assessment of non-inflammatory pathology in bioprosthetic valve dysfunction. Calcification is a key feature of deteriorating bioprosthetic valve performance and highly sensitive radiotracers such as 18F-fluoride can localize regions of hydroxyapatite deposition within dysfunctional bioprosthetic valve leaflets. 19 Comparable 18F-fluoride signal intensity is seen within dysfunctional transcatheter aortic valve leaflets, thereby enabling clinicians to assess bioprosthetic leaflet integ
Key Points
■ 18F-fluorodeoxyglucose PET-CT and radiolabeled white blood cells single-photon emission CT have a high accuracy in detecting prosthetic valve endocarditis (PVE), as well as highlighting remote sites of systemic infection. PET imaging has an important role in patients with an unclear diagnosis. ■ New radiotracers such as 18F-fluoride are being investigated to demonstrate valve degeneration and the presence of thrombus formation in hypoattenuated leaflet thickening.
rity in challenging cases. 20 Additionally, the novel glycoprotein IIb/IIIa receptor radiotracer 18F-GP1 can be used to confirm the presence of thrombus formation in hypoattenuated leaflet thickening. 21 By combining biological activity data from PET with the high-resolution structural information from cardiac CT images, clinicians have a hybrid modality to monitor disease progression and stratify therapy more precisely.
7. CT acquisition protocols for PHV imaging
The role of CT in evaluation of prosthetic heart valves is rapidly expanding. CT complements echocardiography imaging, further aiding physicians to gain a comprehensive assessment of the functional and anatomic performance of the PHV. To image suspected PHV dysfunction, physicians must choose an optimal CT image acquisition protocol, which necessitates a multifaceted approach. Such a protocol should include image data acquisition in various phases of the cardiac cycle, to facilitate the assessment of PHV dynamics. Furthermore, the protocol should enable precise differentiation between the prosthetic materials and surrounding native tissue changes and be adapted to the anatomical position of the investigated PHV (left vs. right heart). Finally, the use of a dedicated workstation enables optimal extraction of information from the acquired data set through image analysis and postprocessing.
As a general rule, tube potential and current settings used for coronary CTA provide good image quality for PHV imaging as well. Automatic tube potential and current adaptation based on the patient’s size and anatomical region scanned usually provides adequate adjustments. For mechanical and transcatheter PHV it may be beneficial to use higher kV settings (120, 140 kV) to reduce PHV artifacts. Tube current modulation during the R–R interval should be used cautiously (see below).
 Figure 2: Normal imaging appearance on CT images of different types of prosthetic heart valves with 3D volume rendering and valve in closed and open position. Top row: mechanical bileaflet valve, lower row: stented biological valve.
7.1. Functional evaluation
In general, CT imaging of the heart and PHV should be performed using ECG-synchronized thin slice $(<\!1\mathrm{mm})$ acquisition and reconstructions to optimally visualize the PHV and allow for multiplanar reconstructions 22,23 (Fig. 2). CT scanners with at least 64 slices or equivalent are recommended although retrospective ECG-gated acquisitions may also be performed on scanners with fewer slices (e.g. 16 or 40), albeit at a longer acquisition time. Data acquisition allowing reconstructions of image datasets throughout the cardiac cycle is required for functional evaluation of leaflet motion (also called 4D-CT imaging). Typically, images are reconstructed at $5{-}10\%$ intervals across the cardiac cycle. This approach allows a dynamic assessment of both normal and abnormal leaflet motion with results comparable with conventional cinefluoroscopy for mechanical PHV and valve stability/dehiscence (i.e. rocking motion). 24,25 CT evaluation can also replace fluoroscopic assessment when it is limited by impractical C-arm angulation. 25,26 For most centers a retrospective electrocardiography (ECG)-gated image acquisition protocols is the preferred acquisition mode. Alternatively, perform a prospective ECG-gated sequential acquisition preferably with a wide padding window to have the required phases throughout the cardiac cycle for image analysis. 27 While the radiation dose of retrospective ECG-gating is higher compared with other data acquisition techniques, it offers in general the highest flexibility to correct triggering errors due to arrhythmias. 26,28
Keeping radiation exposure in mind, design the arterial phase acquisition with caution. Since image quality in this phase should ideally be consistent across the cardiac cycle, one can use dose modulation techniques targeted to optimize one cardiac phase over another (as often used in conventional CT-imaging of the coronary arteries) but do not lower tube current too aggressively. Cardiac arrhythmia such as atrial fibrillation or extrasystole are relatively common in patients with valvular heart disease. In such patients ECG editing 29 or absolute delay multiphase reconstructions 30 may be beneficial to improve image quality. Note that streaking and beam hardening artifacts from mechanical leaflets can limit interpretation, and it is important to identify the phases of the cardiac cycle with the least artifact. 31
Previous studies describe the use of Beta-blocker medication for heart rate control to improve image quality which in general improves image quality. 32,33 Nitroglycerine does not impact PHV image quality but may be beneficial if the coronaries are to be assessed as well. However, both beta-blockers and nitroglycerine may be contraindicated in patients with e.g. aortic stenosis and one should use them with care depending on the clinical status of the patient and in consultation with the referring cardiologist.
7.2. PHV material imaging and tissue characterization
PHVs can cause artifacts that may hamper image quality and lower the diagnostic yield. 34,35 Several techniques can reduce PHV material-induced artifacts, including higher (e.g. $140\,\mathrm{kVp})$ ) tube voltage, 34 reconstruction of high keV level monoenergetic images from dual-energy CT, 36 and acquisitions that use prospective ECG-triggering. 37 Iterative reconstruction techniques allow for considerable dose reduction 38,39 and even improved overall objective image quality. 39 More recently, photon-counting detector CT with its high spatial resolution in combination with sharp kernels and dual energy-based reconstruction with high keV monoenergetic images has demonstrated further improvements in the quality of prosthesis depiction with fewer artifacts. 40 Iterative metal artifact reduction algorithms may reduce beam hardening artifacts but their application for PHV imaging is limited and results may be variable and may even induce secondary artifacts. 34
Despite these approaches, excessive streaking may make it difficult to evaluate monoleaflet valves, especially the Bjork-Shiley and Sorin monoleaflet valves. 26,32 Fortunately, most contemporary bileaflet mechanical valve leaflets are associated with only limited artifacts on CT imaging. 26,35
7.3. Contrast injection
In general, a multiphase contrast injection protocol is recommended. A non-contrast enhanced scan demonstrates presence and extent of calcifications, which may help to differentiate between acute and chronic infection and show structural biological PHV dysfunction. Additionally, non-contrast images can distinguish suture pledgets and paravalvular leakage, which
 Figure 3: Suture pledgets. CT angiography (A) and non contrast enhanced (B) images of a biological prosthetic heart valve. Notice the small hyperdense structures around the valve ring which represent pledgets.
have similar attenuation characteristics when only contrast-enhanced images are available 41,42 Fig. 3. Perform this non-contrast enhanced acquisition in a single phase of the cardiac cycle. A low dose multiphase acquisition can be used to evaluate mechanical PHV leaflet motion in case of contra-indications for contrast administration. 28 With spectral or photon counting CT scanners, a virtual non-contrast reconstruction from the CTA images may substitute an additional true non-contrast scan. 43,44 Acquire an arterial phase acquisition to evaluate valve position and leaflet dynamics as well as to search for leakage, obstruction, pannus, hypoattenuated leaflet thickening (HALT), and to search for the various features of infective endocarditis. 27 If required one can also evaluate the coronary arteries on these images, with prior sublingual administration of nitroglycerine
recommended in this scenario if the patient’s hemodynamic condition allows for this. 45 Optionally, add a delayed phase acquisition to the protocol (with image data acquisition starting 60–70 seconds after contrast injection), which can be useful for improved visualization of abscesses in infective endocarditis and to differentiate between delayed flow and thrombus in the cardiac chambers. 27,46,47 In contrast to the arterial phase acquisition, both the non-enhanced and delayed phase acquisition usually can be acquired with a narrow ECG-pulsing window, to reduce radiation exposure.
In patients with a need to keep the volume of iodinated contract material a low as possible use either low kV settings to increase contrast attenuation (at the expense of more PHV induced artifacts in some mechanical and transcatheter PHV)
 Figure 4: How to obtain image reconstructions in plane and perpendicular to the valve. Scan of a patient with a mechanical aortic valve prosthesis. The images are loaded in a multiplanar viewer with the lines of the “crosshairs” locked (i.e. they remain in a fixed 90-degree orientation when rotating one of the axes). Panels A , B and C are the initial axial, coronal and sagittal orientations. In the coronal plane (B) the red line is rotated counterclockwise to be parallel to the valve ring. In the sagittal plane (C) the red line is then rotated to also be parallel to the valve ring. In the resulting planes (D,E and F) it is checked that the valve ring can be seen as a complete ring in the adapted axial plane (D) . If the ring is not seen in one image the red axis needs to be rotated some more. Then in the adapted axial plane (D) one needs to scroll up and down until the two valve leaflets can be seen as two parallel lines inside the valve ring. If the leaflets are not parallel the red line needs to be rotated again whilst looking at the leaflets as this allows some tweaking to obtain the orientation. Then the green line in the adapted axial plane (G) is rotated so it is parallel to the valve leaflets. This results in an orientation perfectly perpendicular to the valve leaflets in the two other imaging planes (H and I) .
and/or use virtual monoenergetic reconstructions at low kV for scanners allowing spectral imaging.
7.4. Anatomical location of the PHV
Contrast injection protocols also depend on the valve of interest. For left-sided valves, a triphasic protocol consisting of a bolus of contrast followed by a mixture of saline and contrast and then pure saline is recommended, similar to that used for coronary CT angiography. 29 For pulmonary valves, apply an injection protocol like that used for CT pulmonary embolism. Adequate visualization of the tricuspid valve with CT is more challenging, owing to the mixing effect occurring in the right atrium from high attenuation inflow from the superior vena cava and non-enhanced attenuation inflow from the inferior vena cava. Reports describe attempts to use a two-site injection (via arms and legs). 48 Alternatively, one can use a tailored triphasic contrast media protocol that includes adapting the amount of the administered contrast media to the left ventricular ejection fraction. 48,49 Decreased contrast flow injection speed during the last phase reduces flow artifacts in the right chambers.
7.5. Image analysis and postprocessing
Dedicated viewing software should be used for image analysis enabling both static and dynamic image reconstructions, scrolling through the images, and changing window width and level settings at the readers’ discretion. The isotropic nature of modern CT images allows high quality reformats in any plane. To obtain reformats in parallel and perpendicular to the prosthesis, rotate the axes to be in line with the valve ring and parallel to the valve leaflets. 23,25 Set the center in the midpoint of the valve orifice in the short-axis and obtain long-axis reformations by rotating the axes while maintaining the annulus plane. This provides an optimal on-edge view of the opening and closing of the leaflet(s). 24,50,51 Fig. 4 illustrates the specific steps. Attempts have been made to automate this process but these solutions are still in initial development stages. 52 Assess the opening and closing angles, defined as angles between the leaflets and the orifice ring, in cine mode and measure them in the fully open and fully closed positions to evaluate leaflet motion. 23,28,50,51 Reconstruct images in the same orientation as the standard TTE and
Consensus statement
Cardiac CT acquisition protocols for PHV assessment should include an arterial phase acquisition, preferably an additional (virtual) non-contrast enhanced acquisition and optionally a delayed phase.
Retrospective ECG-gated or prospective wide-window ECG-triggered arterial phase acquisition is required for the reconstructions of image datasets throughout the cardiac cycle to allow functional evaluation of leaflet motion.
Most contemporary PHV generate only limited artifacts. PHV-induced artifacts may be mitigated by using high kV acquisitions, high keV level monoenergetic reconstructions, iterative reconstruction and prospective ECG-triggering when possible.
Contrast injection should be tailored to the valve of interest with standard injection protocols used for coronary imaging often being well-suited for left sided valves.
Image analysis should be performed using dedicated viewing software that enables both static and dynamic imaging with multiplanar reconstructions in plane and perpendicular to the valve leaflets.
TEE views to allow for direct comparison between modalities if required. 53
8. CT findings in normal functioning PHV
Each PHV type and size has a unique appearance on CT. 35,36,54 In general, most PHV types present with only limited artifacts; with the tilting disc valves mentioned above the notable exceptions. Importantly, the PHV size as labelled by the manufacturer does not represent the actual internal orifice, stent frame, and external dimensions. 55 The implanted PHV size can be accurately determined from the CT images using a standardized measuring approach. 54 This encompasses more than simply measuring the inner diameter of the PHV and we refer the reader to the original publication for detailed instructions and reference values. 54 Lastly, not all prosthetic surgical devices are radio-opaque and visible by CT and a thorough evaluation of the patient’s operative report or alternative imaging modality such as TEE will help guide assessment.
8.1. Prosthetic heart valve positioning
Prosthetic heart valves should be well approximated to the valve annulus. Gaps/spaces between the outer wall of the prosthesis and native valve annulus may indicate paravalvular leak and dehiscence. In cases of transcatheter prosthetic heart valves, the positioning with respect to extension above and below the annular plane varies based on the valve being replaced. Surgical valves may sit intra-annular or supra-annular, based on prosthetic design. TAVR valves will demonstrate a small portion of the stent frame below the annulus. Self-expanding transcatheter PHV, and, to a lesser extent, surgical biological valves, may show a non-circular configuration at the level of the native valve annulus. 41,56
8.2. Leaflet motion and integrity
Use ECG-gated CT acquisitions to evaluate mechanical and bioprosthetic heart valve leaflet motion and integrity in the fully open and fully closed positions.
Mechanical bi-leaflet PHV leaflet angles can be measured in two different ways. On fluoroscopy most physicians measure the angle between both leaflets and communicate this as a single number. On CT however, readers often measure the angle of the individual leaflets with respect to the PHV ring and report it as two separate numbers (one for each leaflet) 23 (Fig. 5). Leaflets should open and close in a symmetrical fashion. 57 Limitation of motion is likely when motion of a leaflet or leaflets is restricted and/or the opening or closing angle measured on CT deviates ${>}4^{\circ}$ compared to the manufacturer value. 24,58
Bioprosthetic valve leaflets can be difficult to see on CT owing to the thinness of the leaflets and the insufficient temporal resolution of CT. If seen however, the leaflets must be smooth and symmetric with complete forward excursion during systole and complete coaptation during diastole. 33,59
8.3. Normal early post-operative and procedural findings
The mode of PHV placement and timing of CT after surgery determine normal postprocedural findings on CT. Normal early post-operative features of surgical valves can include periaortic soft tissue stranding/edema, fluid, and air. 60–62 Unfortunately,
these early, expected post-operative findings can be difficult to differentiate from infection. 61 Suture pledgets appear as small hyperdense foci in a circular pattern around the PHV.
In cases of transcatheter prosthetic heart valves, the majority of the post-procedural features seen with SAVR are not expected. 63
Consensus statement
Each PHV model has a distinct appearance on CT imaging.
Normal mechanical PHV leaflet opening and closing angles may show up to $4^{\circ}$ deviation from manufacturer specifics.
Normal biological PHV leaflets are expected to be smooth and symmetric with complete forward excursion during systole and complete coaptation during diastole.
Self-expanding TAVR valves and biological surgical PHV may show a non-circular configuration at the level of the aortic annulus.
Surgical valves may demonstrate periaortic soft tissue stranding/ edema, fluid, and air in the early post-operative period.
9. Types of PHV dysfunction
9.1. Obstruction
9.1.1. Thrombus and pannus
PHV obstruction may be asymptomatic or present with embolic events, decompensated heart failure, or even car
diogenic shock. Causes of PHV obstruction include thrombus, pannus, valve degeneration and infective endocarditis. 4 Thrombus formation restricting leaflet motion is the most common cause for mechanical PHV obstruction with a reported prevalence of $0.3–1.3\%$ patient-years. 64 Contrary to pannus formation, mechanical valve thrombosis is commonly associated with a sub-therapeutic INR. Mechanical valve obstruction from pannus typically occurs relatively late after implantation after 5 or more years. 65 Importantly, pannus, which is composed of myofibroblasts and collagen fibers, can serve as a nidus for superimposed thrombus. In patients undergoing redo valve replacement for valve obstruction, nearly half demonstrate combined thrombus and pannus formation. 66 The distinction between thrombus and pannus has important clinical implications. Depending on the size and clinical presentation, PHV thrombosis may be treated with anticoagulation and possibly thrombolytics alone. However, clinically significant valve obstruction due to pannus requires redo valve replacement. 9,10
The initial diagnostic evaluation of suspected PHV obstruction includes TTE and TEE echocardiography (and may include cine fluoroscopy in mechanical PHV). 4,14 Cardiac CT can provide important complementary information. Cine fluoroscopy and cardiac CT can both reliably assess restricted leaflet motion by measuring opening and closing angles relative to a reference standard 5,67 (Table 2). However, cine fluoroscopy does not provide insights regarding
 Figure 5: How to measure valve leaflet opening and closing angles for mechanical bileaflet prosthetic heart valves. With fluoroscopy the leaflet angles are usually measured as the angle between the two leaflets. This can be done with CT as well (A and B) . Using the valve ring as the reference plane (yellow line in C and D ), CT also allows measurement of the opening and closing angle of each individual leaflet (green and red lines in C and D ). The angles reported in Table 2 refer to measurements as seen in C and D .
Table 2: Opening and closing angles of specific PHV types. Adapted from Habets et al. 5,21 The list is not intended to include all mechanical PHV types. Some types are no longer available but may still be encountered in patients. Opening and closing angles refer to the angle between the leaflet and the prosthetic valve plane.
|||Leafet opening angle|| |---|---|---|---| |PHVs by type| Metal content|(degrees)|Leafet closing angle (degrees)| |ATS Open Pivot|Titanium alloy ring|85 (AVR)|25 (AVR)| |Carbomedics|Titanium alloy ring|78 (AVR and MVR)|25 (AVR and MVR)| |Duromedics| Cobalt-chrome alloy ring|78 (AVR), 73 (MVR)|20 (AVR and MVR)| |On-X|Titanium alloy ring|90 (AVR and MVR)|40 (AVR and MVR)| |Bicarbon|Titanium alloy ring|80 (AVR and MVR)|20 (AVR and MVR)| | St Jude/Masters/ Regent| Nickel Aloy ring|85 (AVR and MVR)|30 or 25a (AVR and MVR)| | Bjork-Shiley spherical|Cobalt-chrome alloy|60-70# (AVR and MVR)|0 (AVR and MVR)| |Medtronic Hall|Titanium alloy ring|75 (AVR and MVR)|0 (AVR and MVR)| |Omniscience|Titanium alloy ring|60 (AVR)|0 (AVR)| | Sorin All carbon|Cobalt-chrome alloy|60 (AVR and MVR) 0|0 (AVR and MVR)|
a $30^{\circ}$ for 19–25 mm valves, $25^{\circ}$ for 27–31mm valves; # $60^{\circ}$ for valves manufactured before 1981, $70^{\circ}$ for valves manufactured after 1981.
 Figure 6: Prosthetic heart valve obstruction due to pannus tissue underneath mechanical aortic valve prosthesis. Pannus tissue is seen as a thin circular hypodense rim underneath the valve ring (arrows in A and B ). The valve leaflets are restricted and do not open completely (C) and do not close completely (D) .
the underlying cause of obstruction. 68 Accordingly, American and European guidelines both endorse cardiac CT in the evaluation of patients with suspected mechanical valve dysfunction. 9,10
9.1.2. Cardiac CT assessment of thrombus and pannus
PHV thrombus manifests as an abnormal hypodensity at the in- or outflow aspect of the valve leaflets or stent. The morphology of a thrombus is typically irregular, and echocardiography
 Figure 7: Mechanical PHV thrombosis. Images of a patient with a mechanical mitral PHV (Carbomedics, size $29\,\mathsf{m m}$ ) who developed progressive heart failure. Transthoracic echocardiography (TTE) showed severely increased pressure gradients $(\mathsf{P}G)$ (mean 18 mmHg; max 30 mmHg) (A) over the PHV. Fluoroscopy showed immobility of one of the valve leaflets and greatly diminished mobility of the other (D) . CT reconstructions in diastole confirmed fluoroscopy findings of the restricted leaflets (E) . Moreover, CT revealed a hypodense mass attached to the hinges and leaflets and on both the ventricular and atrial side, making thrombus most likely (arrowheads E and F ). Two days after thrombolysis, TTE showed decreased PG (mean 9 mmHg, max 21 mmHg) (B) and fluoroscopy demonstrated improved opening of one leaflet with the other still immobilized (G) . A repeat CT scan confirmed fluoroscopy findings and showed diminished thrombus load (H,I) . TTE performed 6 weeks after discharge showed full mobility of both leaflets and normalized PG (mean 4 mmHg, max 9 mmHg) (C) .
and cine imaging demonstrate independent mobility. Although thrombi may be small, a large hypodense mass is more commonly compatible with thrombus and less likely to be suggestive of pannus. 68
The location of the PHV abnormality helps distinguish pannus versus thrombus. Pannus is commonly located sub-valvular, with a crescenteric or circular morphology as it extends along the valve sewing ring (Fig. 6). Moreover, pannus may have a higher attenuation (i.e., $\mathrm{HU}>145.$ ) as compared to thrombus. 31,69 Based on these characteristics, compared to other imaging modalities, cardiac CT can more readily distinguish pannus from thrombus. Cardiac CT has demonstrated higher sensitivity of $85\%$ for diagnosing pannus compared to TEE (sensitivity $62\%$ ) in a meta-analysis. 70 The pooled
sensitivity and specificity for diagnosing thrombus as the cause of PHV obstruction was $75\%$ and $75\%$ , respectively for TEE vs. $45\%$ and $90\%$ , respectively for CT. 70 In cases of medical management, CT can be used in follow-up to evaluate resolution of thrombus (Fig. 7).
9.1.3. Complementary value of cardiac CT in suspected PHV obstruction
Suchá et al investigated the additional value of cardiac CT in 30 patients with suspected PHV obstruction and showed it provided incremental value to the routine evaluation with TTE and TEE (and cine fluoroscopy in selected cases) in 26/30 $(87\%)$ ) cases for detecting the specific cause of PHV obstruction, which led to a treatment strategy change in 8/30 $(27\%)$ patients. 71
 Figure 8: Paravalvular leakage. Bioprosthetic transcatheter valve in the aortic position. Incomplete expansion of the stent due to severe annular calcification leading to severe PVL (grade 3) left–sided (arrows in A – C ). 3D VRT (A) , axial plane (B) and left sagittal oblique view (C) . Mechanical bileaflet valve in the aortic position with a focal paravalvular leak (arrowheads in D and E ) with contrast enhanced blood next to the valve ring.
9.1.4. Other causes of obstruction
Besides pannus and thrombus, CT can identify other more rare sources of obstruction of PHV, including for instance a narrowed LVOT due to a membrane or complex multifactorial anatomical obstruction following TMVR in patients with a TAVR valve in place as well. 72
Consensus statement
Cardiac CT provides valuable complementary information to echocardiography to determine the cause of PHV obstruction, especially by distinguishing thrombus from pannus.
Thrombus and pannus both present as a hypodense mass on CT imaging, but location and Hounsfield unit value measurements can help to differentiate between both entities.
Cardiac CT has a higher sensitivity than TEE for detecting pannus tissue as the cause of PHV obstruction.
In cases of PHV obstruction due to complex anatomical alterations, CT may provide valuable insight into the mechanism of obstruction.
10. (Para)valvular leakage
The majority of PVL $(74\%)$ is detected in the 1st year following surgery. 73 Causes of PVL include early technical failures from
suture loosening in surgical valves or severe (peri)annular calcifications and suboptimal placement for transcatheter valves. Late PVL is most commonly related to infective endocarditis with valve dehiscence and is discussed more in detail in the endocarditis section below.
Transthoracic and transesophageal, 2D and 3D echocardiography are the primary diagnostic modalities, assessing and subsequently monitoring severity, and can guide percutaneous closure procedures. 14
On CT, the presence of PVL may be suspected by the presence of contrast-enhanced blood around the PHV, in continuity with the adjacent cardiac and aortic structures (Fig. 8). Although cardiac CT cannot directly assess the functional PVL severity, it can provide information about the size and shape of the orifice, as well as the location of the leakage in relation to the surrounding tissues in selected cases. Dynamic evaluation across the cardiac cycle can display abnormal mobility of the PHV (rocking valve) in case of dehiscence.
PVL sizing by CT is commonly performed through phaseby-phase CT analysis to evaluate the largest opening area of the PVL defect to the PHV. In the mitral position, the larger PVL may more commonly be identified during end-systole and for the aortic position during mid-diastole. 74,75 Based on an observed larger regurgitant orifice size, alternatively, the mitral
valve can be evaluated during mid-diastole and the aortic valve during systole. 76 Overall, as motion-augmented beam hardening artifacts generally determine and limit phase selection, readers should compare multiple phases if available. Metal artifacts can significantly affect image quality and preclude assessment of the paravalvular region. Reader must carefully set window level and width to reduce beam-hardening artifacts (wide window width) while preserving the visualization of the sometimes thin surrounding tissues.
The presence of radiopaque suture pledgets can be difficult to distinguish from small paravalvular leaks on contrast-enhanced scans. Multiple pledgets of equal size are typically placed symmetrically around the prosthesis. Comparing the CTA with a non-contrast or post-contrast scan can help differentiate small leaks from pledgets (Fig. 3).
Retrospective cohorts and prospective studies have reported that for the detection of PVL, the accuracy of cardiac CT is equally high as TEE compared to surgical inspection, and outperforms TTE. 74,77,78 However, it is important to note that these publications often excluded low-quality scans so they represent only a selection of patients. Planimetry can measure the regurgitant orifice on in-plane reconstructed CT images and these measurements correlate well with the PVL severity assessed by TEE. 75,76
Of note however, CT lacks flow information, and therefore, smaller PVL can be difficult to visualize, especially if CT image quality is also limited by artifacts. When excessive acoustic shadowing from metal or calcifications limits the interpretation of echocardiographic images, cardiac CT may serve as an adjunct imaging option. Cardiac CT images can provide optimal fluoroscopic angles to optimize pre-procedure planning of transcatheter treatment of PVL. Furthermore, CT imaging provides integral evaluation for precise defect measurement to select appropriate device shape and size. In a prospective study in 12 patients with suspected PHV leakage, CT provided incremental value to TTE and TEE in 7/12 $(58\%)$ ) cases by visualizing pseudoaneurysms and excluding other pathology and CT information resulted in a change in management in 3/12 $(25\%)$ ) cases. 71 Overall, however, caution is advised in using CT for PVL diagnosis and echocardiography is considered the technique of choice for PVL diagnosis and assessment.
PHV. 79 There is very limited data on the use of CT to diagnose PPM. CT can determine the size of the implanted PHV if this information cannot be retrieved from the surgical report. 54 In addition, CT can rule out other causes of PHV obstruction.
Consensus statement
In suspected PPM, CT may primarily be used to rule out other causes of valve obstruction. However, there are very limited data available on the role of CT in suspected PPM.
12. Endocarditis
In PVE, microorganisms adhere to foreign material and form fibrin deposits which can result in vegetations. These microorganisms can also secrete enzymes which degrade tissue and results in leaflet perforation, though bioprosthetic valves are generally more resistant to this degradation compared to native leaflets. Similarly, the infection can lead to paravalvular leak with dehiscence at the valve ring. Invasive endocarditis occurs when infection extends beyond the prosthetic valve, and can result in abscesses and pseudoaneurysms. 80 In PVE, patients can have invasive disease even in the absence of large vegetations or leaflet destruction.
For all patients with known or suspected endocarditis, echocardiography is the first-line imaging investigation. It provides a comprehensive hemodynamic assessment as well as excellent temporal resolution to detect mobile valvular vegetations. Doppler interrogation can readily identify leaflet perforations and cardiac chamber fistulae. 81 However, for patients with known or suspected PVE, the role of Cardiac CT in establishing the diagnosis is increasing. It helps to delineate local complications, identify systemic embolic phenomena, and inform pre-operative planning. 81,82 Cardiac CT is included in the diagnostic flow-chart in both the ESC and AHA guidelines. 9,81 Cardiac CT has a class I indication in the ESC guideline to detect valvular lesions, confirm the diagnosis in possible PVE, and to detect periprosthetic complications when echocardiography is inconclusive. 81 In the 2020 ACC/AHA Valvular Heart Disease Management Guidelines, CT has a class IIa recommendation for evaluation of a suspected paravalvular abscess in cases with inadequate or non-diagnostic echocardiographic imaging. 9
Consensus statement
On CT images paravalvular leakage is defined as contrast-enhanced blood around the valve in continuity with the cardiac cavities on each side of the valve.
In good quality CT scans the diagnostic accuracy of CT for paravalvular leakage detection is in the same range as TEE. However, this represents data from selected populations. Echocardiography is considered superior for PVL detection and CT may easily miss PVL.
Cardiac CT images can also be used in pre-procedure planning of transcatheter treatment of PVL by providing optimal fluoroscopic angles, device sizing, and transseptal approach.
11. Patient prosthesis mismatch
Prosthesis-patient mismatch (PPM) is defined as a too small effective orifice area of the inserted PHV in relation to body size, resulting in higher-than-expected pressure gradients over the
12.1. CT assessment of PVE
Cardiovascular CT can detect vegetations but is inferior to echocardiography due to lower temporal resolution. Small, erratically moving vegetations may be undetected on CT, particularly if only limited phases are acquired. However, in a systematic review with pooled analysis, the accuracy for detecting vegetations by CT ( $78\%$ sensitivity and $94\%$ specificity) is reported to be similar to transesophageal echocardiography (sensitivity and specificity of $89\%$ and $74\%$ ). 83 In PVE, the strengths of cardiovascular CT are to identify and define the extent of abscesses and mycotic aneurysms around the valve (Fig. 9). With CT, abscesses may have low attenuation central necrosis with a peripheral enhancing rim on delayed images. Mycotic aneurysms appear as a contrast-filled outpouching with a connection to the cardiovascular lumen, which is better defined on CT compared to TTE and TEE. CT provides addi
tional diagnostic value compared to TTE and TEE in patients suspected of having endocarditis mainly by detecting mycotic aneurysms. 84,85 CT has a reported sensitivity and specificity to detect periannular complications of $80\%$ and $82\%$ , respectively which are similar to TEE (sensitivity of $77\%$ and specificity of $86\%$ ) in one systematic review. 83
With invasive endocarditis, complications can result as infection spreads along the fibrous skeleton of the heart, resulting in conduction abnormalities and fistulous connections between the chambers of the heart. In delineating the extent of invasive PVE, cardiovascular CT has several advantages compared to echocardiography. With ultrasound, prosthetic valves, calcification, or surgical material may cause reverberation artifacts or acoustic shadowing that limits assessment. With CTA, iodinated contrast allows for a clear distinction between abscess and pseudoaneurysm. In addition, CT provides a more complete field-of-view of the structures of interest, whereas intervening bone or air may preclude adequate assessment with echocardiography.
12.2. PET-CT in PVE
For detecting invasive complications of PVE, 18F-FDG PET-CT has demonstrated good diagnostic accuracy. 15,86,87 Accuracy improves with appropriate preparation to suppress normal myocardial glucose utilization. 88,89 Patients with PVE are more likely to have focal and heterogeneous increased paravalvular 18 F-FDG signal, and a quantitative assessment of uptake may also aid in the diagnosis. 15,90 PET-CT can enhance the diagnostic work-up, but has a limited role in pre-operative planning, with CTA remaining the better image modality for this.
In difficult cases, the decision to pursue a cardiovascular CTA and/or PET-CT for PVE depends on local expertise and discussion with a multidisciplinary endocarditis team. 88 A practical consideration is whether the patient can safely receive iodinated contrast. If so, CTA may better delineate relevant anatomy. When there is an inability to give iodinated contrast and there is diagnostic uncertainty after a clinical evaluation and echocardiography, PET-CT may be more informative. Fusion of the CTA images with the PET allows for optimal co-registration of the areas with increased FDG uptake and anatomical changes. 88
12.3. Additional CT findings
CT can also help define extra-cardiac manifestations of PVE. 82 In right-sided endocarditis, septic pulmonary emboli are a common complication. With left-sided disease, the most feared extra-cardiac manifestations include central nervous system emboli and resultant mycotic aneurysms, and infectious vasculitis. Septic emboli with occlusion of visceral or peripheral arteries can occur as well. Finally, when needed, an endocarditis-protocol CTA can delineate the coronary artery anatomy, as well as the presence and severity of coronary atherosclerosis. 91,92 Invasive angiography may be relatively contra-indicated in patients with aortic PVE and large vegetations. In selected cases when assessment of concomitant coronary artery disease is important, pre-medication may be justified after careful assessment of the patient’s status. However, even without pre-medication and with sub-optimal evaluation of the distal coronary arteries, images can be sufficient to assess the need for surgical coronary revascularization. 91,93
Consensus statement
Cardiac CT plays an important role in the diagnosis of PVE and is firmly included in the ESC and AHA guidelines.
Cardiac CT can detect valvular vegetations but is especially helpful in delineating the extent of infection and detecting periprosthetic complications such as perivalvular mycotic aneurysms.
CTA images may be fused with 18F-FDG PET images to provide optimal anatomical detail and metabolic information.
CT can also play an important role in detecting distant emboli.
In endocarditis patients the Cardiac CT acquisition to evaluate the valve can also be used to assess the status of the coronary arteries in case of a planned re-operation.
13. Structural valve deterioration/dysfunction
Structural bioprosthetic valve dysfunction (BVD) is defined by irreversible changes to the surgical or transcatheter heart valve (THV) leaflet or stent, including fibrosis, collagen disorganization, wear and leaflet injury, fibrosis, and calcification. 94 This differs from non-structural BVD which includes any changes to the bioprosthetic valve that are not related to the prosthetic valve device such as e.g. pannus formation. 94 BVD is a slow process developing over one decade and usually seen 5–8 years after implantation, with a steep incline after 10 years. 95,96 Due to the increasing number of bioprosthetic valves implanted during the past decades, the number of patients with BVD is currently increasing, and expected to rise further. Cardiac CT is currently not used for routine screening for BVD.
Subclinical structural BVD is detected primarily through imaging by either CT and/or echocardiography, which may exhibit features of leaflet thickening or circumferential hypoattenuating material on CT at the inflow of the bioprosthetic valve with absent or mild hemodynamic changes. 4,94 Especially once calcification is evident by CT there are typically hemodynamic changes of the bioprosthesis with rising transvalvular gradients. According to the valve academic research consortium (VARC) 3 guidance document, BVD is graded based on the extent of hemodynamic changes increasing from absent in stage 1 to stage 3 with a mean gradient increase of $20~\mathrm{mm~Hg}$ or a mean gradient of $30~\mathrm{mm~Hg}$ with a commensurate decrease in effective orifice area as per VARC 3 guidelines. 96 BVD transitions to bioprosthetic valve failure with the onset of end organ failure. Progression is from stage 1 with findings of LV hypertrophy or irreversible pulmonary hypertension, over stage 2 with valve re-intervention to finally death at stage 3.
13.1. Imaging findings of BVD
Leaflet calcification can be identified on non-contrast CT, while contrast-enhanced CT can identify hypoattenuated leaflet thickening (HALT) that may affect leaflet mobility. As calcification is a common end stage feature of BVD across many etiologies, CT is helpful to aid in diagnosis (Fig. 10). During the development of BVD, microcalcifications during early stages of BVD are not visible by CT, while during later stages, calcified spots or diffuse calcification of leaflets or annular bioprosthetic material becomes visible.
Radiotracers and the role of PET imaging to assess valve degeneration has been of growing interest. 18F-sodium fluoride
 Figure 9: Endocarditis. PHV in the aortic position (A, B) with vegetation on the valve leaflets (arrowheads in A and B ) and large mycotic aneurysms (asterisks in A and B ). Prosthetic valve in the aortic position (C) with infectious thickening of the fibrous continuity of the mitral and aortic valve (asterisks in C ) and both leaflet calcification (arrowhead in C ) as well as vegetation on the valve leaflets. Transcatheter valve in the aortic position with large vegetation (arrowhead in D ) and mycotic aneurysms (asterisks in D ).
(18F–NaF) PET is a radiotracer that can identify areas of microcalcification $(<\!50~\upmu\mathrm{m})$ allowing for specific detection of developing microcalcifications and active calcific activity. Recent studies have suggested 18F–NaF uptake as a sensitive marker of leaflet degeneration following SAVR and a predictive indicator of THV dysfunction and degeneration. 19,20,97
Consensus statement
CT findings of structural bioprosthetic valve dysfunction include leaflet thickening and calcification of bioprosthetic valves.
Non contrast enhanced CT imaging can detect valve calcifications but lacks the resolution to detect microcalcifications in the early stages of bioprosthetic valve dysfunction.
14. Hypo-attenuated leaflet thickening
Hypo-attenuated leaflet thickening (HALT) can be seen on biological valves. The exact mechanism of bioprosthetic PHV leaflet thrombosis is unclear, but there is growing evidence that the development of leaflet thrombosis represents a complex mechanical and biological process. 98 From a mechanical perspective, incomplete leaflet expansion and reduced neo-sinus volume may play important roles in leaflet thrombosis formation. 99 From a biological perspective, stasis, endothelial injury, and hypercoagulability remain important factors. 100 Prosthetic valve material is thrombogenic and activates the clotting system. Non-physiological blood flow, characterized by areas of high shear stress and/or low velocity/stagnant blood promotes platelet activation and thrombosis. 101
The incidence of HALT in the literature varies by valve type, timing of imaging and design of the study. 102 In randomized trials of low-risk cohorts the incidence of HALT was similar between surgical and transcatheter bioprostheses and across the transcatheter valve designs. The prevalence of subclinical HALT following TAVR was $28\%$ (Sapien 3 valve) and $31\%$ (Evolut system) at one year and comparable to SAVR at one year, at $20\%$ in the PARTNER 3 and $28\%$ in the Evolut Low Risk studies, suggesting this phenomenon equally affects both surgical and transcatheter aortic bioprosthetic valves. 103,104 While HALT may result in valve stenosis and increased pressure gradients there is ongoing debate on the relation between HALT and thromboembolic events as well as structural valve deterioration. 105,106
14.1. CT findings
The hallmark feature of leaflet thrombosis on CT, i.e. HALT, is meniscoid thickening of the bioprosthetic leaflet(s) thickest at the base and thinnest at the tip, with a relative aortic-sided distribution (Fig. 11). Importantly, according to the VARC 3 criteria, subclinical HALT is defined as CT imaging findings in the absence of an elevated echocardiographic gradient or clinical symptoms. 96,107,108 Thrombi characteristically form at the leaflet insertion point and may extend towards the leaflet tip to varying extent. 109–111 HALT is best visualized during the phase when the leaflets coapt (diastolic for aortic valves) and there are minimal artifacts from leaflet motion. The extent of leaflet thickening should be graded on a 4-tier grading score along the curvilinear orientation of the leaflet: ${\le}25\%$ , $25–50\%$ , $50–75\%$ , $>\!75\%$ . 96,112 HALT can then be scored using a 12-point scale
 Figure 10: Structural valve deterioration. Biological prosthetic heart valve in the aortic position. There is severe calcification and thickening of the leaflets compatible with structural valve deterioration. 3D VRT (A) and in-plane reconstructions (B) .
summing-up the scores of up to four assigned to each leaflet depending on the extent of thickening as outlined above. 96
14.2. Reduction in leaflet motion
As the thrombotic burden increases, leaflet opening may be restricted. Varying nomenclatures have been applied to this phenomenon such as reduced leaflet motion and hypo-attenuation affecting motion. 110,113–115 Both terms describe a degree of quantifiable leaflet restriction caused by HALT. 96,109,111 Reduction in leaflet motion is assessed during maximal leaflet excursion, and the degree of leaflet restriction is hence best evaluated in the systolic phase for aortic valves and diastolic phase for mitral valves. 112 For scans with reconstructions during the entire R–R interval evaluating the other phases and cine reconstructions aids in diagnosing reduced leaflet motion. 109 According to SCCT guidelines and overall consensus, once HALT is identified, the presence of reduced leaflet motion should be evaluated. 112
Consensus statement
On CT, hypo-attenuated leaflet thickening (HALT) is characterized by hypodense meniscoid thickening of the bioprosthetic leaflet(s) thickest at the base and thinnest at the leaflet tip.
HALT may be associated with a reduction in leaflet motion, which can be assessed on cine CT reconstructions.
15. Role of CT in preoperative planning for reoperation
CT provides important additional information in case of a planned reoperation. Since the acquisition for valve assessment is ECG synchronized and contrast injection protocols often resemble those for dedicated CCTA, the coronary arteries and bypass grafts can be assessed for stenosis and patency.
CT findings can change the re-entry plan and cannulation strategy for cardiopulmonary bypass. 116 In patients with previous coronary artery bypass grafting, CT can identify the course and patency of previous coronary bypass grafts to help the surgeon avoid injury. Also, assessing the distance between the posterior
sternum and the heart, particularly the right ventricle, can help reduce life-threatening injuries during re-do sternotomy. The aorta can be evaluated for calcifications and iatrogenic aortic dissection as well as aortic root and ascending dilatation that merits concomitant replacement during re-operation, especially in patients with a history of aortic valve replacement for bicuspid aortic valve. 117 Including CT of the abdomen and pelvis can help identify peripheral artery disease that can affect peripheral cannulation via femoral vessels. If surgical risk during re-entry is prohibitively high based on CT scan, a transcatheter valve-in-valve approach may be considered.
In case of a planned valve-in-valve percutaneous re-intervention CT provides important information on the risk of coronary occlusion, aortic root and PHV anatomy for sizing and procedural planning. This is discussed in detail in another SCCT consensus document. 2
Consensus statement
CT provides valuable information in preoperative planning for reoperation after valve surgery.
16. Reporting
Assessment of PHV (dys)function is best done by incorporating all available imaging and other clinical information. To achieve optimal complementary value, the CT report should follow a structured format to ensure all available information is included.
First, the clinical context and diagnostic question should be noted including the type and size of the implanted PHV. Second, procedural features are mentioned including the type of acquisition and ECG synchronization, which phases of the R–R interval have been reconstructed, the amount of contrast medium, and the premedication that has been administrated.
Then, studies for comparison are considered; any prior CT or other studies that could be useful such as echocardiography, MRI or PET/CT and to which the current imaging findings are specifically compared, should be mentioned. Image quality of the scan and specific artifacts should be
 Figure 11: Transcatheter prosthetic heart valve with hypoattenuated leaflet thickening (grading $\mathord{>}\overline{{\gamma}}5\%$ ). Notice the hypodense leaflet thickening which is thicker at the base than at the tip of the leaflet.
reported, especially if this hinders assessment. Location and type of PHV should be described, as well as the structural integrity of the valve. The presence of specific material such as for example suture pledgets should be mentioned. For mechanical valves, leaflet or disk motion and opening/ closing angles should be described and compared to manufacturer specifies. For bioprosthetic valves, leaflet thickening and leaflet motion should be evaluated, as well as signs of PHV degeneration. For all PHV types any complication such as thrombus, pannus tissue ingrowth, paravalvular leak, fistula, abscess or periprosthetic mycotic aneurysm should be reported in terms of size, location and relationship with adjacent structures. Further, cardiac dimensions, coronary arteries and aortic dimensions should be assessed, and extra-cardiac anatomy is evaluated. The report should be concluded with a concise statement of the most relevant findings and should include an answer to the clinical question which integrates the CT findings with those of other available imaging modalities. Table 3 provides a template for a structured report.
In addition to this comprehensive report, we believe it is especially important to discuss imaging findings in a multidisciplinary meeting with the specialties involved in the care of the patient to optimize diagnosis and treatment decisions.
| Table 3: Comprehensive template for structured reporting of CT of PHVs. The template can be adapted to the local situa. tion by adding or leaving out specific parts. ||| |---|---|---| |Report of cardiac CT for PHV assessment||| | Patient data and clinical context||| | Diagnostic question||| |CT acquisition||| | Scan protocol||| |Contrast protocol and amount||| | Drug administration (incl. Dose)||| |Quality of the scan (incl. Specific artifacts)||| |PHV||| |Location||| |Type of PHV||| | Structural integrity||| |Other specific features||| | Dynamic PHV assessment||| | Opening and closing of the valve| Normal/Abnormal|| | Mechanical valves|Opening angle(s) (incl. Ref value)|| |Anatomic PHV assessment|Closing angle(s) (incl. Ref value)|| |Abnormal masses around PHV|Most likely thrombus|Yes/No| || Most likely pannus|Yes/No| |||Yes/No| | PHV angulation| Indeterminate If Yes: Location Normal/Abnormal If abnormal: Describe Presence of vegetations Presence of mycotic aneurysm If Yes: Location || || Indeterminate If Yes: Location Normal/Abnormal If abnormal: Describe Presence of vegetations Presence of mycotic aneurysm If Yes: Location || |Signs of endocarditis||Yes/No| |||Yes/No| |||| ||Other such as:|| ||Abscess (periprosthetic fuid? Encapsulated?|| ||Tissue enhancement?)||
Table 3 ( continued ): Comprehensive template for structured reporting of CT of PHVs. The template can be adapted to the local situation by adding or leaving out specific parts.
|Report of cardiac CT for PHV assessment||| |---|---|---| |Paravalvular leakage|Yes/No|| || If Yes: Location|| |LVOT narowing|Yes/No|| ||If Yes: Location|| |Specific for biological PHV|Systolic valve leaflet opening orifice area|| || PHV degeneration|Yes/no| || Presence of HALT|| ||If Yes: Grading| Yes/No Mild/Moderate/Severe, or as a | || Presence of hypo-attenuation affecting motion If Yes: Grading | Mild/Moderate/Severe, or as a numerical value of 1-12 Yes/No Mild/Moderate/Severe | |Other cardiac findings||| |Atria and ventricles||| |Coronary arteries||| | Prior cardiac surgery/bypass grafts||None/Mild/Moderate/Severe| | Proximal aorta| Dimension Calcifications |None/Mild/Moderate/Severe| | Relevant extra-cardiac fndings||None/Mild/Moderate/Severe|
Consensus statement
CT scans performed for PHV dysfunction assessment are preferably reported using a structured template and findings discussed in a multidisciplinary meeting.
17. Emerging technologies
Recently, photon-counting detector CT has emerged as a promising new technology in CT imaging and is expected to overcome some of the limitations of conventional CT for cardiac and PHV imaging. It represents the next generation of CT hardware technology that delivers ultra-high spatial resolution, and spectral based reconstruction with monoenergetic images. 118,119 Both techniques can reduce blooming and metallic artifacts, which can aid in visualization of valve leaflets. In addition spectral imaging improves the contrast between different tissues and contrast agents. 40,119 Other new developments in
clude the assessment of intracardiac flow patterns using 4D-CT that may provide insight in the development of PHV dysfunction and the influence of PHV orientation. 120,121 However, these potential improvements are still in very early stages and need to be developed further.
18. Conclusion
Echocardiography is the imaging modality of first choice to evaluate PHV function. However, in the last 15 years CT has proven to be a valuable complementary technique to assess PHVs, especially in determining the cause of PHV dysfunction, the diagnosis of PVE and for providing important information in case of a planned re-operation to replace the dysfunctional PHV. In addition, along with the tremendous increase in TAVR procedures that are performed and the ability of CT to accurately detect HALT and restricted leaf
||| Appendix 1. Relationships with Industry of Writing Group Members for Cardiac Computed Tomography for Prosthetic HeartValveAssessment. |||||| |---|---|---|---|---|---|---|---| || Formal represen- Writing Group tative of external organization || Consultant /Speakers Honoraria Bureau || Grants and re- search support ||| |||Employment|Bayer; Sie-||||Patents Royalties| | Ricardo PJ Budde || Erasmus Medical Center, Rotter- dam |mens||||| |(chair)|||||||| | Marguerite E. Faure |||||||| || terdam Center, Rotterdam | Center, Rot- terdam Center, |||||| |Suhny Abbara| Mayo Clinic, Jack- sonville, USA ||||||Elsevier| ||||||||Appendix 1 (continues)|
Appendix 1 ( continued ). Relationships with Industry of Writing Group Members for Cardiac Computed Tomography for Prosthetic Heart Valve Assessment.
| Formal epresen- Writing Group tative of external Consultant / Speakers Stock and Grants and re- Member organization Employment Honoraria Bureau stock options search support Patents Royalties | |---| | Hatem Alkadhi University Hospital Siemens Bayer, Canon, Zurich, Switzer- Health- Guerbet, land ineers Siemens Healthineers | | Paul C. Cremer Northwestern Uni- versity, Chicago, USA | | Gudrun M. Medical Universi- Feuchtner ty, Innsbruck, Austria | | Holly M. American College Sanger Heart and Abbott Gonzales of Cardiology Vascular at (ACC) Atrium Health, Charlotte, USA | | Todd L. Kiefer Duke University, Durham, USA | | Jonathan University of Circle CVI, Philips Circle CVI, Abbott, Boston Leipsic British Colum- Heartlow Heartflow Scientific, bia, Vancouver, Edwards, GE Canada Healthcare, Medtronic | | Koen Nieman Stanford University, Artrya; Novartis, Sie- Stanford, USA Cleerly; mens Novartis; RapidAI | | Jonathan North American New York Univer- Revels Society for sity Langone Cardiovascular Health, New Imaging (NA- York, USA SCI) | | Dee Dee Wang Society for Henry Ford Healthn Abbott, Cardiovascular System, Detroit, Boston Angiography and USA Scientific; Interventions Edwards (SCAI) Life- Sciences, Materi- alise | | Eric William- Radiological Mayo Clinic, Roch- son Society of ester, USA North American (RSNA) | | Moritz C. Society of ThoracicTexas Health Wyler von Surgeons (STS) Harris Methodist Ballmoos Hospital, Fort Worth, USA | | Brittany A. Duke University Zwischen- berger | | Rodrigo Sal- European Society Antwerp University gado of Cadiovascu- Hospital, Ant- lar Radiology werp, Belgium (ESCR) |
Appendix 2. Relationships with Industry of Reviewers for Cardiac Computed Tomography for Prosthetic Heart Valve Assessment.
| Reviewer| Formal representative of external organi- zation |Employment| Consultant/ honoraria | Speakers S bureau | Stock and|Grants and re-|| |---|---|---|---|---|---|---|---| |Ashraf Hamdan|||||||| ||| Rabin Medical Center, Tel Aviv ||||Nanox|| |Hashrul Rashid| Monash University| Brigham and Wom- en's Hospital, Harvard Medical |||||| | Rhanderson N. Cardoso | American Collge of Cardiology (ACC) |||||Cleerly|| | Jean-Nicolas Dacher | European Society of Cardiovas- cular Radiology (ESCR) | University of Rouen Normandy, France | Sanofi; Bayer; Pfzer ||Takeda||| | Laura Jimenez- Juan | ciety for Cardio- vascular Imaging (NASCI) | North American So- St Michael's Hos- pital |||||| |Cristina Fuss| Radiological Society of North America (RSNA) | Yale New Haven Hospital | Elsevier; Siemens ||||Elsevier| |Triston Smith| Society for Cardiovascular Angiography Interventions (SCAI) | Trinity Health System | Edwards Life- sciences ||||| | Karen M. Kim| Society of Thoracic Surgeons (STS) | UT Health Austin/ Dell Medical School ||||||
let motion, the role of CT is expected to only grow in the future. Accordingly, CT is already included in the ESC and AHA guidelines on valvular disease and endocarditis. Moreover, CT has a Class I recommendation in ESC endocarditis guidelines to diagnose both native and prosthetic valve and is a major criterion for the Duke criteria. There is however a relatively limited number of studies available that have investigated cardiac CT in the diagnosis of PHV (dys)function. With this consensus document we have aimed to provide guidance on how to use cardiac CT for PHV assessment based on this available data as well as expert opinion.
Declaration of competing interest
Conflict of interest statements included in Appendix 1.
Author affiliations:
a Department of Radiology & Nuclear Medicine, Erasmus Medical Center, Room Nd-547, Dr. Molewaterplein 40 P.O. Box 2040, 3015 GD Rotterdam, the Netherlands
b Department of Radiology, Mayo Clinic, Jacksonville, Fl, USA c Diagnostic and Interventional Radiology, University Hospital Zurich, Zurich, Switzerland d Departments of Medicine and Radiology, Northwestern University Feinberg School of Medicine, Chicago, IL, USA e Department of Radiology, Innsbruck Medical University, Innsbruck, Austria f Department of Cardiology, Sanger Heart and Vascular at Atrium Health, Charlotte, NC, USA g Department of Cardiology, Duke University Medical Center, Durham, NC, USA h Department of Radiology, University of British Columbia, Vancouver, BC, Canada i Department of Cardiovascular Medicine, Stanford University, Stanford, CA, USA j Department of Radiology, New York University, New York, NY, USA k Center for Structural Heart Disease, Henry Ford Hospital, Detroit, MI, USA l Department of Radiology, Mayo Clinic, Rochester, MN, USA m Department of Cardiac Surgery, Texas Health Harris Methodist Hospital, Fort Worth, Texas, USA n Division of Cardiothoracic Surgery, Department of Surgery, Duke University, Durham, NC, USA o Department of Radiology, Antwerp University Hospital, Antwerp, Belgium Received 5 July 2024; Received in revised form 11 April 2025; Accepted 14 April 2025
Address correspondence to: r.budde@erasmusmc.nl (R.P.J. Budde).
Funding: Authors declared no funding for this work.
References
1. Carroll JD, Mack MJ, Vemulapalli S, et al. STS-ACC TVT registry of transcatheter aortic valve replacement. J Am Coll Cardiol. 2020;76:2492–2516. https://doi.org/10.1016/j.jacc.2020.09.595. 2. Blanke P, Weir-McCall JR, Achenbach S, et al. Computed tomography imaging in the context of transcatheter aortic valve implantation (TAVI)/transcatheter aortic valve replacement (TAVR): an expert consensus document of the Society of Cardiovascular Computed Tomography. J Cardiovasc Comput Tomogr. 2019;13: 1–20. https://doi.org/10.1016/j.jcct.2018.11.008. 3. Francone M, Budde RPJ, Bremerich J, et al. CT and MR imaging prior to transcatheter aortic valve implantation: standardisation of scanning protocols, measurements and reporting-a consensus document by the European Society of Cardiovascular Radiology (ESCR). Eur Radiol. 2020;30:2627–2650. https:// doi.org/10.1007/s00330-019-06357-8.
4. Zoghbi WA, Jone PN, Chamsi-Pasha MA, et al. Guidelines for the evaluation
of prosthetic valve function with cardiovascular imaging: a report from the American society of echocardiography developed in collaboration with the society for cardiovascular magnetic resonance and the society of cardiovascu. J Am Soc Echocardiogr. 2024;37:2–63. https://doi.org/10.1016/j.echo.2023.10.004. 5. Habets J, Budde RP, Symersky P, et al. Diagnostic evaluation of left-sided prosthetic heart valve dysfunction. Nat Rev Cardiol. 2011;8:466–478. https:// doi.org/10.1038/nrcardio.2011.71. 6. Pibarot P, Dumesnil JG. Prosthetic heart valves: selection of the optimal prosthesis and long-term management. Circulation. 2009;119:1034–1048. https://doi.org/10.1161/CIRCULATIONAHA.108.778886. 7. Shrestha M, Fischlein T, Meuris B, et al. European multicentre experience with the sutureless Perceval valve: clinical and haemodynamic outcomes up to 5 years in over 700 patients. Eur J Cardio Thorac Surg. 2016;49:234–241. https://doi.org/10.1093/ejcts/ezv040. 8. Wang DD, O’Neill BP, Caranasos TG, Chitwood WR, Stack RS, O’Neill WW. Comparative differences of mitral valve-in-valve implantation: a new mitral bioprosthesis versus current mosaic and epic valves. Cathet Cardiovasc Interv. 2022; 99:934–942. https://doi.org/10.1002/ccd.30011. 9. Otto CM, Nishimura RA, Bonow RO, et al. 2020 ACC/AHA Guideline for the Management of Patients with Valvular Heart Disease: Executive Summary: a Report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. 2021. https:// doi.org/10.1161/CIR.0000000000000932. 10. Vahanian A, Beyersdorf F. 2021 ESC/EACTS Guidelines for the management of valvular heart disease. Eur J Cardio Thorac Surg. 2021;60:727–800. https:// doi.org/10.1093/ejcts/ezab389. 11. Pibarot P, Herrmann HC, Wu C, et al. Standardized definitions for bioprosthetic valve dysfunction following aortic or mitral valve replacement: JACC state-of-theart review. J Am Coll Cardiol. 2022;80:545–561. https://doi. org/10.1016/j.jacc.2022.06.002. 12. Hahn RT, Leipsic J, Douglas PS, et al. Comprehensive echocardiographic assessment of normal transcatheter valve function. JACC Cardiovasc Imag. 2019;12: 25–34. https://doi.org/10.1016/j.jcmg.2018.04.010. 13. Quien MM, Saric M. Ultrasound imaging artifacts: how to recognize them and how to avoid them. Echocardiography. 2018;35:1388–1401. https://doi. org/10.1111/echo.14116. 14. Lancellotti P, Pibarot P, Chambers J, et al. Recommendations for the imaging assessment of prosthetic heart valves: a report from the European association of cardiovascular imaging endorsed by the Chinese society of echocardiography, the inter-American society of echocardiography, and the Brazilian. Eur Heart J Cardiovasc Imag. 2016;17:589–590. https://doi.org/10.1093/ehjci/jew025. 15. Swart LE, Gomes A, Scholtens AM, et al. Improving the diagnostic performance of 18F-fluorodeoxyglucose positron-emission tomography/computed tomography in prosthetic heart valve endocarditis. Circulation. 2018;138:1412–1427. https://doi.org/10.1161/CIRCULATIONAHA.118.035032. 16. Gomes A, Glaudemans AWJM, Touw DJ, et al. Diagnostic value of imaging in infective endocarditis: a systematic review. Lancet Infect Dis. 2017;17:e1–e14. https://doi.org/10.1016/S1473-3099(16)30141-4. 17. Kooshki N, Grambow-Velilla J, Mahida B, et al. Diagnostic performance of White Blood Cell SPECT imaging against intra-operative findings in patients with a suspicion of prosthetic valve endocarditis. J Nucl Cardiol. 2022;29:528–534. https://doi.org/10.1007/s12350-021-02674-y. 18. Scholtens AM, Swart LE, Verberne HJ, Tanis W, Lam MGEH, Budde RPJ. Confounders in FDG-PET/CT imaging of suspected prosthetic valve endocarditis. JACC Cardiovasc Imag. 2016;9. https://doi.org/10.1016/j. jcmg.2016.01.024. 19. Cartlidge TRG, Doris MK, Sellers SL, et al. Detection and prediction of bioprosthetic aortic valve degeneration. J Am Coll Cardiol. 2019;73:1107–1119. https://doi.org/10.1016/j.jacc.2018.12.056. 20. Kwiecinski J, Tzolos E, Cartlidge TRG, et al. Native aortic valve disease progression and bioprosthetic valve degeneration in patients with transcatheter aortic valve implantation. Circulation. 2021;144:1396–1408. https://doi. org/10.1161/CIRCULATIONAHA.121.056891. 21. Bing R, Deutsch MA, Sellers SL, et al. 18F-GP1 positron emission tomography and bioprosthetic aortic valve thrombus. JACC Cardiovasc Imag. 2022;15:1107–1120. https://doi.org/10.1016/j.jcmg.2021.11.015. 22. Rajiah P, Moore A, Saboo S, et al. Multimodality imaging of complications of cardiac valve surgeries. Radiographics. 2019;39:932–956. https://doi. org/10.1148/rg.2019180177. 23. Habets J, Mali WPTM, Budde RPJ. Multidetector CT angiography in evaluation of prosthetic heart valve dysfunction. Radiographics. 2012;32:1893–1905. https://doi.org/10.1148/rg.327125702. 24. Suchá D, Symersky P, Vonken E-JPA, Provoost E, Chamuleau SAJ, Budde RPJ. Multidetector-row computed tomography allows accurate measurement of mechanical prosthetic heart valve leaflet closing angles compared with fluoroscopy. J Comput Assist Tomogr. 2014;38. https://doi.org/10.1097/ Computed tomography vs. cinefluoroscopy for the assessment of mechanical prosthetic valve leaflet motion. Heart Vessel. 2023;38:599–605. https://doi. org/10.1007/s00380-022-02193-x. 26. Suchá D, Symersky P, Tanis W, et al. Multimodality imaging assessment of prosthetic heart valves. Circ Cardiovasc Imag. 2015;8:1–14. https://doi. org/10.1161/CIRCIMAGING.115.003703. 27. Faure ME, Swart LE, Dijkshoorn ML, et al. Advanced CT acquisition protocol with a third-generation dual-source CT scanner and iterative reconstruction technique for comprehensive prosthetic heart valve assessment. Eur Radiol. 2018;28:2159–2168. https://doi.org/10.1007/s00330-017-5163-7. 28. Bazeed MF, Moselhy MS, Rezk AI, Al-Murayeh MA. Low radiation dose non-contrast cardiac CT: is it of value in the evaluation of mechanical aortic valve. Acta Radiol. 2012;53:389–393. https://doi.org/10.1258/ ar.2012.110253. 29. Feuchtner G. Imaging of cardiac valves by computed tomography. Sci Tech Rep. 2013;2013:1–13. https://doi.org/10.1155/2013/270579. 30. Suh YJ, Im DJ, Hong YJ, et al. Absolute-delay multiphase reconstruction reduces prosthetic valve-related and atrial fibrillation-related artifacts at cardiac CT. Am J Roentgenol. 2017;208:W160–W167. https://doi.org/10.2214/AJR.16.16839. 31. Gündüz S, Özkan M, Kalçik M, et al. Sixty-four-section cardiac computed tomography in mechanical prosthetic heart valve dysfunction: thrombus or pannus. Circ Cardiovasc Imag. 2015;8:1–11. https://doi.org/10.1161/ CIRCIMAGING.115.003246. 32. Tsai IC, Lin YK, Chang Y, et al. Correctness of multi-detector-row computed tomography for diagnosing mechanical prosthetic heart valve disorders using operative findings as a gold standard. Eur Radiol. 2009;19:857–867. https:// doi.org/10.1007/s00330-008-1232-2. 33. Chenot F, Montant P, Goffinet C, et al. Evaluation of anatomic valve opening and leaflet morphology in aortic valve bioprosthesis by using multidetector CT: comparison with transthoracic echocardiography. Radiology. 2010;255:377–385. https://doi.org/10.1148/radiol.0000082294. 34. Habets J, Symersky P, Leiner T, De Mol BAJM, Mali WPTM, Budde RPJ. Artifact reduction strategies for prosthetic heart valve CT imaging. Int J Cardiovasc Imag. 2012;28. https://doi.org/10.1007/s10554-012-0041-5. 35. Habets J, Symersky P, van Herwerden LA, et al. Prosthetic heart valve assessment with multidetector-row CT: imaging characteristics of 91 valves in 83 patients. Eur Radiol. 2011;21:1390–1396. https://doi.org/10.1007/ s00330-011-2068-8. 36. Schwartz FR, Tailor T, Gaca JG, et al. Impact of dual energy cardiac CT for metal artefact reduction post aortic valve replacement. Eur J Radiol. 2020;129. https://doi.org/10.1016/j.ejrad.2020.109135. 37. Symersky P, Habets J, Westers P, De Mol BAJM, Prokop M, Budde RPJ. Prospective ECG triggering reduces prosthetic heart valve-induced artefacts compared with retrospective ECG gating on 256-slice CT. Eur Radiol. 2012;22. https://doi.org/10.1007/s00330-011-2358-1. 38. Habets J, Symersky P, De Mol BAJM, Mali WPTM, Leiner T, Budde RPJ. A novel iterative reconstruction algorithm allows reduced dose multidetector-row CT imaging of mechanical prosthetic heart valves. Int J Cardiovasc Imag. 2012;28. https://doi.org/10.1007/s10554-011-9954-7. 39. Suchá D, Willemink MJ, De Jong PA, et al. The impact of a new model-based iterative reconstruction algorithm on prosthetic heart valve related artifacts at reduced radiation dose MDCT. Int J Cardiovasc Imag. 2014;30. https://doi. org/10.1007/s10554-014-0379-y. 40. van der Bie J, Sharma SP, van Straten M, et al. Photon-counting detector CT in patients pre- and post-transcatheter aortic valve replacement. Radiol Cardiothorac Imag. 2023;5:2–6. https://doi.org/10.1148/ryct.220318. 41. Faure ME, Suchá D, Schwartz FR, et al. Surgically implanted aortic valve bioprostheses deform after implantation: insights from computed tomography. Eur Radiol. 2020;30. https://doi.org/10.1007/s00330-019-06634-6. 42. Habets J, Meijer TS, Meijer RCA, Mali WPTM, Vonken E-JPA, Budde RPJ. CT attenuation measurements are valuable to discriminate pledgets used in prosthetic heart valve implantation from paravalvular leakage. Br J Radiol. 2012;85. https://doi.org/10.1259/bjr/29602784. 43. van Hamersvelt RW, de Jong PA, Dessing TC, Leiner T, Willemink MJ. Dual energy CT to reveal pseudo leakage of frozen elephant trunk. J Cardiovasc Comput Tomogr. 2017;11:240–241. https://doi.org/10.1016/j. jcct.2016.11.001. 44. Rajiah PS, Alkadhi H, Van Mieghem NM, Budde RPJ. Utility of photon counting CT in transcatheter structural heart disease interventions. Semin Roentgenol. 2023. https://doi.org/10.1053/j.ro.2023.11.005. 45. Tanis W, Such a D, Laufer W, et al. Multidetector-row computed tomography for prosthetic heart valve dysfunction: is concomitant non-invasive coronary angiography possible before redo-surgery? Eur Radiol. 2015;25. https://doi. org/10.1007/s00330-014-3551-9. 46. Cundari G, Alkadhi H, Eberhard M. The role of CT in arrhythmia management-treatment planning and post-procedural imaging surveillance. Br J 47. Feuchtner GM, Stolzmann P, Dichtl W, et al. Multislice computed tomography in infective endocarditis: comparison with transesophageal echocardiography and intraoperative findings. J Am Coll Cardiol. 2009;53:436–444. https:// doi.org/10.1016/j.jacc.2008.01.077. 48. Han BK, Rigsby CK, Leipsic J, et al. Society of cardiovascular computed tomography, society of pediatric Radiology, North American society of cardiac imaging, computed tomography imaging in patients with congenital heart disease, Part 2: technical recommendations. An expert consensus document of the society of cardiovascular computed tomography (SCCT): endorsed by the society of pediatric Radiology (SPR) and th. J Cardiovasc Comput Tomogr. 2015;9:493–513. https://doi.org/10.1016/j.jcct.2015.07.007. 49. Hinzpeter R, Eberhard M, Burghard P, et al. Computed tomography in patients with tricuspid regurgitation prior to transcatheter valve repair: dynamic analysis of the annulus with an individually tailored contrast media protocol. EuroIntervention. 2017;12:e1828–e1836. https://doi.org/10.4244/EIJ-D-16-00891. 50. LaBounty TM, Agarwal PP, Chughtai A, Bach DS, Wizauer E, Kazerooni EA. Evaluation of mechanical heart valve size and function with ECG-gated 64-MDCT. Am J Roentgenol. 2009;193:389–396. https://doi.org/10.2214/ AJR.08.2178. 51. Suh YJ, Kim YJ, Hong YJ, Lee H. Measurement of opening and closing angles of aortic valve prostheses in vivo using dual-source computed tomography: comparison with those of manufacturers’ in 10 different types. Korean J Radiol. 2015;16:1012–1023. 52. Androulakis I, Faure ME, Budde RPJ, van Walsum T. Automated quantification of bileaflet mechanical heart valve leaflet angles in CT images. IEEE Trans Med Imag. 2019;38:753–761. https://doi.org/10.1109/TMI.2018.2871366. 53. Habets J, Tanis W, Mali WPTM, Chamuleau SAJ, Budde RPJ. Imaging of prosthetic heart valve dysfunction: complementary diagnostic value of TEE and MDCT? JACC Cardiovasc Imag. 2012;5. https://doi.org/10.1016/j. jcmg.2012.03.018. 54. Suchá D, Daans CG, Symersky P, et al. Reliability, agreement, and presentation of a reference standard for assessing implanted heart valve sizes by multidetector-row computed tomography. Am J Cardiol. 2015;116:112–120. https:// doi.org/10.1016/j.amjcard.2015.03.048. 55. Durko AP, Head SJ, Pibarot P, et al. Characteristics of surgical prosthetic heart valves and problems around labeling: a document from the European association for cardio-thoracic surgery (EACTS)—the society of thoracic surgeons (STS)— American association for thoracic surgery (AATS) valve La. J Thorac Cardiovasc Surg. 2019;158:1041–1054. https://doi.org/10.1016/j. jtcvs.2019.04.001. 56. Kazuno Y, Maeno Y, Kawamori H, et al. Comparison of SAPIEN 3 and SAPIEN XT transcatheter heart valve stent-frame expansion: evaluation using multi-slice computed tomography. Eur Heart J Cardiovasc Imag. 2016;17:1054–1062. https://doi.org/10.1093/ehjci/jew032. 57. Konen E, Goitein O, Feinberg MS, et al. The role of ECG-gated MDCT in the evaluation of aortic and mitral mechanical valves: initial experience. Am J Roentgenol. 2008;191:26–31. https://doi.org/10.2214/AJR.07.2951. 58. Symersky P, Budde RPJ, de Mol BAJM, Prokop M. Comparison of multidetector-row computed tomography to echocardiography and fluoroscopy for evaluation of patients with mechanical prosthetic valve obstruction. Am J Cardiol. 2009;104: 1128–1134. https://doi.org/10.1016/j.amjcard.2009.05.061. 59. De Heer LM, Habets J, Kluin J, et al. Assessment of a transcatheter heart valve prosthesis with multidetector computed tomography: in vitro and in vivo imaging characteristics. Int J Cardiovasc Imag. 2013;29:659–668. https:// doi.org/10.1007/s10554-012-0111-8. 60. Suchá D, Chamuleau SAJ, Symersky P, et al. Baseline MDCT findings after prosthetic heart valve implantation provide important complementary information to echocardiography for follow-up purposes. Eur Radiol. 2016;26. https://doi.org/10.1007/s00330-015-3918-6. 61. Hanneman K, Chan FP, Mitchell RS, Miller DC, Fleischmann D. Pre- and postoperative imaging of the aortic root. RadioGraphics. 2016;36:19–37. https:// doi.org/10.1148/rg.2016150053. 62. Boccalini S, Swart LE, Bekkers JA, et al. Peri-aortic fluid after surgery on the ascending aorta: worrisome indicator of complications or innocent postoperative finding? Eur J Radiol. 2017;95:332–341. https://doi.org/10.1016/j. ejrad.2017.09.001. 63. Salgado RA, Budde RPJ, Leiner T, et al. Transcatheter aortic valve replacement: postoperative CT findings of Sapien and corevalve transcatheter heart valves. RadioGraphics. 2014;34:1517–1537. https://doi.org/10.1148/rg.346130149. 64. Roudaut R, Serri K, Lafitte S. Thrombosis of prosthetic heart valves: diagnosis and therapeutic considerations. Heart. 2007;93:137–142. https://doi. org/10.1136/hrt.2005.071183. 65. Li J, Wang S, Sun H, et al. Clinical and surgical evaluations of reoperation after mechanical mitral valve replacement due to different etiologies. Front Cardiovasc Med. 2021;8:778750. https://doi.org/10.3389/fcvm.2021.778750. 66. Vitale N, Renzulli A, Agozzino L, et al. Obstruction of mechanical mitral prostheses: analysis of pathologic findings. Ann Thorac Surg. 1997;63:1101–1106. https://doi.org/10.1016/s0003-4975(96)01391-4. 67. Tanis W, Budde RPJ, van der Bilt IAC, et al. Novel imaging strategies for the detection of prosthetic heart valve obstruction and endocarditis. Netherlands Hear J. 2016. https://doi.org/10.1007/s12471-015-0796-0. 68. Tanis W, Habets J, Van Den Brink RBA, Symersky P, Budde RPJ, Chamuleau SAJ. Differentiation of thrombus from pannus as the cause of acquired mechanical prosthetic heart valve obstruction by non-invasive imaging: a review of the literature. Eur Heart J Cardiovasc Imag. 2014;15:119–129. https://doi.org/10.1093/ehjci/jet127. 69. Nam K, Suh YJ, Han K, Park SJ, Kim YJ, Choi BW. Value of computed tomography radiomic features for differentiation of periprosthetic mass in patients with suspected prosthetic valve obstruction. Circ Cardiovasc Imag. 2019;12:e009496. https://doi.org/10.1161/CIRCIMAGING.119.009496. 70. Kim JY, Suh YJ, Han K, Kim YJ, Choi BW. Diagnostic value of advanced imaging modalities for the detection and differentiation of prosthetic valve obstruction: a systematic review and meta-analysis. JACC Cardiovasc Imag. 2019;12:2182–2192. https://doi.org/10.1016/j.jcmg.2018.11.033. 71. Suchá D, Symersky P, van den Brink RBA, et al. Diagnostic evaluation and treatment strategy in patients with suspected prosthetic heart valve dysfunction: the incremental value of MDCT. J Cardiovasc Comput Tomogr. 2016;10:398–406. https://doi.org/10.1016/j.jcct.2016.07.008. 72. Faure ME, Van Mieghem NMDA, Budde RPJ. Transcatheter tricuspid valve-inring placement: complex valve obstruction by hypo-attenuating leaflet thickening, hypoattenuation affecting motion, and native tricuspid valve remnant. Eur Heart J. 2020; 41:973. https://doi.org/10.1093/eurheartj/ehy846. 73. Shah VN, Orlov OI, Orlov CP, et al. Incidence, natural history, and factors associated with paravalvular leak following surgical aortic valve replacement. Innovations. 2019;14:519–530. https://doi.org/10.1177/1556984519874806. 74. Hassan AY, Kandil HI, Salem MA, Mohamed AS, Kassem HH. Diagnostic accuracy of cardiac computed tomographic angiography and transesophageal echocardiography in evaluation of patients with prosthetic paravalvular leakage. Echocardiography. 2022;39:1122–1130. https://doi.org/10.1111/ echo.15428. 75. Hascoet S, Smolka G, Bagate F, et al. Multimodality imaging guidance for percutaneous paravalvular leak closure: insights from the multi-centre FFPP register. Arch Cardiovasc Dis. 2018;111:421–431. https://doi.org/10.1016/j. acvd.2018.05.001. 76. Koo HJ, Lee JY, Kim GH, et al. Paravalvular leakage in patients with prosthetic heart valves: cardiac computed tomography findings and clinical features. Eur Heart J Cardiovasc Imag. 2018;19:1419–1427. https://doi.org/10.1093/ehjci/ jex341. 77. Hara M, Nishino M, Taniike M, et al. Impact of 64 multi-detector computed tomography for the evaluation of aortic paraprosthetic regurgitation. J Cardiol. 2011;58:294–299. https://doi.org/10.1016/j.jjcc.2011.08.002. 78. Suh YJ, Hong G-R, Han K, et al. Assessment of mitral paravalvular leakage after mitral valve replacement using cardiac computed tomography: comparison with surgical findings. Circ Cardiovasc Imag. 2016;9. https://doi.org/10.1161/ CIRCIMAGING.115.004153. 79. Pibarot P, Dumesnil JG. Prosthesis-patient mismatch: definition, clinical impact, and prevention. Heart. 2006;92:1022–1029. https://doi.org/10.1136/ hrt.2005.067363. 80. Pettersson GB, Hussain ST. Current AATS guidelines on surgical treatment of infective endocarditis. Ann Cardiothorac Surg. 2019;8:630–644. https:// doi.org/10.21037/acs.2019.10.05. 81. Marsan NA, De Waha S, Germany TFC, et al. 2023 ESC Guidelines for the Management of Endocarditis Developed by the Task Force on the Management of Endocarditis of the European Society of Cardiology (ESC) Endorsed by the European Association for Cardio-Thoracic Surgery (EACTS) and the European. 2023:1–95. 82. Bin Saeedan M, Wang TKM, Cremer P, et al. Role of cardiac CT in infective endocarditis: current evidence, opportunities, and challenges. Radiol Cardiothorac Imag. 2021;3:e200378. https://doi.org/10.1148/ryct.2021200378. 83. Jain V, Wang TKM, Bansal A, et al. Diagnostic performance of cardiac computed tomography versus transesophageal echocardiography in infective endocarditis: a contemporary comparative meta-analysis. J Cardiovasc Comput Tomogr. 2021;15: 313–321. https://doi.org/10.1016/j.jcct.2020.11.008. 84. Habets J, Tanis W, a van Herwerden L, et al. Cardiac computed tomography angiography results in diagnostic and therapeutic change in prosthetic heart valve endocarditis. Int J Cardiovasc Imag. 2014;30:377–387. https://doi. org/10.1007/s10554-013-0335-2. 85. Fagman E, Flinck A, Snygg-Martin U, Olaison L, Bech-Hanssen O, Svensson G. Surgical decision-making in aortic prosthetic valve endocarditis: the influence of electrocardiogram-gated computed tomography. Eur J Cardio Thorac Surg. 2016;50: 1165–1171. https://doi.org/10.1093/ejcts/ezw177. 86. Wang TKM, Sánchez-Nadales A, Igbinomwanhia E, Cremer P, Griffin B, Xu B. Diagnosis of infective endocarditis by subtype using 18F-fluorodeoxyglucose positron emission tomography/computed tomography: a contemporary metaanalysis. Circ Cardiovasc Imag. 2020;13:e010600. https://doi.org/10.1161/ 87. O’Gorman P, Nair L, Kisiel N, et al. Meta-analysis assessing the sensitivity and specificity of 18F-FDG PET/CT for the diagnosis of prosthetic valve endocarditis (PVE) using individual patient data (IPD). Am Heart J. 2023;261:21–34. https://doi.org/10.1016/j.ahj.2023.03.004. 88. Swart LE, Scholtens AM, Tanis W, et al. 18F-fluorodeoxyglucose positron emission/ computed tomography and computed tomography angiography in prosthetic heart valve endocarditis: from guidelines to clinical practice. Eur Heart J. 2018;39: 3739–3749. https://doi.org/10.1093/eurheartj/ehx784. 89. Madamanchi C, Weinberg RL, Murthy VL. Utility of serum ketone levels for assessment of myocardial glucose suppression for 18F-fluorodeoxyglucose PET in patients referred for evaluation of endocarditis. J Nucl Cardiol. 2023;30:928–937. https://doi.org/10.1007/s12350-023-03209-3. 90. Roque A, Pizzi MN, Fernández-Hidalgo N, et al. The valve uptake index: improving assessment of prosthetic valve endocarditis and updating [18F]FDG PET/CT(A) imaging criteria. Eur Heart J Cardiovasc Imag. 2022;23:1260– 1271. https://doi.org/10.1093/ehjci/jeab279. 91. Knol WG, Wahadat AR, Roos-Hesselink JW, et al. Screening for coronary artery disease in early surgical treatment of acute aortic valve infective endocarditis. Interact Cardiovasc Thorac Surg. 2021;32:522–529. https://doi. org/10.1093/icvts/ivaa313. 92. Habets J, van den Brink RBA, Uijlings R, et al. Coronary artery assessment by multidetector computed tomography in patients with prosthetic heart valves. Eur Radiol. 2012;22:1278–1286. https://doi.org/10.1007/s00330- 011-2360-7. 93. Sims JR, Anavekar NS, Chandrasekaran K, et al. Utility of cardiac computed tomography scanning in the diagnosis and pre-operative evaluation of patients with infective endocarditis. Int J Cardiovasc Imag. 2018;34:1155–1163. https://doi.org/10.1007/s10554-018-1318-0. 94. Capodanno D, Petronio AS, Prendergast B, et al. Standardized definitions of structural deterioration and valve failure in assessing long-term durability of transcatheter and surgical aortic bioprosthetic valves: a consensus statement from the European Association of Percutaneous Cardiovascular Interven. Eur Heart J. 2017;38:3382–3390. https://doi.org/10.1093/eurheartj/ehx303. 95. Dvir D, Bourguignon T, Otto CM, et al. VIVID (valve in valve international data) investigators, standardized definition of structural valve degeneration for surgical and transcatheter bioprosthetic aortic valves. Circulation. 2018;137:388–399. https://doi.org/10.1161/CIRCULATIONAHA.117.030729. 96. VARC-3 WRITING COMMITTEE:, Généreux P, Piazza N, Alu MC. Valve academic research consortium 3: updated endpoint definitions for aortic valve clinical research. J Am Coll Cardiol. 2021;77:2717–2746. https://doi. org/10.1016/j.jacc.2021.02.038. 97. Tzolos E, Andrews JP, Dweck MR. Aortic valve stenosis-multimodality assessment with PET/CT and PET/MRI. Br J Radiol. 2020;93:20190688. https://doi.org/10.1259/bjr.20190688. 98. Puri R, Auffret V, Rodés-Cabau J. Bioprosthetic valve thrombosis. J Am Coll Cardiol. 2017;69:2193–2211. https://doi.org/10.1016/j.jacc.2017.02.051. 99. Rashid HN, Cameron JD, Gooley RP. Early bioprosthetic valve dysfunction following TAVR: the role of CT imaging in diagnosing acute leaflet thrombosis, hear. Lung Circ. 2022;31:e120–e122. https://doi.org/10.1016/j. hlc.2022.03.006. 100. Dangas GD, Weitz JI, Giustino G, Makkar R, Mehran R. Prosthetic heart valve thrombosis. J Am Coll Cardiol. 2016;68:2670–2689. https://doi. org/10.1016/j.jacc.2016.09.958. 101. Lim WY, Lloyd G, Bhattacharyya S. Mechanical and surgical bioprosthetic valve thrombosis. Heart. 2017;103:1934–1941. https://doi.org/10.1136/ heartjnl-2017- 311856. 102. Woldendorp K, Doyle MP, Black D, et al. Subclinical valve thrombosis in transcatheter aortic valve implantation: a systematic review and meta-analysis. J Thorac Cardiovasc Surg. 2021;162:1491–1499.e2. https://doi.org/10.1016/j. jtcvs.2020.01.084. 103. Makkar RR, Blanke P, Leipsic J, et al. Subclinical leaflet thrombosis in transcatheter and surgical bioprosthetic valves: PARTNER 3 cardiac computed tomography substudy. J Am Coll Cardiol. 2020;75:3003–3015. https://doi. org/10.1016/j.jacc.2020.04.043. 104. Blanke P, Leipsic JA, Popma JJ, et al. Evolut low risk LTI substudy investigators, bioprosthetic aortic valve leaflet thickening in the Evolut low risk sub-study. J Am Coll Cardiol. 2020;75:2430–2442. https://doi.org/10.1016/j. jacc.2020.03.022. 105. Bogyi M, Schernthaner RE, Loewe C, et al. Subclinical leaflet thrombosis after transcatheter aortic valve replacement: a meta-analysis. JACC Cardiovasc Interv. 2021;14:2643–2656. https://doi.org/10.1016/j.jcin.2021.09.019. 106. Rashid HN, Michail M, Ramnarain J, et al. The impact of hypo-attenuated leaflet thickening on haemodynamic valve deterioration following transcatheter aortic valve replacement. J Cardiovasc Comput Tomogr. 2022;16:168–173. https://doi.org/10.1016/j.jcct.2021.11.013. 107. Hansson NC, Grove EL, Andersen HR, et al. Transcatheter aortic valve thrombosis: incidence, predisposing factors, and clinical implications. J Am Coll Cardiol. 2016; 68:2059–2069. https://doi.org/10.1016/j.jacc.2016.08.010. 108. Ahmad Y, Makkar R, Sondergaard L. Hypoattenuated leaflet thickening (HALT) and reduced leaflet motion (RELM) of aortic bioprostheses: an imaging finding or a complication? Prog Cardiovasc Dis. 2022;72:78–83. https://doi.org/10.1016/j.pcad.2022.05.007. 109. Jilaihawi H, Asch FM, Manasse E, et al. Systematic CT methodology for the evaluation of subclinical leaflet thrombosis. JACC Cardiovasc Imag. 2017;10: 461–470. https://doi.org/10.1016/j.jcmg.2017.02.005. 110. Rashid HN, Rajani R, Leipsic J, et al. Computed tomography imaging for subclinical leaflet thrombosis following surgical and transcatheter aortic valve replacement. J Cardiovasc Comput Tomogr. 2023;17:2–10. https://doi. org/10.1016/ j.jcct.2022.11.001. 111. Ng ACT, Holmes DR, Mack MJ, et al. Leaflet immobility and thrombosis in transcatheter aortic valve replacement. Eur Heart J. 2020;41:3184–3197. https://doi.org/10.1093/eurheartj/ehaa542. 112. Blanke P, Weir-McCall JR, Achenbach S, et al. Computed tomography imaging in the context of transcatheter aortic valve implantation (TAVI)/ Transcatheter aortic valve replacement (TAVR): an expert consensus document of the society of cardiovascular computed tomography. JACC Cardiovasc Imag. 2019;12:1–24. https://doi.org/10.1016/j.jcmg.2018.12.003. 113. Sondergaard L, De Backer O, Kofoed KF, et al. Natural history of subclinical leaflet thrombosis affecting motion in bioprosthetic aortic valves. Eur Heart J. 2017;38: 2201–2207. https://doi.org/10.1093/eurheartj/ehx369. 114. Chakravarty T, Søndergaard L, Friedman J, et al. Subclinical leaflet thrombosis in surgical and transcatheter bioprosthetic aortic valves: an observational study. Lancet. 2017;389:2383–2392. https://doi.org/10.1016/S0140-6736(17) 30757-2. 115. Rosseel L, De Backer O, Søndergaard L. Clinical valve thrombosis and subclinical leaflet thrombosis in transcatheter aortic heart valves: clinical manifestations, diagnosis, and treatment. Precis Clin Med. 2018;1:111–117. https://doi.org/10.1093/pcmedi/pby016. 116. Miller CL, Kocher M, Koweek LH, Zwischenberger BA. Use of computed tomography (CT) for preoperative planning in patients undergoing coronary artery bypass grafting (CABG). J Card Surg. 2022;37:4150–4157. https:// doi.org/10.1111/jocs.17000. 117. Knol WG, Simon J, Den Harder AM, et al. Effect of routine preoperative screening for aortic calcifications using noncontrast computed tomography on stroke rate in cardiac surgery: the randomized controlled CRICKET study. Eur Radiol. 2022;32: 2611–2619. https://doi.org/10.1007/s00330-021-08360-4. 118. Boccalini S, Mayard C, Lacombe H, et al. Ultra-high-resolution and K-edge imaging of prosthetic heart valves with spectral. Photon-Counting CT. 2024;00:1–10. https://doi.org/10.1097/RLI.0000000000001068. 119. Douek PC, Boccalini S, Oei EHG, et al. Clinical applications of photon-counting CT: a review of pioneer studies and a glimpse into the future. Radiology. 2023;309: e222432. https://doi.org/10.1148/radiol.222432. 120. Lantz J, Bäck S, Carlhäll C-J, et al. Impact of prosthetic mitral valve orientation on the ventricular flow field: comparison using patient-specific computational fluid dynamics. J Biomech. 2021;116:110209. https://doi.org/10.1016/j. jbiomech.2020.110209. 121. Lantz J, Gupta V, Henriksson L, et al. Intracardiac flow at 4D CT: comparison with 4D flow MRI. Radiology. 2018;289:51–58. https://doi.org/10.1148/ radiol.2018173017.
| Table 1: Overview of the different imaging modalities for prosthetic heart valve assessment. | ||||
|---|---|---|---|---|
| Echocardiography | CT | Cine fHuoroscopy | Nuclea imaging | |
| Normal functioning PHV | ||||
| Leaflet motion | ++ | +++ (mechanical PHV only) n/a | ||
| +++ | +++ (mechanical PHV only) ++ (biological PHV) | n/a | ||
| Flow and pressure gradients | n/a | n/a | ||
| Determination PHV type and size | + (mechanical vs. biological) | +++ | n/a | |
| PHV dysfunction | ||||
| +(to detect ↑ pressure +++ gradient) | n/a | |||
| Pannus | ++ (indirect by detecting leaflet motion restriction in mechanical PHV) | |||
| Thrombus | ++ | +++ | ++ (indirect by detecting leafet motion restriction in mechanical PHV) | n/a |
| Paravalvular leak | +++ | + (location and size) | + (contrast needed) | n/a |
| Patient prosthesis mismatch | ++ | n/a | n/a | |
| Endocarditis | Endocarditis | Endocarditis | Endocarditis | |
| Endocarditis in | +++ | +++ | ++ (18F-FDG PET-CT) | |
| general | ||||
| Vegetations | +++ | n/a | ||
| Valve dehiscence | +++ | ++ | +++ | n/a |
| Abscesses and my- cotic aneurysms | +++ | n/a | n/a | |
| Biological PHV dys- | ||||
| function | ||||
| HALT | ++ | +++ | n/a | n/a |
| RELM | + | ++ | n/a | n/a |
| Structural valve deterioration/BVD | ++ | +++ | n/a | +++ (18F-Fluoride PET)* |
| Leafet opening angle | |||
|---|---|---|---|
| PHVs by type | Metal content | (degrees) | Leafet closing angle (degrees) |
| ATS Open Pivot | Titanium alloy ring | 85 (AVR) | 25 (AVR) |
| Carbomedics | Titanium alloy ring | 78 (AVR and MVR) | 25 (AVR and MVR) |
| Duromedics | Cobalt-chrome alloy ring | 78 (AVR), 73 (MVR) | 20 (AVR and MVR) |
| On-X | Titanium alloy ring | 90 (AVR and MVR) | 40 (AVR and MVR) |
| Bicarbon | Titanium alloy ring | 80 (AVR and MVR) | 20 (AVR and MVR) |
| St Jude/Masters/ Regent | Nickel Aloy ring | 85 (AVR and MVR) | 30 or 25a (AVR and MVR) |
| Bjork-Shiley spherical | Cobalt-chrome alloy | 60-70# (AVR and MVR) | 0 (AVR and MVR) |
| Medtronic Hall | Titanium alloy ring | 75 (AVR and MVR) | 0 (AVR and MVR) |
| Omniscience | Titanium alloy ring | 60 (AVR) | 0 (AVR) |
| Sorin All carbon | Cobalt-chrome alloy | 60 (AVR and MVR) 0 | 0 (AVR and MVR) |
| Table 3: Comprehensive template for structured reporting of CT of PHVs. The template can be adapted to the local situa. tion by adding or leaving out specific parts. | ||
|---|---|---|
| Report of cardiac CT for PHV assessment | ||
| Patient data and clinical context | ||
| Diagnostic question | ||
| CT acquisition | ||
| Scan protocol | ||
| Contrast protocol and amount | ||
| Drug administration (incl. Dose) | ||
| Quality of the scan (incl. Specific artifacts) | ||
| PHV | ||
| Location | ||
| Type of PHV | ||
| Structural integrity | ||
| Other specific features | ||
| Dynamic PHV assessment | ||
| Opening and closing of the valve | Normal/Abnormal | |
| Mechanical valves | Opening angle(s) (incl. Ref value) | |
| Anatomic PHV assessment | Closing angle(s) (incl. Ref value) | |
| Abnormal masses around PHV | Most likely thrombus | Yes/No |
| Most likely pannus | Yes/No | |
| Yes/No | ||
| PHV angulation | Indeterminate If Yes: Location Normal/Abnormal If abnormal: Describe Presence of vegetations Presence of mycotic aneurysm If Yes: Location | |
| Indeterminate If Yes: Location Normal/Abnormal If abnormal: Describe Presence of vegetations Presence of mycotic aneurysm If Yes: Location | ||
| Signs of endocarditis | Yes/No | |
| Yes/No | ||
| Other such as: | ||
| Abscess (periprosthetic fuid? Encapsulated? | ||
| Tissue enhancement?) |
| Report of cardiac CT for PHV assessment | ||
|---|---|---|
| Paravalvular leakage | Yes/No | |
| If Yes: Location | ||
| LVOT narowing | Yes/No | |
| If Yes: Location | ||
| Specific for biological PHV | Systolic valve leaflet opening orifice area | |
| PHV degeneration | Yes/no | |
| Presence of HALT | ||
| If Yes: Grading | Yes/No Mild/Moderate/Severe, or as a | |
| Presence of hypo-attenuation affecting motion If Yes: Grading | Mild/Moderate/Severe, or as a numerical value of 1-12 Yes/No Mild/Moderate/Severe | |
| Other cardiac findings | ||
| Atria and ventricles | ||
| Coronary arteries | ||
| Prior cardiac surgery/bypass grafts | None/Mild/Moderate/Severe | |
| Proximal aorta | Dimension Calcifications | None/Mild/Moderate/Severe |
| Relevant extra-cardiac fndings | None/Mild/Moderate/Severe |
| Appendix 1. Relationships with Industry of Writing Group Members for Cardiac Computed Tomography for Prosthetic HeartValveAssessment. | |||||||
|---|---|---|---|---|---|---|---|
| Formal represen- Writing Group tative of external organization | Consultant /Speakers Honoraria Bureau | Grants and re- search support | |||||
| Employment | Bayer; Sie- | Patents Royalties | |||||
| Ricardo PJ Budde | Erasmus Medical Center, Rotter- dam | mens | |||||
| (chair) | |||||||
| Marguerite E. Faure | |||||||
| terdam Center, Rotterdam | Center, Rot- terdam Center, | ||||||
| Suhny Abbara | Mayo Clinic, Jack- sonville, USA | Elsevier | |||||
| Appendix 1 (continues) |
| Formal epresen- Writing Group tative of external Consultant / Speakers Stock and Grants and re- Member organization Employment Honoraria Bureau stock options search support Patents Royalties |
|---|
| Hatem Alkadhi University Hospital Siemens Bayer, Canon, Zurich, Switzer- Health- Guerbet, land ineers Siemens Healthineers |
| Paul C. Cremer Northwestern Uni- versity, Chicago, USA |
| Gudrun M. Medical Universi- Feuchtner ty, Innsbruck, Austria |
| Holly M. American College Sanger Heart and Abbott Gonzales of Cardiology Vascular at (ACC) Atrium Health, Charlotte, USA |
| Todd L. Kiefer Duke University, Durham, USA |
| Jonathan University of Circle CVI, Philips Circle CVI, Abbott, Boston Leipsic British Colum- Heartlow Heartflow Scientific, bia, Vancouver, Edwards, GE Canada Healthcare, Medtronic |
| Koen Nieman Stanford University, Artrya; Novartis, Sie- Stanford, USA Cleerly; mens Novartis; RapidAI |
| Jonathan North American New York Univer- Revels Society for sity Langone Cardiovascular Health, New Imaging (NA- York, USA SCI) |
| Dee Dee Wang Society for Henry Ford Healthn Abbott, Cardiovascular System, Detroit, Boston Angiography and USA Scientific; Interventions Edwards (SCAI) Life- Sciences, Materi- alise |
| Eric William- Radiological Mayo Clinic, Roch- son Society of ester, USA North American (RSNA) |
| Moritz C. Society of ThoracicTexas Health Wyler von Surgeons (STS) Harris Methodist Ballmoos Hospital, Fort Worth, USA |
| Brittany A. Duke University Zwischen- berger |
| Rodrigo Sal- European Society Antwerp University gado of Cadiovascu- Hospital, Ant- lar Radiology werp, Belgium (ESCR) |
| Reviewer | Formal representative of external organi- zation | Employment | Consultant/ honoraria | Speakers S bureau | Stock and | Grants and re- | |
|---|---|---|---|---|---|---|---|
| Ashraf Hamdan | |||||||
| Rabin Medical Center, Tel Aviv | Nanox | ||||||
| Hashrul Rashid | Monash University | Brigham and Wom- en's Hospital, Harvard Medical | |||||
| Rhanderson N. Cardoso | American Collge of Cardiology (ACC) | Cleerly | |||||
| Jean-Nicolas Dacher | European Society of Cardiovas- cular Radiology (ESCR) | University of Rouen Normandy, France | Sanofi; Bayer; Pfzer | Takeda | |||
| Laura Jimenez- Juan | ciety for Cardio- vascular Imaging (NASCI) | North American So- St Michael's Hos- pital | |||||
| Cristina Fuss | Radiological Society of North America (RSNA) | Yale New Haven Hospital | Elsevier; Siemens | Elsevier | |||
| Triston Smith | Society for Cardiovascular Angiography Interventions (SCAI) | Trinity Health System | Edwards Life- sciences | ||||
| Karen M. Kim | Society of Thoracic Surgeons (STS) | UT Health Austin/ Dell Medical School |