2025 EHRA/APHRS/CHRS/HRS/LAHRS临床共识声明:心脏植入式电子设备的升级与降级

共识 心血管内科· 心脏康复与其他 · 2025

Graphical Abstract

![Cardiac Implantable Electronic Device Upgrades and DowngradesUpgrades include adding ICD capability and/or delivering a distinct pacing strategyPatient's clinical stateand requirements canDowngrades are typically considered when ICD therapy is no longer appropriate.change over time~nMArrhythmia risk ?LBBB ?RV pacing (>20%)?LV Ejection Fraction√≤35%Echocardiography for LVEFIdentifying potentiaAssess utility oupgrade patients(√)36-40%RV pacing(NT-Pro BNPavoidance algorithmCan be used to rationalize imaging if required*>10% decrease from baselineEvaluate forMulti-disciplinary team approach for decision-making, involving where needed:Alternative causesElectrophysiologist / Device Specialist / Extraction Specialist / HF Specialist / Geriatrician / Heart Surgeon / Anesthetistfor LVEF reduction(& relevant therapy)● Patient frailty and comorbidityConsiderations indecision making● Forthcoming need for device procedure i.e elective generator replacemen● Upfront venogram and procedural planning with multi-disciplinary teamprocess● Risk / benefit of extraction vs lead abandonmen● Procedural risk relative to future harm from ongoing dyssynchrony or arrhythmia\n](1/parseimages/figurefe4a398eca0440218ebf363b1fc12c86_0_3.png) Keywords

CIED upgrade $\cdot$ Downgrade

Table of contents

|Introduction|4| |---|---| |Overview of clinical data and guidelines|4| |Summary of potential procedural risk and benefit|6| |Effective workflow and approach to practice|6| |Optimizing initial choice of cardiac implantable electronic device|6| |Screening for eligible patients|6| |Timing of device upgrades or downgrades|8| |Shared decision-making|8| |Hardware considerations during decision-making|9| |Operator experience and need for specialist skills|10| |Pre-procedural steps—importance of procedural planning and contingency planning|10| |Value of pre-procedural cardiac imaging|10| |Determining the need for upgrade-assessment of left ventricular function|10| |Assessment of tricuspid regurgitation|11| |Additional utilization of imaging|11| |Cardiac computerized tomography|11| |Cardiac magnetic resonance imaging|11| |Upfront superior vascular anatomy assessment: the venogram|11| |Enaesthetic considerations|11| |Medication considerations|11| |Medication considerations|12|

|Value of pre-procedural cardiac imaging|10| |---|---| |Determining the need for upgrade-assessment of left ventricular function|11| |Assessment of tricuspid regurgitation|11| |Additional utilization of imaging|11| |Cardiac computerized tomography|11| |Cardiac magnetic resonance imaging|11| |Upfront superior vascular anatomy assessment: the venogram|11| |Anaesthetic considerations|11| |Medication considerations|12|

|Anticoagulation and antiplatelets|12| |---|---| |Equipment and lab set-up considerations|12| |Vascular superior access considerations|14| |Vascular access issues|14| |Techniques to manage venous stenoses/obstruction|14| |Vascular superior access considerations|15| |Preparation|15| |Puncture|17| |Access through an occluded vessel|18| |Serial dilatation|18| |Balloon venoplasty|18| |Infrequently used methods|18| |Contralateral new lead implantation and tunnelling/contralateral new system implant and abandonment of original system|18| |Lead extraction to create venous access or manage redundant leads|19| |Alternative (non-superior access) delivery approaches|20| |Surgical/epicardial approach|20| |Leadless, endocardial, and other non-transvenous approaches|20| |Femoral approach for lead placement|21| |Surgical considerations|22| |Consideration 1: timing of skin incision and pocket opening in relation to vascular access for cardiac implantable electronic device lead placement|22| |Consideration 2: location of incision|23| |Consideration 3: surgical techniques; blunt dissection vs. diathermy vs. low-thermal-injury dissection device|23|

4

D. Keene et al.

|Consideration 4: capulectomy|23| |---|---| |Consideration 5: management of abandoned leads|23| |Consideration 6: infection prevention and antibiotic envelopes (Table 7)|23| |Antibiotic envelopes|24| |Patient perspective|25| |Considerations for individualized risk-benefit analysis|26| |Suggested framework|27| |Role of the multi-disciplinary team and ethical framework|27| |Risk at generator change vs. any other time|27| |Specific situations|27| |Device deactivation|28| |Distinct case scenario discussions|29| |Device downgrade considerations|29| |Downgrade of cardiac resynchronization therapy defibrillator to cardiac resynchronization therapy pacemaker|34| |Practical considerations|34| |Downgrade of implantable cardioverter defibrillator to pacemaker|34| |Practical considerations|34| |Downgrade from dual-chamber to single-chamber device|36| |When a cardiac implantable electronic device is no longer indicated|36| |Distinct patient scenarios|38| |Conclusion|38|

Introduction

Cardiac implantable electronic device (CIED) upgrade and downgrade procedures are increasingly being performed in clinical practice. Whilst the most appropriate guideline-recommended device may have been used during a patient's initial implant procedure, the requirement of patients can change over time. In these scenarios, the risks and benefits of changing the capabilities of a device need to be carefully considered prior to further device interventions.

Device upgrades frequently include the addition of cardiac resyn- chronization therapy (CRT) capabilities [biventricular pacing (BiV) or, more recently, conduction system pacing (CSP)], defibrillator capabilities, or both, or delivery of dual-chamber pacing capabilities with the addition of an atrial lead. The desire to upgrade a device can result from cardiac dysfunction due to detrimental effects of pa- cing itself or the development, discovery, or progression of another cardiac comorbidity.

Furthermore, patients are living longer, and it is becoming increasingly frequent to encounter patients in whom the initially implanted device is no longer appropriate, e.g. patients with a previous defibrillator

|Definition|Categories of advice| |---|---| |Evidence or general agreement that a given measure is clinically useful and appropriate|Advice TO DO| |Evidence or general agreement that a given measure may be clinically useful and appropriate|May be appropriate TO DO| |Evidence or general agreement that a given measure is not appropriate or harmful|Advice NOT TO DO| |No advice can be given because of lack of data or inconsistency of data. The topic is important to be addressed|Areas of uncertainty|

indication who have developed advanced frailty in whom defibrillator replacement may no longer be an appropriate or desired strategy. In such patients, clinical decision-making can be challenging, particularly in those with a continued need for ventricular pacing (namely in those with an implanted DF-4 lead).$^{1-4}$

The objective of this clinical consensus statement is to provide advice on how to optimally assess, plan, and perform device upgrades and downgrades with consideration for both pre-procedural aspects as well as procedural techniques. This includes discussion of effective and safe clinical practice and an appraisal of the toolkit of techniques that can be utilized.

The document is based upon both published evidence and expert consensus with the intention to standardize procedural approaches, improve success rates, avoid complications, and ultimately improve patient outcomes.

In this clinical consensus statement, different categories of advice and the respective definitions are presented in Table 1. Furthermore, the evidence supporting each advice has been classified in different categories based on the type, quality, and quantity of respective sources (Tables 2 and 3).

All clinical consensus statements were subject to voting and required a >70% consensus by the writing group in order to be adopted.

Overview of clinical data and guidelines

Up to ~30% of patients living with an implanted pacemaker or an implantable cardioverter defibrillator (ICD) may develop significant deterioration in left ventricular (LV) systolic function with associated adverse clinical outcomes and heart failure (HF) induced or worsened by right ventricular (RV) pacing.$^{5-9}$

The guidelines for CIED upgrade in patients with RV pacing to more physiological pacing approaches differ across societies and have changed in both level of recommendation and the defined patient cohort over the last 10 years$^{10-15}$ (Table 4). Since the 2021 European Society of Cardiology (ESC) guidelines were published, additional relevant randomized controlled trial (RCT) and meta-analysis data have also been published.

The clinical data available on device upgrades are largely based on observational datasets as well as small (predominantly crossover) trials. Much of the data focuses on cohorts with severe impairment of LV

|Type of supporting evidence|Strength of evidence| |---|---| |Published $data^{a}$|>1 high-quality RCT Meta-analysis of high-quality RCT High-quality RCT > 1 moderate quality RCT Meta-analysis of moderate quality RCT High quality, large observational studies| |Expert $opinion^{b,c}$|Strong consensus >90% of writing group (WG) supports advice Consensus >70% of WG supports advice|

RCT, randomized controlled trial; WG, writing group. ªThe reference for the published data that fulfil the criteria is indicated in the table of advice, if applicable. ᵇExpert opinion also considers: Randomized, non-randomized, observational, or registry studies with limitations of design or execution, case series, meta-analyses of such studies, and physiological or mechanistic studies in human subjects. ᶜFor areas of uncertainty strong consensus/consensus that the topic is relevant and important to be addressed by future trials.

Consensus statement on CIED upgrades and downgrades

5

|Definition and delineation of the target|Strength of evidence| |---|---| |Advised TO DO|Advised TO DO| |It is advised to, it is appropriate, it is useful|>1 high-quality RCT Meta-analysis of high-quality RCT| |It is advised to, it is appropriate, it is useful|High-quality RCT > 1 moderate quality RCT Meta-analysis of moderate quality RCT| |It is advised to, it is appropriate, it is useful|>90% of writing group agree| |It is advised to, it is appropriate, it is useful|>70% of writing group agree| |May be appropriate TO DO|| |It may be appropriate, it may be useful|High quality, large observational studies| |It may be appropriate, it may be useful|>90% of writing group agree| |Advised NOT TO DO|Advised NOT TO DO| |It is not advised, it is not appropriate, it is harmful|>90% of writing group agree| |Areas of uncertainty|Areas of uncertainty| |It is unknown, lack of data, inconsistency of data|>90% of writing group agree| |It is uncertain, lack of data, inconsistency of data|>70% of writing group agree|

||Prior guideline|Current guideline| |---|---|---| |ESC (EHRA) Guidelines|2013 Recommendation for BiV upgrade if LVEF < 35% despite OMT and significant proportion of RVP The ESC 2021 pacing guidelines highlight a knowledge gap in this area.$^{13}$ The ESC 2021 pacing guidelines may be updated prior to Budapest-CRT publications and recent meta-analyses of available data No comment is made on upgrade to conduction system pacing devices in the guideline, however, upgrade considerations are included in the 2025 ESC/EHRA consensus on CSP indications$^{4}$|2021 Recommendation for BiV upgrade if LVEF < 35% despite OMT and significant proportion of RVP The ESC 2021 pacing guidelines may be updated prior to Budapest-CRT publications and recent meta-analyses of available data No comment is made on upgrade to conduction system pacing devices in the guideline, however, upgrade considerations are included in the 2025 ESC/EHRA consensus on CSP indications$^{4}$| |HRS/LAHRS/ APHRS Guidelines|2012 No formal recommendation regarding device upgrade. However, the guideline 'recognized the ability to improve LV dysfunction in the setting of chronic RV pacing with an upgrade to a biventricular device'$^{10}$|2023 Recommendation for device upgrade to biventricular pacing. Recommendation for upgrade to CSP.$^{14}$ Relevant patients are defined as those with substantial RVP and any decline in left ventricular function or symptoms noted, suggesting that device upgrade should not necessarily be restricted to only those with severely impaired ejection fraction| |Canadian HRS Guidelines|2014 Weak Recommendation CRT may be considered for patients with chronic RV pacing or who are likely to be chronically paced, have signs or symptoms of heart failure and a LVEF ≤ 35%|(No guideline update since 2014)|

LVEF, left ventricular ejection fraction; OMT, optimal medical therapy; RVP, right ventricular pacing; BIV, biventricular; CSP, conduction system pacing

function. Data on upgrades in patients with less severe LV impairment or PICM are more limited. Meta-analyses of these data do, however, show improvements in left ventricular ejection fraction (LVEF), left ventricular end-systolic volume (LVESV), brain-natriuretic peptide (BNP), QRS duration, and quality of life.$^{16-25}$ These studies were not designed to detect differences in hospitalization or mortality.

Although LVEF and NT-proBNP are not formally validated surrogate endpoints for hard clinical outcomes, and their prognostic value can vary across patient groups, they remain amongst the most widely reported and biologically plausible indicators of response in device studies. In the absence of multiple large, event-driven randomized trials in this space, improvements in these markers (particularly when reflective

D. Keene et al.

of reverse remodelling) offer valuable insight into the mechanistic and potential clinical benefit of therapy. In the CARE-HF trial of CRT, for example, CRT was associated with significant improvements in LVEF and NT-proBNP, which paralleled reductions in mortality and HF hospitalizations.$^{26}$ Similar associations were seen in the BUDAPEST-CRT Upgrade Trial, where reverse remodelling (including a change in LVEF) occurred alongside substantial reductions in the composite of mortality and HF hospitalization.$^{27,28}$

The BUDAPEST-CRT Upgrade Trial results were published after the most recent ESC pacing guideline. This was the first, large, prospective, multi-centre, randomized trial to compare the efficacy and safety of a CRT upgrade, compared to ICD alone, in patients with HF and reduced ejection fraction (HFrEF) patients (LVEF ≤ 35%) on guideline-directed medical therapy (GDMT) with HF symptoms [New York Heart Association NYHA II-IVa (which consists of outpatients with NYHA class IV)], a previously implanted pacemaker or ICD and intermittent or permanent RV pacing (RVP ≥ 20%).$^{27,28}$ A total of 360 patients were randomly assigned to cardiac resynchronization therapy defibrilator (CRT-D) upgrade (n = 215) or an ICD (n = 145). During a median of 12.4 months, the primary outcome of HF hospitalization, all-cause mortality or <15% reduction of LV end-systolic volume occurred in 58/179 (32.4%) patients in the CRT-D arm and 101/128 (78.9%) in the ICD arm [adjusted odds ratio (OR) 0.11; 95% confidence interval (CI), 0.06-0.19; P < 0.001] with a homogenous effect across all prespecified subgroups. In the composite of HF hospitalization and all-cause mortality, CRT-D showed a substantial benefit compared to ICD (adjusted hazard ratio 0.28, 95% CI 0.17-0.46; P < 0.001). Left ventricular morphological and functional response according to echocardiography also favoured CRT-D compared to ICD. While the inclusion criteria for the trial was of Vp (ventricular pacing) >20%, the reported pacing burden was much higher (Vp 85.4 ± 21.1% for CRT-D vs. Vp 88.1 ± 18.8% in controls). Whilst a greater RV pacing percentage might be associated with greater potential benefit; trials and observational studies have frequently used an arbitrary cutoff of >20% ventricular pacing burden with associated benefit.$^{5-7,13,29}$

No RCT of upgrade to CSP has been completed, although one is in progress (PROTECT-UP, NCT06052475). Several meta-analyses of observational studies of CSP upgrade (both to His bundle and left bundle branch area pacing) suggest similar improvements in LVE, LVESV, and NYHA class compared to studies of biventricular pacing. These results are similarly observed in cohorts without severe LV impairment.$^{16,30}$ Furthermore, trials are ongoing comparing CSP with other pacing modes including biventricular pacing and RV pacing and these results will be essential in further defining the role of CSP.

Summary of potential procedural risk and benefit

Due to the complexity of CIED upgrade or downgrade procedures, individual assessment and consideration of the patients' risk-benefit profile are crucial.

Patients referred for upgrades are frequently older with a higher burden of comorbidities compared to those receiving de novo devices. This, coupled with advanced age and frailty, can affect procedural outcomes.$^{31-35}$ Procedural risks also increase with the complexity of the approach and the number of leads involved. For a given patient, individualized risk assessment is required to determine, for example, whether it is preferable to abandon a redundant lead or extract it. Patients undergoing removal of an unused or malfunctioning ICD lead might have higher rates of in-hospital complications than those with a lead abandonment strategy alone; however, an abandoned lead may, albeit at low risk, have unpredictable delayed sequelae e. interference with new leads, complications associated with abandoned leads and the potential for more complex and higher-risk extraction if needed in the future due to longer lead dwell time downstream.$^{36,37}$

One of the most severe complications is device infection. The incidence is highest after CIED replacements or upgrades when compared

to other device procedures and is associated with poor outcomes.$^{38}$ The frequency of complications is also associated with operator experience, highlighting the relevance of referral of such patients to high-volume experienced centres.$^{39}$ In registry data, complication rates in general are seen to be up to 2–3 times higher for upgrade procedures than de novo implantations.$^{40-43}$ These include the risk of lead-related intervention due to displacement, infection, pocket haematoma, pneumothorax, cardiac perforation, deep vein thrombosis, tricuspid regurgitation (TR), and pain or discomfort requiring revision.

Procedural risk should not necessarily be compared to de novo procedures in these scenarios, but rather balanced against the potential benefit to be gained by the patient by upgrading from the ongoing current pacing approach. Examples where intervening can be associated with significant benefit could include scenarios such as the development of bradycardia requiring atial pacing in a patient with a VVI ICD or the development of left bundle branch block (LBBB) and a need for CRT pacing in a patient with an initial dual-chamber ICD implant. Figure 1 highlights the considerations to be taken into account when determining whether to upgrade or not for a given patient.

Effective workflow and approach to practice

Optimizing initial choice of cardiac implantable electronic device

Following guideline-directed investigations and comprehensive patient evaluation can help ensure the most appropriate CIED is selected at the time of initial implantation. For example, in a patient under 60 years old presenting with new- onset 2:1 atrioventricular (AV) block (with an- ticipated high burden of pacing), cardiac MRI showing LVEF < 50% and a mid-septal scar pattern may suggest an underlying cardiomyopathy such as Lamin AIC disease. In such a scenario, upfront implantation of a CRT-D may be preferable to a dual-chamber pacemaker, potential- ly avoiding the need for future upgrade procedures.

Nonetheless, initial device selection often occurs in the context of evolving or incomplete clinical information. In many cases, the absence of clear disease-specific markers or definitive imaging findings limits the ability to fully risk-stratify patients at the time of initial implant. As a result even, even best-practice recommendations are followed, the progression of underlying disease may necessitate device upgrade at a later stage.

Screening for eligible patients

At all planned device follow-up visits, assessment as to whether a patient has the most appropriate device implanted should be performed. Consideration should be given to whether any changes in device programming can be used to improve cardiac function. This may involve consideration of utilizing pacing avoidance algorithms if not already in use.$^{44}$ It is worth noting, however, though that for some patients, pacing avoidance algorithms, which allow long AV delays, may be of equal harm to high burden RV pacing, and may be associated with acute deterioration of cardiac function and higher incidence of atrial fibrillation.$^{45,46}$

With regard to device upgrades, assessment may include a review of patient-reported symptoms, hospitalizations related to cardiac issues, and pacemaker-derived data, including the burden of RV pacing, presence of arrhythmias, markers of fluid accumulation (thoric-ic impedance), decline in daily patient activity, or functional status. Electrocardiogram analysis may also be useful and may reveal, for example, PR interval prolongation (which could be a sign of AV dysnchrony) or QRS duration widening (a sign of interventricular dysnchrony), which may be targeted by an upgraded device for

Consensus statement on CIED upgrades and downgrades

![|Upgrades include:| |\n|:---:|:---:|\n|Dual-chamber PPM/ICD (+ A lead)|ICD (+ICD lead/S-ICD/EV-ICD)|\n|CRT-P (+ LV lead or CSP)|CRT-D (+ICD and LV or CSP lead)|\n|Leadless pacing (+ Atrial / Ventricular leadless pacemaker)| |\n|Additional Considerations:| |\n|Patient Frailty and Comorbidity| |\n|Risk / Benefit of Extraction vs. Lead abandonment| |\n|Forcoming need for device procedure i.e. generator replacement| |\n](1/parseimages/figuree83d4ce883ba4bcdbcf1668730a11226_0_1.png) therapeutic benefit.$^{77}$ Furthermore, the presence of sinus rhythm in a patient with a single-chamber ventricular pacemaker with signs or symptoms of pacemaker syndrome may benefit from the addition of an atrial lead to deliver AV synchrony.

Any concerns raised may then be further evaluated with appropriate follow-up testing, including measures of NT-Pro-BNP and/or an echocardiogram.

Sign suggestive of atrial fibrillation or other rhythm disturbances that may be contributing to a patient's symptoms or functional decline should be taken into account and addressed by targeted medical or ab- lative strategies where appropriate. This may, in some patients, avoid the need for device revision or upgrade. When patients attend for follow-up within an approximate 12-month window prior to reaching the elective replacement indicator, it may be useful for centres to adopt a more structured approach to clinical assessment. Effective pathways (as outlined in Figure 2) should identify suitable patients (i.e. those with a RV pacing burden >20%, or those with an ICD who later develop LBBB) who would benefit from a more detailed symptom assessment and referral for additional investigations.

It is important to note here that a potential limitation of remote CIED monitoring (now widely adopted in many centres as standard of care, in particular with patients with bradycardia devices) may be that there are reduced opportunities to conduct a comprehensive clinical patient assessment.

Echocardiography is felt to be the most useful test for evaluation of cardiac function. If resources are limited, NT-Pro-BNP can be used as a gatekeeper for imaging. These test results can then inform onward decisions regarding device upgrade.

If LVEF deterioration is confirmed, secondary causes should be investigated. This process may involve clinicians from other sub-specialities; for example, HF specialists to support investigations and optimization of GDMT or interventional cardiologists to assess alternative contributors i.e. ischaemic heart disease. In some situations, it may be pertinent

to revisit the patient's initial diagnosis, for example investigating the possibility of a diagnosis of neuromuscular disease, Lamin cardiomyopathy, sodium channel disease or Anderson-Fabry disease (e.g. hyper-trophic phenotype). Assessment using MRI with late-gadolinium enhancement can help to evaluate deterioration in LVEF and rework the clinical diagnosis where needed.

In those patients who present with ventricular arrhythmias or AV block at a younger age, genetic testing (which for example, may not have been available or considered at time of initial implant, or the scope of the gene panel limited in comparison to current-day options) may reveal underlying diagnoses that could offer insight to upgrade a pacemaker rather than proceeding with a generator change only (i.e. potential risk of LV dysfunction or susceptibility to ventricular arrhythmias).$^{51-53}$ Prior to elective generator replacement, the option of device downgrades should be assessed in patients with increasing comorbidities, frailty, or non-cardiac life-limiting diagnoses, such as a new diagnosis of a terminal cancer with short life expectancy, advanced organ failure, or dementia, or a change in their goals of care for any reason. The decision to undertake a downgrade is no doubt challenging, and there is uncertainty in the optimal assessment of arrhythmic risk at different stages of underlying disease.$^{54}$ For example, in a registry of 325 CRT-D recipients, 7% of patients had defibrillator therapy for ventricular arrhythmia after a first generator replacement despite no prior therapies.$^{55}$ Patient's attitudes towards generator replacement at an advanced age and even in advanced illness can vary significantly, with one survey of 3067 participants stating that 55% would elect to proceed with continuation of therapy even if seriously unwell.$^{56}$ Other studies of patient perceptions have also suggested that many patients do not realize that ICD generator non-replacement is a management option. This underscores the need for structured and careful discussions on this topic with patients when the relevant circumstances arise, explaining both the advantages and disadvantages of each strategy.

D. Keene et al.

![Possible time points forconsidering upgrade· Arrhythmia. Patient symptomsElective replacementChange in status ·Fluid accumulation / changeindicatedsymptoms/in thoracic Impedance.Increase In diureticPatient selection(ERI)NYHA classRe-evaluate· Heart failure events.Decline In activityAssess forArrhythmia risk ?LBBB ?RV pacing (>20%)?LV Ejection fraction≤35%0Echocardiography for LVEFAssess utility o(√)36-40%RV pacing avoidance algorithm(NT-Pro BNP)Can be used to rationalize imaging if required>10% decrease from baselineEvaluate forMulti-disciplinary team approach for decision-making, involving where neededAlternative causes for LVEFElectrophysiologist / Device specialist / Extraction specialist / HF specialist / Geriatrician / Heart surgeon / Anesthetistreduction (and relevant therapyDecisionNot appropriateUpgrade appropriatePossible strategies:If required• Biventricular pacing or conduction system pacingDevice upgrade. Resynchronization and/or defibrillator capabilitiesGeneratoCRT +/- Defibrillator· Lead abandonment vs. extraction of existing RV leadreplacement onlyif addition of defibrillator considered\n](1/parseimages/figureb039458ed8b34de4af0e8cdb839acd26_0_1.png)

Timing of device upgrades or downgrades

Device upgrades or downgrades can be performed at any point during a patient's device lifetime, contingent upon the clinical situation. The decision to defer a device upgrade or downgrade to the time of a generator replacement or a later undetermined date is common (even when development or progression of HF signs and symptoms are present).$^{57,58}$ Whether this is an appropriate management approach remains unclear. The BUDAPEST-CRT RCT did, however, show a 35% rate of HF hospitalizations or death within 12 months without upgrade for patients with HF.$^{27,28}$ This consensus statement advocates that in the presence of a change in clinical status and/or deterioration in cardiac function meeting an indication for revising an existing CIED, a device intervention should not be deferred until an elective generator replacement procedure.

Factors affecting this decision process include: additional device and procedural costs, any uncertainty of the potential benefit of an upgrade for a given patient, and an increased risk of procedural complications in upgrade procedures (6.2–20.9%).$^{40}$ Careful shared decision-making is therefore required in order to evaluate whether the potential for benefit outweighs the procedural risk.$^{4,21}$

Shared decision-making

A shared decision-making process should follow an integrated, patient-centred approach (Figure 3), which focuses on individual goals of care, patient preferences, and the specific risk-benefit considerations of each available clinical management strategy. This process will most frequently be led by a cardiologist and the patient with input from the wider multi-disciplinary team.

It should be emphasized that many device upgrades occur due to a predictable evolution of the underlying disease, which cannot be accurately captured at baseline implantation. Patients who receive CIEDs should be empowered to understand that some cardiac substrates evolve over time and it may become necessary to modify the technology used in their CIED beyond their initial implant procedure. This may help to mitigate any frustration or mistrust. An open discussion about therapeutic priorities and patients' wishes about end-of-life management should also be undertaken during follow-up to make any eventual decisions about upgrade or downgrade less unexpected and distressing for the patient.$^{59}$

If patients are functionally limited primarily because of worsening HF, then intervention may be encouraged sooner rather than latter.$^{60}$ However, in those who are very elderly and/or very frail with limited functional capacity and/or a short life expectancy (par-ticularly when caused by non-cardiac issues) the benefit of an up-grade may have limited clinical impact and the incremental risk of complications related to the upgrade procedure should be carefully evaluated.

The current assessment approach relies predominately on LVEF assessment, risk stratification for susceptibility to ventricular arrhythmias, and patient symptoms. Whether additional procedural risk is offset by the physiological and symptomatic improvement that an upgrade can deliver needs to be assessed in a formal manner and is best done with a multi-disciplinary team.$^{5}$

Appropriately designed future studies remain outstanding and are required to help advise clinicians on the optimal timing and circumstances, particularly for those patients with only mild to moderate LV impairment for upgrade.

Consensus statement on CIED upgrades and downgrades

![graph TD\n A1[MDT] --> A2[Cardiac physiologist]\n A2 --> A3[Pacemaker nurse]\n A3 --> A4[Imaging/sonographer]\n A4 --> A5[ Suspected need for upgrade]\n A5 --> A6[Other factors to consider • OMT • Angiography • other aetiologies]\n A5 --> A7[Echocardiography]\n A5 --> A8[Electrophysiology]\n A5 --> A9[Patient]\n A7 --> A10[LVEF impairment • Signs of dyssynchrony • Apical rocking • Septal flash • Speckle tracking]\n A8 --> A11[• ECG morphology • RV pacing burden • Arrhythmias]\n A9 --> A10\n A9 --> A8\n A9 --> A11\n A12[Risk-benefit analysis(Fig. 1A-1B)] --> A13[Need of lead extraction]\n A13 --> A14[Anaesthesiologist]\n A13 --> A15[Cardiac surgeon]\n A13 --> A16[Infectologist]\n A13 --> A17[Geriatrician]\n A13 --> A18[CRT/ICD UPGRADE]\n](1/parseimages/figureea05c906778040e094ccc0694f4fab91_0_1.png)

Hardware considerations during decision-making

Technical considerations need to be factored into any upgrade/down-grade decision typically focusing on the number and type of existing leads. The downgrade, for example, of CRT-D to CRT-P can more simply be performed with an initial DF-1 lead in situ rather than a DF-4 lead. DF-4 leads are often favoured at initial implant as they provide an integrated single connector for the pace-sense function and high voltage components connecting to a single port in an ICD header, therefore, allowing smaller device headers with lower risk of device misconnections. However, they are less flexible when downgrade is required as they are only compatible with a DF-4 device. Conversely, DF-1 leads, which have multiple ports in the device header with a higher risk of misplacement of high-voltage connectors, can be more flexible for patients with evolving ICD indication and subsequent consideration of downgrade.

With a DF-1 lead, the pace-sense component can simply be connected to a pacemaker generator and the high energy components capped and buried. With a DF-4 lead, either a new IS-1 pacemaker lead needs to be implanted, or if this is not feasible or appropriate based

on the patient's risk profile, a replacement device with DF-4 connections may instead be required albeit at significant expense even if the defibrillator components are to be deactivated. Thus, replacement with another CRT-D generator, but turning off VT/VF detections and high-voltage therapies, is a potential approach, rather than lead revision. A range of strategies for this purpose is discussed in the later section dedicated to device downgrades.

Abandinging CIED leads can impact magnetic resonance imaging (MRI) conditionality. Abandoning a lead or a lead component means that MRI-conditionality is lost. However, the 2021 ESC pacing guide-line$^{13}$ allows MRI to be performed in patients with abandoned transvenous pacing leads, if no alternative imaging modality is available. Therefore, this imaging modality is not contra-indicated in these patients. No adverse events have been reported from a total of 9 series of 343 abandoned transvenous leads.$^{61-70}$ A further series, which included nine abandoned surgical epicardial leads, also reported no significant adverse events.$^{71}$ However, consideration still needs to be given for mixed-brand systems as well as lead tip heating and device interference. Guidance from a recent cardiac MRI consensus advocates that scanning is preferred at 1.5 T over 3 T for all devices.$^{72}$ It is worth noting, however, despite real-world demonstration of safety that the availability of MRI in many centres

10

D. Keene et al.

is currently limited for patients with abandoned leads and as such, limits the accessibility of this imaging modality.

Operator experience and need for specialist skills

Upgrading conventional pacemakers or ICDs to CRT systems (BiV or CSP) is in general perceived as 'complex'. The procedure could be prolonged in duration, associated with increased risk of infections (up to 4.6%)$^{21}$ and might require skills beyond that commonly required for standard de novo device implantation. Additional skill requirements might include venoplasty or extraction of leads and the ability to attempt multiple distinct strategies as needed to enable procedural success (ideally in a single sitting).

Complication rates are closely linked to individual and centre implantation volumes.$^{22-24}$ Data from a large national quality assurance programme for pacemakers and CRT-P show that the annual hospital implantation volume is inversely related to complication rates.$^{73}$ These data suggest that CIED procedures (particularly more complex ones) should be performed by operators and centres with sufficient procedural volume and expertise. With more than 90% consensus, it is felt an arbitrary minimum caseload of 20 upgrade procedures/year per centre is an appropriate volume for sufficient experience.

Pre-procedural steps—importance of procedural planning and contingency planning

Procedural-related complications are higher during upgrade procedures compared with de novo implants.$^{4,5,9,10}$ Procedural planning and contingency planning are therefore key to adequately appraise the risk and take the necessary steps to minimize and mitigate these risks.

A structured and individually tailored approach to procedural planning allows for the anticipation of procedural challenges, diminishing unpredictability in the procedure, and reducing procedural length and complications. This includes upfront evaluation of venous access, discussion regarding optimal and second-line strategies for lead delivery, optimization of medications including antiplatelets and anticoagulants, consideration of the number of leads crossing the tricuspid valve and the risk of lead-related TR, checks to confirm all required equipment is available and a decision as to what steps would and would not be performed during a case (thus enabling the procedure to ideally be completed in a single sitting)$^{74}$ (Figure 4).

Prior to the procedure, the team should plan the procedural steps and how far they will go, balancing net risk against net benefit. This may include setting boundaries as to what approaches would and would not be used for a given patient's procedure i.e. lead extraction or lead tunnelling from a contralateral implant site, and ensuring that appropriate adjunctive teams are available if required e.g. cardiothacic surgeons and/or anaesthetic teams.

Value of pre-procedural cardiac imaging

Determining the need for upgrade—assessment of left ventricular function

Two-dimensional echocardiography is the first imaging technique to evaluate LVEF in patients who may be candidates for an upgrade to CRT. Assessment of myocardial deformation patterns by strain imaging or the use of three-dimensional (3D) echocardiography might provide additional useful information personalizing the upgrade approach.$^{75-78}$ The presence of apical rocking, septal flash, and use of speckle tracking can provide a useful assessment as to the presence of dysconchrony.$^{79,80}$ Additionally, cardiac MRI may be useful to evaluate myocardial scar and arrhythmic risk in patients with a potential ICD indication.

![PreproceduralVenographyimagingRequired kit:(Low-thermal energy dissectionleads, sheaths, TLE and venoplastyUpgradetools etc.preproceduraInfectionplanningpreventionconsiderations- Personalization according to patientscharacteristics-Determine what procedural stepswould and would not be utilizedLead managementtricuspid regurgitation5.000assessmenMedication optimizationAvoid heparin bridging;Collaboration withmulti-disciplinary team\n](1/parseimages/figuref07424cd2dc141bca247fd3527389cc3_0_13.png) Figure 4 Pre-procedural planning considerations; these include lab set-up, infection prevention, medication optimization, collaborative working approaches, and imaging requirements, including, where possible, a venogram prior to procedure date.

Consensus statement on CIED upgrades and downgrades

Assessment of tricuspid regurgitation

Echocardiography can identify patients with significant TR. In such patients, additional leads across the tricuspid valve may worsen TR and cause right-sided HF. However, it remains unclear for patients who already have severe TR whether they will experience a deterioration in symptoms or valve function if additional leads are placed across already severely affected valves.

There is equipoise regarding whether bulkier and stiffer ICD leads may impact greater upon TR than smaller pacing leads or whether any association seen relates to lead lie, trajectory, and its location rather than size $^{81-83}$.

Where worsening of TR is of significant concern, there are potential alternative approaches that avoid additional leads across the tricuspid valve. For pacing, this could include epicardial leads, coronary sinus (CS) leads, and His pacing (on the atrial side of the tricuspid valve) and even leadless pacemakers.$^{84-86}$ Leadless pacemakers (potentially related to implant technique and location of deployment) can, however, in some cases, worsen TR from baseline.$^{87}$ For defibrillation, this could include S-ICD (Boston Scientific, USA), EV-ICD (Aurora EV-ICD™, Medtronic, USA), epicardial ICD leads or standalone ICD coils placed in alternate locations.$^{85}$ Standone coils are rarely used in current practice, however, when used, these coils can be placed in the azygos vein or CS connected to the RV port or a DF-1 ICD coupled with an antero-lateral subcutaneous 'SQ' array + epicardial/CS pacing lead (connected to IS-1 RV port) or placement of an ICD lead in the middle cardac vein. Currently, the only available products for such purposes include the 6996SQ subcutaneous coil (Medtronic, USA) and the Transvene 6937 unpolar SVC coil (Medtronic, USA).

An alternative approach is to consider the extraction of redundant leads to avoid inactive leads crossing the valve. It is unclear whether this mitigates against increased TR risk.

Additional utilization of imaging

Cardiac computerized tomography

A multi-modality approach based on 3D imaging techniques may provide useful pre-procedural information. Cardiac computed tomography angiography might offer information on venous patency prior to a planned upgrade procedure. It is unclear clinically whether this is superior for procedural planning than a peripheral venogram.

Computed tomography (CT) angiography can provide additional detailed assessment of coronary anatomy (which may prompt further coronary assessment in the context of LV deterioration if significant coronary artery disease is seen). Evaluation of coronary venous branch anatomy can also be performed to assess upfront CS anatomy; however, it requires X-ray radiation and iodinated contrast.$^{88}$

Cardiac magnetic resonance imaging

Cardiac MRI provides detailed information about scar, ventricular function, and degree of segmental ventricular contraction, but requires longer scanning times and may be limited by already abandoned leads or previous concerns regarding device incompatibility, which limits access to MRI at many centres.

Randomized trials have not systematically demonstrated that the guidance of LV lead implantation based on MRI imaging to localize myocardial scar is superior to standard implant practices; however, a recent RCT did show improvement in its secondary endpoint of LVESV reduction when MRI-guided lead position was used in CRT recipients.$^{88-91}$ The absence of a suitable CS vein over viable myocardium may discourage LV lead placement and potentially favour CSP placement or a leadless LV pacing approach.$^{92-94}$

MRI may be useful in patients with PICM, where the absence of scar may help inform arrhythmic risk. Compelling RCT data, however, is lacking, and as such, the use of MRI to determine the need for a defibrillator based on scar pattern is an area of uncertainty.$^{95,96}$

Upfront superior vascular anatomy assessment: the venogram

Peripheral venography via a peripheral vein prior to skin incision can be very useful in order to plan any upgrade procedure. This is often the preferred approach for procedural planning to characterize venous anatomy rather than CT or other imaging modalities.

It is often advantageous, where possible, to perform venography in advance of the planned procedure date in patients, particularly in whom there is a high suspicion of venous occlusion (i.e. more than 3 leads in situ or the presence of visual venous collateralization on the chest wall).

Performing the venogram in advance, and discovery of significant stenoses enables upfront discussion of available options with the patient and planning of the procedure i.e. alternative venous approaches and whether proceeding with extraction is likely to be required not. While this is ideal, it may not always be possible to arrange venography in advance of an upgrade procedure, but it should be done at the very least prior to skin incision. Contrast venography can identify subclavian or innominate vein stenosis/occlusions as well as highlight congenital abnormalities (e.g. persistent left superior vena cava$^{77}$ information that should be known before starting the procedure).$^{98}$

When attempting to gain venous access, it may be useful to use a calo-dital to guide puncture.$^{98}$ It is also worth noting that peripheral venography may overestimate the presence of occlusions and more selective dye injection via a dilator within the vein can reveal residual flow through the lesion in approximately two-thirds of cases where initial venography has suggested occlusion supporting the use of proximal venography i.e. using the axillary vein in addition to peripheral venography in such case.$^{99}$ These significant venous narrowings (including those that are near-complete) may then be crossed using hydrophilic wires and venoplasty performed as needed.$^{100}$

Ultrasound imaging is sometimes utilized to assess vein location, patency, and guide vascular access; however, this approach is limited in that it cannot demonstrate more central occlusions.$^{101,102}$ As such, it is regarded as an inferior choice to contrast venography for pre-procedural planning and should not be used for this purpose.$^{103}$

Anaesthetic considerations

Local anaesthesia (LA) with or without sedation is often preferred for device procedures, including upgrades. This approach generally has fewer cardiovascular and respiratory risks making it safer for high-risk patients. This is particularly true for patients with emphysematous lung disease, where already hyperinflated lungs can be further inflated with significantly reduced margins between the lung surface and the axillary vein giving a higher pneumothorax risk while obtaining vascular access. Recovery times are shorter, more readily enabling cost-effective day case procedures. However, patients undergoing long or complex cases may struggle to remain still and, in these scenarios, moderate sedation or general anaesthesia (GA) allows for precise procedural control without compromising patient comfort. It additionally allows complex and potentially painful procedures to be completed in a single sitting, including the use of tunneling and extraction where needed.

Anaesthetic provision should be assessed alongside all the other pre-procedural planning steps to ensure that a case can be completed as planned ideally in one sitting.$^{104}$ This may mean for a patient with a widely patent venous system that GA is not required, but for a patient with a stenosis where extraction or contralateral tunnelling may be necessary, GA should be used (or at least on standby) from the outset. In a high-risk patient where the extent of procedural complexity to facilitate venous access has been set at venoplasty alone, LA or GA are both potential options based on operator and patient preference, and ease of availability of anaesthesia. If required, a defibrillation test (DFT) may also

12

D. Keene et al.

influence the upfront choice of sedation or general anaesthesia, as many may find it preferable to perform DFTs under general anaesthesia or deep sedation.

Medication considerations

Anticoagulation and antiplatelets

Anticoagulant use is associated with a high risk of haematoma formation, as is antiplatelet therapy.

The BRUISE-CONTROL 1 study showed increased haematoma formation with heparin bridging therapy in those patients taking warfarin undergoing ICD or pacemaker implant compared to uninterrupted warfarin use.$^{105}$ The BRUISE-CONTROL 2 study did not show any difference in the risk of haematoma with continuous uninterrupted direct oral anticoagulant (DOAC) use for device procedures.$^{106}$ Limitations with this study though are the enrolment only of patients undergoing de novo procedures and the occurrence of few events.

In a sub-analysis of the WRAP-IT trial (n = 6800), multi-variable modelling identified prior CIED procedures and lead revision as

Equipment and lab set-up for device upgrades

Essential

$\bullet$ Access to general anaesthesia (either as an upfront option, or if not used upfront, should be immediately available in case of extraction and may be useful in other complex, long duration procedures or those that involve lead tunnelling)$^{111}$

Venography tools (peripheral and coronary sinus) and fluoroscopy equipment with recording capacity for venography

Hydrophilic guide catheters, hydrophilic wires, hydrophilic sheaths of sufficient length, and stiff wires

If transvenous lead extraction is to be performed, general anaesthesia, invasive monitoring, lead extraction tools,$^{56}$ and availability of immediate surgical backup$^{48}$

Desirable or may be useful

Venoplasty tools$^{52}$; tunnelling tools, micropuncture catheters, non-compliant peripheral balloons

Low-thermal energy dissection devices [for example, pulse electron avalanche knife (PEAK) PlasmaBlade™ (Medtronic, USA), or PhotonBlade (Invuity, USA)]

Extraction centres should be prepared for all possibilities including the need for femoral extraction techniques$^{53}$

EP recording system with 12-lead ECG in order to assess intracardiac electrograms and QRS width and morphology (when a conduction system pacing approach is being considered)$^{54,55}$.

Material for biventricular implant, His bundle pacing, and left bundle branch pacing; be prepared for a change in approach to or HOT/LOT-CRT (his or left bundle branch-optimized cardiac resynchronization therapy) strategy

being associated with increased haematoma risk.$^{107}$ Similarly, in the REPLACE registry, the documented rates of haematoma for device upgrades or lead revision procedures were 3.5% and 4.3%, with the highest incidence in CRT device upgrade procedures.$^{100}$ However, in a large single-centre registry of 2100 patients, device upgrade was not associated with a higher risk of haematoma formation.$^{108}$ In the WRAP-IT trial, those patients with an acute haematoma were observed to have a >11-fold risk in the development of CIED infection during 3 years of follow-up. This risk was significantly reduced by the use of an antibiotic envelope.$^{109}$

In patients where the risk of bleeding is deemed to outweigh the risk of stroke or embolic complication [e.g. submuscular pocket formation, bleeding diathesis, or transvenous lead extraction (TLE)] it is advised to hold anticoagulation for 24-48 h prior to device upgrade procedures and resume after 24-48 h. For patients on dual antiplatelet therapy (DAPT), aspirin be continued, but it is advised to hold the second agent; 3 days for ticagrelor, 5 days for diploogrel, and 7 days for prasugrel beforehand. The second antiplatelet agent can be resumed after 24-48 h. Where DAPT interruption is not feasible (such as recent stent placement) careful risk assessment and individualized planning are advised.

Procedures can frequently be performed on uninterrupted single agent low-dose aspirin.$^{110}$

Equipment and lab set-up considerations

Table 5 provides a summary of the required and desirable equipment and lab set up for device upgrades.

|General considerations|Strength| |---|---| |Advised to do|| |In patients who have received a conventional pacemaker or an ICD and who develop symptomatic HF with LVEF ≤ 35% despite OMT and who have a significant (>20%) proportion of RV pacing, upgrade to CRT is advised$^{16-19,22,27,57}$|>1 moderate quality RCT or Meta-analysis of moderate quality RCT| |In patients who have received a conventional pacemaker and who develop an ICD indication, ICD upgrade is advised$^{27}$|>1 moderate quality RCT or Meta-analysis of moderate quality RCT| |An integrated care approach is advised in pacemaker and ICD patients when considering upgrades and downgrades to ensure a patient-centred approach with patient involvement and shared decision-making|>90% of writing group agree| |When HFrEF is detected in the presence of high RV pacing burden, it is advised to discuss an upgrade strategy utilizing a multi-disciplinary team approach, including consideration of prognosis and risk benefit analysis. Discussion should include discussion of upgrade|>90% of writing group agree|

Continued

Consensus statement on CIED upgrades and downgrades

|General considerations|Strength| |---|---| |to CRT with or without ICD capability|| |It is advised that device upgrade procedures are performed by experienced operators in experienced centres (at least 20 upgrade procedures/year)|>90% of writing group agree| |May be appropriate to do|| |In patients who have received a conventional pacemaker or an ICD, and who subsequently develop symptomatic HF with either LVEF 36-40%, or a significant (~ 10%) decrease in LVEF from baseline despite OMT, and who have a significant (>20%) proportion of RV pacing, it may be appropriate to upgrade to CRT$^{16}$,$^{20}$|High quality, large observational studies| |Areas of uncertainty|| |It is uncertain whether patients who suffer from PICM alone, with no other substrate identified for LV impairment, benefit from upgrade to CRT plus a defibrillator, or rather just upgrade to CRT strategies|>90% of writing group agree|

|Patient selection and assessment for upgrade indications and generic considerations|Strength| |---|---| |Advised to do|| |At all planned device follow-up visits (by physicians/cardiac physiologists/ technicians/device nurses), assessment as to whether a patient has the most appropriate device implanted is advised. This should include review of symptoms, hospitalizations, and pacemaker-derived data review. This assessment may trigger further initial investigations with echocardiography and/or NT-Pro-BNP|>90% of writing group agree| |When patients with a ventricular pacing burden >20% are approaching ERI (<12 months), evaluation of symptoms of heart failure and left ventricular function is advised|>90% of writing group agree|

Continued

Continued

|Patient selection and assessment for upgrade indications and generic considerations|Strength| |---|---| |When LVEF deterioration is confirmed, investigation of secondary causes, including arrhythmias, coronary atherosclerosis, lack of GDMT etc. and involvement of HF specialists or invasive cardiologists as needed is advised|>90% of writing group agree| |A shared decision-making process is crucial before the upgrade procedure considering patients' preferences and taking into consideration frailty and comorbidity burden. The multi-disciplinary team may play a key role in this|>90% of writing group agree| |When lead extraction is being considered, it is advised to involve a multi-disciplinary team including the CIED specialist, extraction specialist, anaesthesiologists, cardiac surgeons, geriatricians, and infectious diseases specialists where appropriate|>90% of writing group agree| |May be appropriate to do|| |It may be appropriate to perform device upgrade at the time patient symptoms or LV function deterioration is noted and not defer till time of elective generator replacement$^{27,112}$|>1 moderate quality RCT or Meta-analysis of moderate quality RCT|

|Pre-procedural considerations|Strength| |---|---| |Advised to do|| |Before an upgrade procedure, it is advised that procedural steps and strategy are planned, including what steps (for example venous access strategies or extraction) would and would not be used for a given patient balancing net risk against net benefit|>90% of writing group agree| |Two-dimensional echocardiography is advised as the primary imaging technique to evaluate patients who may be candidates for an upgrade to CRT|>90% of writing group agree| |It is advisable to have a recent echocardiogram before replacing a device generator, to assess LVEF in those with|>90% of writing group agree|

Continued

14

D. Keene et al.

Continued

|Pre-procedural considerations|Strength| |---|---| |ventricular pacing >20% and/or with heart failure symptoms (preferably done within 12 months of the proposed generator replacement)|| |Before a generator replacement, it is advised to evaluate patient's condition for signs and symptoms of heart failure and to check the device for ventricular pacing burden as well as atrial and ventricular arrhythmias|>90% of writing group agree| |Where the risk of bleeding significantly outweighs the risk of stroke or embolic complication, it is advised to hold anticoagulation for 24-48 h prior to device upgrade procedures and resume after 24-48 h.|>90% of writing group agree| |Where risk of bleeding outweighs the risk of coronary complications (i.e. stent thrombosis), it is advised to withhold DAPT for an appropriate duration prior to procedure (3 days for ticagrelor, 5 days for clopidogrel, and 7 days for prasugrel) whilst continuing on with aspirin and resume the second antiplatelet agent after 24-48 h|>90% of writing group agree| |Venography should be performed before skin incision. where possible, it is advised to undertake this in advance of a planned upgrade procedure date to enable procedural planning|>90% of writing group agree|

Vascular superior access considerations

Vascular access issues

Venous obstruction in patients with CIEDs is not uncommon. Lead-related venous obstruction has been categorized according to the degree of maximal narrowing, ranging from mild in up to 40% of patients$^{113,114}$ to severe or total occlusion in 3–9% of patients.$^{115–117}$

Venous abnormalities that occur soon after transvenous device insertion are usually related to venous thrombosis without a coexisting stenosis.$^{118}$ However, venous thrombosis that occurs later than 1 year after implantation likely stems from a stenosis caused by organization of a previous thrombus.$^{119,120}$ The pathogenesis of venous thrombosis after implantation of a transvenous CIED is multi-factorial. Ligation of the vein used for access (cut-down approach) can cause a central extension of thrombus from the cephalic vein to proximal large veins.$^{121}$ The development of collateral circulation may cause further slowing of blood flow in the affected vein and create a prothrombotic state.$^{122}$ Endothelial trauma caused by pacemaker leads may also cause an inflammatory response of the surrounding vessel wall with subsequent scarring and subsequent fibrosis.$^{123}$ More recently, multi-level lead-related venous stenosis/occlusion (MLVSO) has been proposed as a better index of the severity of global venous obstruction (rather than the degree of

![~2/3 of stenoses occur in the peripheralcomponent of the venous tree(subclavian/distal innominate vein<1/10 of stenoses are central only(innominate/superior vena cava area~1/4 of patients with stenoses have bothperipheral and central lesionsFigure 5 Venous stenosis locations and their prevalence. The majority of venous stenoses occurs peripherally in the venous tree.114,12\n](1/parseimages/figure25509c59c77d4b56b2c1d21a5e2d43d3_0_8.png) vein narrowing at only a single point), and this has been explored in a large cohort of 3002 patients.$^{124}$ The number of leads in the heart, presence of CS leads, leads on both sides of the chest, and a previous device upgrade or downgrade with lead abandonment were the strongest predictors of significant MLVSO.

Venous stenosis/occlusion is usually asymptomatic but may become clinically important in the event of a device upgrade. The majority (~2/3) of significant stenoses or occlusions are located in the peripheral component of the venous tree (subclavian/distal innominate), while a minority are central only (innominate/superior vena cava area (~17%),$^{100}$ whilst ~25% of patients with venous stenoses involve both peripheral and central locations (Figure 5).$^{100,114,125}$ Typically, venous occlusions have associated collateral veins; however, these are typically small and tortuous and therefore cannot be used to place a new lead.

Techniques to manage venous stenoses/obstruction

Multiple techniques are available for patients with venous stenoses or obstruction and a planned CIED upgrade with pre-existing leads. Superior vascular access (i.e. axillary, subclavian, and cephalic) strategies are preferred over non-superior access (i.e. femoral) strategies, which may, however, be helpful in bailout situations. Such situations are becoming less frequently required due to adoption of more novel alternative approaches (i.e. leadless pacing).$^{86}$ The chosen strategy to overcome superior vascular issues depends on multiple factors such as operator and centre experience, patient-specific risk factors, venous anatomy/location, and size of the venous obstruction, and the patient's preference (Figure 6).

Consensus statement on CIED upgrades and downgrades

![graph TD\n A1[Patient with planned CIED upgrade complete diagnostic workup] --> A2[Superior vascular access imaging • Contrast venography* • CT / MRI]\n A2 --> A3[Ipsilateral superior vascular access patient?]\n A3 --> A4[Yes]\n A3 --> A5[No]\n A4 --> A6[Fluoroscopy- or US-guided axillary vein puncture Subclavian vein puncture Cephalic vein cutdown in selected cases.]\n A5 --> A7[Central venous stenosis or occlusion]\n A7 --> A8[Feasibility of: LLP and/or S-ICD or extravascular ICD]\n A8 --> A9[No]\n A8 --> A10[Yes]\n A9 --> A11[Stenosis Venoplasty** TLE** Medial puncture]\n A9 --> A12[Total occlusion]\n A10 --> A13[Leadless Pacemaker(LLP) SL-CRT S-ICD EV-ICD LLP+SICD]\n A12 --> A14[Risk assessment • Age • Frailty • Elevated TLE risk*** • Patient's preference RISK]\n A14 --> A15[Low risk]\n A14 --> A16[High risk]\n A15 --> A17[Transvenous lead extraction]\n A16 --> A18[Contralateral implantation(lead and/or device) • Epicardial implantation • Leadless Pacemaker(LLP) • LL-CRT Bailout: Non-superior access strategies i.e. femoral approach or inside-out central venous access(IOCVA) (Go to section: Other rarely used strategies to overcome central venous occlusion)]\n](1/parseimages/figurea6bc3c9d810047e994fd7f2a26c7203d_0_1.png) Table 6 summarizes the current vascular access options and their advantages and disadvantages, which are described in detail in the following sections.

Vascular superior access considerations Preparation

Information on how previous leads have been implanted is important. If a cephalic cut-down technique has previously been used,$^{126}$ the cephalic vein cannot be utilized again as a method of access. If axillary or subclavian puncture has been used for the original implanted leads, the cephalic

approach may be possible,$^{163}$ although fibrosis in the pocket area may make it difficult to cleanly dissect the tissues and expose the vein.

It is essential to obtain good quality ipsilateral contrast venography of the chosen vein for puncture to examine the anatomical route of the distal part as well as to assess the patency of the vein from where the original leads enter the vessel, all the way to the superior vena cava.$^{114}$ If collaterals are observed, it is often a sign of at least partial stenosis, but not necessarily total occlusion. It might be possible to pass guidewires through such a part of a vein.

If an occlusion is observed and a guidewire does not pass along the lead, it can sometimes be helpful to visualize the extent of the occluded

16

D. Keene et al.

|Vascular Access options|Advantage|Disadvantage|For whom?| |---|---|---|---| |Superior vascular access options|||| |Medial subclavian vein puncture$^{126-128}$|Ipsilateral vascular access Preservation of contralateral vascular access Relatively easy technique|Increase in risk of pneumothorax and haemothorax Risk of long-term lead damage (subclavian crush) impacting on lead performance Success rate variable depending on the occlusion (length, complete occlusion)|Favourable anatomy of the venous stenosis or occlusion (puncture distally) TLE and venoplasty not appropriate or available| |'Wire under the insulation' technique$^{129}$ (figure demonstrating technique included in reference)|Preserves vascular access Avoids repeated venous cannulation Minimally invasive|Highly challenging in patients with extensive venous thromboses or occlusion (therefore less useful in leads with longer dwell time) Success rate variable with risk of lead damage|Patients with limited vascular access Relatively short lead dwell time Patients at high risk of complication from repeat venous cannulation (i.e. those with COPD or frailty)| |Serial dilation or venous venoplasty$^{85,86}$|Ipsilateral vascular access Preservation of contralateral venous access|Procedure duration potentially increased Potential damage to remaining leads Multi-disciplinary approach occasionally needed (interventional radiology support) but mostly can be done by cardiologist alone Training/experience required|All CIED upgrades with a suitable pre-operative venogram Upgrade in patients with <4 leads on one side and <5 through the SVC| |Transvenous lead extraction (TLE)$^{130-132}$|Ipsilateral vascular access Preservation of contralateral venous access Lead management in young patients i.e. minimises total number of leads MRI conditional system|Procedure duration potentially increased Additional risks including SVC tear, haemothorax, tricuspid valve damage, tamponade, death. Collateral damage (TLE of all leads instead of only a redundant lead due to inter-lead fibrosis), possible extraction of functional leads Availability of expertise and centre capability Femoral access requirement Cardiothoracic surgery backup generally required Availability of general anaesthesia|Young age and long-term need for functional CIED therapy and desire to avoid an abandoned lead Upgrade in patients with venous occlusion + >1 dysfunctional lead or >1 functional lead to be replaced Single (non-ICD) lead for extraction, lead dwell time <10 years, no coronary sinus lead (risk of inadvertent extraction along with other leads) [lower risk] MRI conditional system required| |Contralateral approach$^{133}$ Contralateral implantation of a new system Contralateral lead implantation plus tunnelling to initial site|No complications associated with venous intervention or lead extraction Increased procedure duration Increased risk|Contralateral approach with both vascular access sides used Risk of bilateral subclavian vein occlusion Increased number of leads through the SVC (potentially higher risk of SVC syndrome) Future lead revisions are more challenging Potential risk of lead-lead abrasion or interaction Further compromise of venous access If total system extraction required secondary to infection, future device implantation options may be limited due to infection involving both right- and left-sided access sites True MRI-conditionality lost contralateral tunnelling carries risk of cutaneous lead perforation|Older patients Frail patients High-risk associated with TLE (low body-mass index (BMI) < 25 kg/m$^{2}$, severe LV dysfunction, renal dysfunction, complete venous occlusion, leads >10 y/o, high-risk leads)$^{134}$ Low total number of leads (right and left; ≤4, ≤1 ICD lead in total) Non-ICD lead (for contralateral tunnelling)|

Consensus statement on CIED upgrades and downgrades

|Vascular Access options|Advantage|Disadvantage|For whom?| |---|---|---|---| |Epicardial/surgical lead $implantation^{135-137}$|Lead can be readily placed if patient undergoing cardiac surgery for alternate reason|Thoracotomy Additional risks of surgery Additional cost Suboptimal lead performance Difficult removal or revision Non-MRI conditional system|CRT upgrade after failed endocardial lead positioning and failed CSP For patients undergoing surgery for other reasons| |Utilization of a non-transvenous approach (e.g. LLP, S-ICD, EV-ICD, WISE CRT)$^{138-156}$|No lead associated problems in LLP Risk of CIED related endocarditis is lower|Only available for selected patients depending on the planned up- or downgrade procedure Unclear future management of LLPs in patients <40 y/o (limited long-term data available at the current time) Additional cost CSP LLP not readily available at this time Only a limited option in CRT|Patients at a high risk of CIED infection (e.g. haemodialysis, prior CIED infection, immunosuppression) ATP not required (S-ICD) Bilateral venous occlusion Patients at high risk for TLE| |Other, rarely used strategies to overcome central venous occlusion (non-superior access)|||| |Inside-out central venous access (IOCVA)$^{157}$|Ipsilateral vascular access Classical pectoral placement of pacing and/ or defibrillation leads Applicable for chronic total SVC and subclavian vein occlusions|Higher potential risk of pneumothorax and haemothorax Limitations in patients with scarred or calcified subcutaneous tissue No experience with lead revision or TLE Unknown risk to existing leads Complex intervention for experienced operators|Bailout strategy for very few patients| |Femoral (or trans-iliac) approach with abdominal or thigh pulse generator $placement^{158-162}$|Pacing option in patients with central venous occlusion (predominately in the era prior to LLP)|Unstable leads Risk of lead dislodgement, especially atrial lead Additional subcutaneous lead in defibrillator systems to overcome the abdominal pulse generator location Risk of deep vein thrombosis and pulmonary embolism Increased risk of procedural complications (infection, lead fracture, intestinal/vascular injury)|Bailout strategy for very few patients, primarily for implantation of a new CIED in patients with complete central venous occlusion (particularly relevant now if LLP is not available/suitable)|

Options outlined should be discussed with the patient, taking into account, where appropriate, their views relating to procedural risks, for example when a patient desires to seek an alternative to a transvenous lead extraction.

Abbreviations: ATP, anti-tachycardia pacing; CIED, cardiac implantable electronic device; CRT, cardiac resynchronization therapy; CSP, conduction system pacing; IOCVA, inside-out central access; LLP, leadless pacing; LVEF, left ventricular ejection fraction; MRI, magnetic resonance imaging; S-ICD, subcutaneous ICD; SVC, superior vena cava; TLE, transvenous lead extraction.

section by venography proximal to the occluded site with a guiding catheter via femoral access through the right atrium and superior vena cava.

Puncture

The site of puncture and orientation of the needle can be guided by the contrast filling of the distal part of the vein or more proximally by the route of the previously implanted leads.$^{164}$ The diameter of the needle used for puncture is often 18 G but newer Micropuncture® (Cook Medical, USA) or Merit MAK™ Mini Access Kit (Merit Medical, USA) come with a thinner 21G needle, and a 4 or 5 Fr sheath with a nitinol guidewire with a straight, yet flexible, tip may be useful. No formal studies have evaluated these against standard equipment in CIED

implantation, but the approach has been suggested to be advantageous in non-cardiac venous access procedures.$^{165}$ It is important to avoid damage to the existing leads by evaluating the course of the needle as well as its interaction with the lead under fluoroscopic guidance. A thinner needle might reduce that risk, in addition to reducing the risk from inadvertent arterial puncture.

Puncture should preferentially be performed in an extra-thoracic location either in the axillary vein or the distal subclavian portion (overlying the first rib), if there are no signs of occlusion.

A medial subclavian vein puncture (where required) involves puncturing the subclavian vein closer to the sternoclavicular junction, which can bypass a stenotic/occluded area. However, this carries an increased risk of pneumothorax, arterial puncture and haematoma, rarely arterio-

18

D. Keene et al.

venous fistula, transient phrenic nerve palsy due to LA, thoracic duct injury, and potentially greater risk of future lead dysfunction. This approach should be reserved as a strategy when first-line options are not successful.$^{126,127}$

Standard de novo vascular access for implantation can be guided by ultrasound.$^{101,102,166}$ However, in upgrade procedures, it has not been systematically evaluated, and fluoroscopy frequently remains the method of choice for guiding the puncture as ultrasound cannot always visualize proximal stenoses, which may preclude successful access to the heart. Ultrasound-guided puncture can, however, guide and aid puncture into the axillary vein upstream to any stenosis at low risk to the pre-existing leads. Therefore, for some operators, a hybrid approach using venography and ultrasound has a possible role.

Access through an occluded vessel

If venography identifies a possible occlusion, a hydrophilic guidewire (for example, a Whisper wire; Abbott, USA, or the laureate wire, Merit Medical, USA, or the wire from the Merit MAKTM Mini Access Kit, Merit Medical, USA) is usually required for gaining access beyond the stenotic section.$^{99}$ It is often not possible to confirm a total occlusion (where a guidewire cannot be passed), just based on peripheral venography alone. In case of a total occlusion or near-total occlusion, wires and techniques used for crossing occluded coronary arteries can also be used to cross chronically occluded veins.$^{99,167}$ The principle is to use wires that are increasingly more aggressive and stiffer. Once a wire has passed the occluded portion and is safely in the right atrium or the inferior vena vena, a addition of microcatheters allows gradual upsizing to larger and stif-fer guidewires and finally insertion of a large enough introducer sheath.

Alternatively, a hydrophilic angled braided guide catheter (for example the KA2 guiding catheter, Merit Medical, USA) can be utilized to provide direction and support for wire passage. On occasion advancing the catheter with puffs of contrast without a wire can be more effective than blind wire advancement. Once a wire has crossed the lesion/s and is safely in the inferior vena cava, serial dilatation or venoplasty may be performed.

Serial dilatation

Serial dilatation involves progressively enlarging the vein through the vascular entry point using a series of dilators of increasing diameter. This method is employed after the initial vessel puncture and guide-wire insertion. The primary goal is to create a sufficiently large and stable access route to facilitate the introduction of the CIED leads through the gradual and controlled expansion of the vessel. Depending on the severity of stenosis and the type of procedure, a range from standard sheaths to specific dilator kits can be used. Using a super-stiff 0.035" wire e.g. Supra-Core™ (Abbott, USA) or Amplatz Extra Stiff™ (Cook Medical, USA), is usually necessary to provide support, especially when navigating an angulated route. Due to the stiffness of the larger-sized dilators, this technique is usually restricted to straight distal stenotic segments. The access created through serial dilatation is often temporary, necessitating prompt advancement of the necessary equipment. Complications of this approach very rarely include vessel perforation, excessive bleeding, and haematoma formation.

Balloon venoplasty

Stenotic veins will hinder access to the heart for pacing leads and also the manoeuvrability of guiding catheters for either BiV pacing or CSP lead placement. In such cases, venoplasty can be performed.$^{100,167,168}$ Venoplasty can be done safely and efficiently by a device implanter in

a single sitting. As most tenemos are fibrotic, they may be prone to elastic recoil after venoplasty has been performed and as such it may be preferential to perform venoplasty and the device upgrade in a single sitting. The procedure is typically performed utilizing a non-compliant balloon (typically 6–10 mm diameter by 40–60 mm length) over a stiff guidewire. In some cases, pre-dilatation with a smaller balloon-catheter is required to facilitate access for larger guidewires and thus larger balloons. Typically, balloon dilatation is performed from the RA-SVC junction all the way back to the access site in the pocket with inflations to the rated burst pressure of the balloon (for example 14 atm or 14.19 bar for the Boston Mustang 10 mm × 40 mm balloon and 12 atm or 12.16 bar for the Abbott Armada balloon).$^{100}$ Figure 7 shows a case example. Venous stenting in such patients is not a favoured approach without performing lead extraction first, due to the risk of 'jailing' the lead and the potential risk of lead damage.$^{169,170}$

Infrequently used methods

(i) PowerWire (Baylis Medical, Canada) is a specialized tool designed to facilitate vascular access, particularly in challenging cases where traditional methods are insufficient. This device utilizes radio frequency (RF) energy to traverse occluded or stenosed veins, providing a pathway for subsequent device implantation procedures. There are only limited published data regarding efficacy and safety of this approach.$^{171}$

(ii) The inside-out access approach is a minimally invasive vascular access technique. In this method, a trans-septal needle within a long sheath is advanced from the femoral vein up to the most central point of a stenosis. From here, the vein can be punctured from within, creating a controlled exit to a predefined supra- or infra-clavicular point. Following this, a wire is advanced through the trans-septal needle inside-out, which is then used as a rail to introduce a sheath on to secure central venous access. Experience of this technique being used for CIED implantation is very limited.$^{157}$

Contralateral new lead implantation and tunnelling/contralateral new system implant and abandonment of original system

Implantation of an entirely new system on the contralateral side can be problematic since the most frequent upgrade procedure is from a dual-chamber pacemaker to a CRT system, which may lead to 5 leads in the SVC with an increased risk of superior vena cava obstruction/venous thoracic outlet syndrome. If fewer leads are involved, this can more readily be considered. However, even fewer than 5 leads in some patients may increase the risk of SVC obstruction, and therefore, this decision should be individualized.

Implantation of a single lead for system upgrade on the contralateral side with tunnelling to the original side represents an alternate although less frequently adopted option (Figure 9). This approach appears safe; however, it should be noted that the data documenting the longer-term performance of tunnelled leads is limited to three registries (n = 72) over median follow-up of 2–3 years.$^{133,172}$ Utilization of the contralateral vasculature should be carefully assessed as an option in the decision-making process as doing so may compromise future options for re-implantation or revision in the event of infection, which would then affect both right- and left-sided vascular access sites.

The new contralaterally implanted lead used has to be long enough (minimum suggestions: 69 cm for His and left bundle branch area pacing, 85 cm for CSP pacing, 75 cm for right ventricular pacing, 65 cm for right atrial pacing) to allow tunnelling back to the original

Consensus statement on CIED upgrades and downgrades

![APreproceduraCollateral venous flowvenogramRetrograde fillingAntegrade filling\n](1/parseimages/figure34333c677f6f44c98bed3d100642b28a_0_1.png) ![Lesion/s crossedVascular access with micro punctureVenoplasty on stiffsystem and lesion crossed with KA2amplatz wirehydrophilic 5Fr catheter\n](1/parseimages/figuref45a226e80464526b01058bf4ef030df_0_2.png) ![CVenoplasty back to pocketTight 'apple-core' lesionovercome with venoplasty aue \n](1/parseimages/figuref596663d231c437ca4aaf1d47d0a25f4_0_3.png) ![DSuccessful CS cannulationand lead deployment\n](1/parseimages/figure38d2be8a0e604a79aa15a5ce95506471_0_4.png) implant side The tunnelling tool must have a diameter sufficient to permit an IS-1, IS-4, or DF-4 connector pin to pass through (DF-1 leads are more problematic due to the yoke). Lead implantation is performed conventionally from the contralateral side. Then, a blunt-tipped tunnelling device (e.g. equipment used for ventriculoperito-neal shunts or the tunnelling tools used for either subcutaneous or extravascular ICD implantation) is gently pushed subcutaneously, suprasternal from the contralateral site to the initial device pocket. The lead can then be passed or pulled through to the initial device side (Figure 8).

Contralateral tunnelling is easiest with thin leads, e.g. for conduction system (Medtronic SelectSecure 4.1F) or CS pacing (5 Fr) but can also safely be performed in thin patients with 7 Fr pacing or 8-9 Fr ICD leads.

Lead extraction to create venous access or manage redundant leads

Lead extraction can be used to facilitate vascular access where previously implanted leads and venous stenoses do not readily permit the addition of new leads. Lead extraction is a potential option during an upgrade procedure to remove leads that would otherwise be abandoned. This reduces the total number of leads left in the venous system and in the heart. In the scenario of challenging vascular access on the ipsilateral side, lead extraction to gain vascular access has a recommendation in the 2017 HRS expert consensus statement.$^{134}$ Alternatives to lead extraction in this scenario include implanting additional leads on the contraltal side with subcutaneous tunnelling to the ipsilateral device or the implantation of a new system on the contraltal side with abandonment of ipsilateral leads. These non-extraction strategies may be suboptimal when considering lifetime lead management, particularly in younger patients with a long life expectancy.

In the scenario where extraction is not specifically required to facilitate vascular access, extraction is a potential option to remove redundant leads. Equally, these leads can be abandoned. Both strategies are considered valid with a recommendation in the 2017 HRS expert consensus statement.$^{134,173}$ In these scenarios, the management strategy should be individualized through a shared decision-making process. There is no difference in published 10-year survival probability between both lead management approaches. However, it has been shown that lead abandonment may result in more complicated future lead extraction procedures if eventually required.$^{174}$ This could impact ~10% of patients with abandoned leads.$^{175}$ Nonetheless, lead extraction in this setting is associated with an increased risk of acute procedural complications, including mortality.$^{36}$ These recommendations stem from observational datasets, and as such determining the optimal strategy is currently uncertain.

There are two significant considerations with TLE during upgrade procedures:

(1) The risk of collateral lead damage to other leads and (2) the potentially higher risk of peri-procedural complications.

In a retrospective single-centre study, out of a total cohort of 425 patients, 88 patients underwent lead extraction procedures for non-infectious indications.$^{176}$ Of these 88 patients, 36 had venous occlusions and 49 had non-occluded veins (3 with unknown status). Patients with venous occlusion had higher odds for minor peri-procedural complications (12/36 patients compared with 2/49). However, major complications, including peri-procedural death, were not significantly higher in patients with venous occlusion (2/36 and 0/49 for major complication, P = 0.17 and 2/36 and 1/49 for peri-procedural mortality, P = 0.38). The risk of collateral lead damage ranges between 2.7% and 3.8% in a general lead extraction population.$^{177,178}$ However, in the specific situation of TLE for recanalization,

20

D. Keene et al.

![D.99\n](1/parseimages/figure9323c6d60c3a4696a17f4983193da03e_0_2.png) ![C\n](1/parseimages/figurec9934fe4e5de4617b3720aa7bd45a758_0_3.png) ![D\n](1/parseimages/figure8d37bab5850f488c8e4ac3462a66e7a1_0_4.png) in cases of venous occlusion, the risk of collateral lead damage may be higher. In a retrospective single-centre study with 45 patients and 107 leads, 77 leads were targeted for extraction to re-canalize a venous occlusion with 30 bystander leads. Collateral lead damage was reported in 2 out of 30 leads.$^{179}$

In contrast, results from the Multicenter Swiss Lead Extraction Registry have shown that TLE procedures during device upgrade can be performed effectively and do not add a disproportionate risk to the upgrade procedure.$^{100,167}$ In the BUDAPEST-CRT trial, which showed significant benefit from upgrade, 15% of the CRT-D procedures and 11% of the ICD procedures required extraction.

Alternative (non-superior access) delivery approaches

Surgical/epicardial approach

In cases of complex venous anatomy, missing venous access, prosthetic tricuspid valve, or recurring infections, surgical implantation of epicardial leads or even complete epicardial CIED systems is an option.$^{180,181}$ In most cases, epicardial lead placement may and should be achieved by minimally invasive approaches avoiding full sternotomy.$^{150,181}$ Surgical approaches to upgrade can also be assessed and adopted concomitant-ly at the time of planned cardiac surgery.

Nowadays, implantation of complete epicardial systems remains a rare necessity due to the emergence of techniques detailed above. However, in the context of CIED upgrades, the minimal-invasive implantation of LV epicardial leads is a potential option that may be encountered in cases of failed vascular access or failed transvenous implantation of a CS lead. It has been shown that the use of an epicardial left ventricular lead results in equal improvement of LVEF and NYHA class in CRT.$^{151}$,$^{152}$ Furthermore, the electrical performance of epicardial leads has been shown to be non-inferior to transvenous leads in a series of more than 1000 leads over a 5-year period.$^{151}$ It should be noted though that it can be a challenge when utilizing a minimally invasive approach to ensure placement of the epicardial lead in a postero-lateral position when aiming to deliver CRT. Optimization of approaches may be able to improve outcomes.$^{138}$

Leadless, endocardial, and other non-transvenous approaches

Femoral/abdominal, surgical/epicardial approaches are associated with additional ongoing morbidity as well as surgical risk; however, leadless pacemakers offer a lower-risk alternative approach for patients who have superior access issues and/or increased risk of infection, lead fracture, tricuspid valve or pocket complications.

Until recently, only single-chamber ventricular pacing systems were available to provide ventricular pacing support (Micra®, Medtronic, USA).$^{139}$,$^{140}$ Recent advancements have expanded leadless pacemaker technology to patients who would benefit from AV synchrony (AV Micra®, Medtronic, USA) and dual-chamber AVEIR® (Abbott, USA),$^{140}$ and also for those who need atrial pacing (e.g. sinus node dys-function) (atrial or dual-chamber AVEIR®, Abbott). Implant-to-implant communication between an atrial and a ventricular leadless pacemaker has been shown to be reliable. Dual-chamber leadless pacing provides AV synchrony for patients with heart block and sinus rhythm, rather than just best serving patients with persistent AF.$^{142}$ Still, the long-term experience with these systems is yet to be demonstrated.

For patients with pacemakers in situ (transvenous, epicardial, or leadless) who develop a need for tachyarrhythmia support but have vascular access problems or an ongoing desire to avoid placing transvenous devices, extravascular ICDs (S-ICD and EV-ICD) can provide an alternative upgrade option. This scenario can lead to the addition of a separate device to the initial pacing device.

The subcutaneous ICD (S-ICD, Boston Scientific, USA)$^{143}$ has been shown to be non-inferior to the transvenous ICD with respect to device-related complications and inappropriate shocks in ICD patients without an indication for pacing.$^{144}$ Utilization of the S-ICD (Boston

Consensus statement on CIED upgrades and downgrades

Scientific, USA) in pacing-dependent patients treated with a pacemaker has been explored and has been shown to be an effective strategy.$^{145}$ However, patients require two separate systems, which can pose device interaction issues. Although extremely unlikely, the key priority is to ensure that the S-ICD does not misinterpret pacing spikes or oversense T waves as a tachyarrhythmia and thus may require careful programming of both devices.

Combination of an S-ICD (Boston Scientific, USA) with a leadless pacemaker may also facilitate the ability to deliver commanded anti-tachycardia pacing (ATP) using unidirectional wireless communication. The Modular ATP® system (Boston Scientific, USA) has been shown to have 61.3% successful VT termination.$^{146}$ This technology is currently not yet approved by regulatory agencies. The S-ICD has also been shown to be feasible in combination with an epicardial CRT-D.$^{147}$

An alternative substernal extravascular ICD (Aurora EV-ICD™, Medtronic, USA) is now available. There is minimal experience using this strategy for upgrading from a pacemaker at present.$^{148}$ The EV-ICD (Aurora EV-ICD™, Medtronic, USA) can provide pause-prevention pacing. ATP, and defibrillation energy similar to that of transvenous ICDs.$^{149}$ ATP success with this device has been estimated at 77%.$^{153,154,182}$ Pacing capture thresholds are relatively high (4.9 ± 2.0 V at implant, 5.5 ± 2.0 V at 6 months; capture at ≥ 10 volts in 49 (16.4%) patients), and patient discomfort during stimulation has limited the use of pacing in some patients. It is worth noting though that the device is not designed primarily to provide frequent anti-bradycardia pacing.

Although the experience is limited, case reports exist where a patient develops a pacing indication leading to upgrade of an EV-ICD by adding in a dual-chamber pacemaker with good effect.$^{148}$

For patients who require CRT and have vascular access problems, there is ongoing investigation with wireless technology that can provide LV endocardial stimulation synchronized with concomitant RV pacing using ultrasound-activated stimulation WiSE-CRT® (EBR Systems, USA) (Leadless-CRT).$^{155}$ Initial experience with this technology shows

favourable clinical responses in HF symptoms and LV reverse remodeling.$^{156}$ Totally leadless CRT was delivered successfully with a combination of Micra® (Medtronic, USA) and WiSE-CRT® (EBR Systems, USA) in eight patients, and this application may potentially broaden the field of leadless cardiac pacing.$^{183}$

Femoral approach for lead placement

The ilio-femoral approach for pacemaker implantation has been described since the 1980s$^{160}$ and is an option in case of upper venous access issues. The largest series to date included just 27 patients with a mean follow-up of 3 years.$^{158}$ Venous access has been described either above or below the inguinal ligament. The former avoids lead crush and damage due to flexion of the hip, but exposes the patient to intestinal lesions and peritoneal bleedings, whereas the latter is simpler and probably safer as it is routinely performed using ultrasonography. It is necessary to use long leads (e.g. the Medtronic 5076 85 or 110 cm leads, or 3830 69 or 74 cm leads, or

![BACD\n](1/parseimages/figurefc5714465ca14abc925ceca8b4861b9c_0_9.png) ![0\n](1/parseimages/figure56cfa834e0ab4eef8187bb65c6428251_0_11.png) 22

D. Keene et al.

Abbott Tendril 2088TC 100 cm lead). Coronary sinus leads for CRT may also be implanted, but require customizing the delivery sheath (see Figure 9). The generator pocket can be placed subcutaneously in the inferior abdominal wall or in the inner thigh (which is simpler and is usually well tolerated—see Figure 10), with fixation of the generator to the muscular plane to avoid migration.

The most frequently reported complication is atrial lead dislodge-ment, which is described in up to 20% of patients.$^{158}$ In order to reduce

|Vascular access considerations|Strength| |---|---| |Advised to do|| |Superior (cephalic, subclavian, and axillary) vascular access strategies are advised over non-superior access strategies (i.e. femoral transvenous implantation)|>90% of writing group agree| |Good quality contrast venography from the ipsilateral arm is advised prior to skin incision to assess the patency of the vein to help guide the procedure|>90% of writing group agree| |Venous puncture is advised in an extra-thoracic location rather than using more medial punctures to avoid risk of pneumothorax and potential future lead failures$^{127}$|High quality, large observational studies| |May be appropriate to do|| |When possible, venography may be useful to be performed in advance of the planned upgrade date to help risk-stratify and plan an upgrade approach|>90% of writing group agree| |It may be appropriate that centres undertaking device upgrades have access to operators with skills to perform multiple techniques to cross stenoses, including venoplasty|>90% of writing group agree| |Venoplasty, where possible, may be appropriate to facilitate venous access and enable device upgrade in case of venous obstruction|>90% of writing group agree| |Lead extraction may be appropriate to obtain venous access and enable device upgrade in case of venous obstruction|>90% of writing group agree| |Where superior venous access is challenging, alternative approaches may be appropriate, including S-ICD/EV-ICD, leadless pacing, and epicardial/surgical systems|>90% of writing group agree| |Contralateral lead implantation and tunnelling may be an appropriate approach to add leads|>90% of writing group agree| |Advised not to do|| |Placement of more than 5 leads in the SVC and 4 from a single subclavian vein is not advised agree|>90% of writing group agree|

|Vascular access considerations|Strength| |---|---| |Ultrasound is not advised as a first choice/ optimal technique for assessment of venous patency to plan procedural vasculature access|>90% of writing group agree| |Areas of uncertainty|| |It is uncertain in which patients extraction of abandoned leads should be utilized when extraction is not required to facilitate vascular access|>90% of writing group agree|

tensions on the lead, adequate slack with a loop in the atrium may be formed.$^{184}$

Deep vein thrombosis/occlusion has not been described to date. With the advent of leadless pacing, this approach is likely now to be rarely used, other than potentially for CRT delivery.

Surgical considerations

Although the benefits of upgrading may exceed the risks, optimal sur- gical approaches can reduce the associated risk rate.

Complications include bleeding, haematoma formation (which can prompt infection), lead damage, infection, and scarring, and these may be mitigated with optimized surgical technique and tools.

Consideration 1: timing of skin incision and pocket opening in relation to vascular access for cardiac implantable electronic device lead placement

A number of approaches exist in clinical practice. In scenarios where a generator replacement (ERI) is not required and the proposed procedure is for new lead placement, some operators will aim to gain vascular access and puncture through the skin rather than making a skin incision. If vascular access is achieved, an appropriate incision is then made either pre- or post-lead deployment. If unsuccessful in gaining vascular access, the case can be abandoned at this juncture without having made an incision. The risk of infection from this approach is unknown as whilst no incision has been made, skin flora may still be introduced to the tissues and bloodstream via the puncture needle.

An alternative approach involves performing the skin incision upfront and to free the existing generator from its pocket prior to attempting vascular access. This approach means that once the additional lead is implanted, there are no additional surgical steps of the procedure to complete, and then an efficient process can follow to secure the lead and connect to a generator and place in a pre-formed pocket. This approach allows any inadvertent pocket bleeding to manifest early and be addressed well in advance of wound closure and potentially reduces the risk of lead displacement during subsequent pocket manipulation and intervention. Furthermore, removing the generator from the pocket allows better visualization of the puncture site using fluoroscopy. However, the generator and leads are unnecessarily disturbed if vascular access/upgrade is not successful in this scenario, and furthermore, exposed to potential contamination for a longer period.

In cases where a generator replacement will be carried out (ERI), one might consider to first make the skin incision, and then free the

Consensus statement on CIED upgrades and downgrades

existing generator and modify the pocket as required before seeking vascular access. Alternatively, one might think to obtain vascular access and lead placement percutaneously prior to skin incision as this may reduce the length of time the generator pocket is open. Data determining whether one strategy is more favourable than another is lacking.

Consideration 2: location of incision

Whilst for de novo implants there are three incision types (deltopecto- toral, horizontal, and oblique), the incision for a device upgrade (which involves addition of a new lead on the ipsilateral side) should be opti- mized for venous access. The location of the initial incision and scar and additionally the location of the generator should be a secondary consideration. This is particularly relevant if the initial implant utilized a deltopecto-toral incision and the cephalic vein and/or if the generator has since displaced caudally. It is often worthwhile to screen the loca- tion of the incision using an instrument held over the skin to ensure the location of the incision is in fact optimal for gaining vascular access. Furthermore, if lead extraction is being considered, the incision may need to be adjusted to allow coaxial alignment of lead extraction tools.

Consideration 3: surgical techniques; blunt dissection vs. diathermy vs.

low-thermal-injury dissection device

Traditional diathermy or blunt dissection is associated with greater risks of complications due to excessive thermal injury (to tissues and/or leads) or less precise dissection, particularly close to the existing leads.

While the use of diathermy has been maintained in some centres, many centres have migrated to the use of low-temperature electrocautery devices, such as pulse electron avalanche knife (PEAK) PlasmaBlade™ (Medtronic, USA) or PhotonBlade (Invuity, USA). These single-use, low-temperature surgical instruments are powered by a specialized radiofrequency electrosurgical generator (AEX™, Medtronic, USA) or a standard electrocautery generator, respectively. The instrument uses brief (~40 µs) pulses to induce electrical plasma along the edge of a 12.5-µm-thin insulated electrode, allowing it to operate at low temperatures in the range 40–170°C. With standard cautery (diathermy), the high temperatures are in the range of 200–350°C.$^{185,186}$

Kypta et al. retrospectively compared two groups: 509 patients with scissors and conventional diathermy with 102 patients who had PlasmaBlade™ (Medtronic, USA). They found shorter procedure duration and less major complications, including infection and lead damage (2.4% vs. 6.9%), leading also to demonstration of cost reduction.$^{186,187}$ In a sub-study of WRAP-IT, PlasmaBlade™ (Medtronic, USA) was associated with a 23% reduction in procedure or lead-related adverse events during 3 years of follow-up, and when controlling for complexity of the case the risk reduction improved further with a 32% lower risk of complications (HR 0.68; 95% CI 0.52-0.89, P = 0.004).$^{187}$ While no high-quality results from RCTs exist on this topic, these data from observational studies and post-hoc analysis of data from an RCT suggest PlasmaBlade™ (Medtronic, USA) may be associated with fewer complications and might support its use over conventional diathermy. PhotonBlade (Invuity, USA) has been compared in bench experiments with PlasmaBlade™ (Medtronic, USA) to show less lead damage reported.$^{188}$ In a small retrospective observational study, similar outcomes were observed between the two technologies.$^{189}$

Consideration 4: capulectomy

Even in the absence of signs of clinical infections, cultures taken at the time of generator change demonstrate a significant incidence of bacterial colonization.$^{190}$ In addition, the biofilm-prone fibrous capsule inhibits the body's normal defence mechanisms and antibiotic penetration. Theoretically, 'capsuctomy' might therefore mitigate these issues. However, if capscuctomy is performed, the risk from resultant pocket bleeding and subsequent haematoma risk is significantly elevated. In a 659-patient observational study, pocket haematoma was associated with a ~7-fold increased risk of pocket infection (hazard ratio: 7.7; 95% CI: 2.9–20.5; P < 0.0001).$^{106,191}$ It is therefore advised not to perform capscuctomy as routine practice.$^{192}$ Partial capscuctomy may have a role in the presence of a highly fibrotic capcule causing suboptimal device positioning, but limited data exist addressing this.

Consideration 5: management of abandoned leads

Where the upfront decision is to abandon a pacing lead rather than proceed with lead extraction (in the scenario where lead extraction is not required for vascular access), proper management of the abandoned lead within the pocket is essential. This is to reduce infection rates (abandoned leads have previously been shown to have higher infection rates than when extraction is performed).$^{173,193,194}$ to prevent lead migration, lead erosion, interference with other components of the CIED system (leading to noise and inappropriate device behaviours), and also giving consideration to later extraction of the lead if this becomes necessary.

The abandoned lead should be coiled and placed deep within the device pocket to avoid interacting with other CIED components. Best practice often involves looping the lead within the bottom of the pocket. Some operators choose to anchor it in place with a suture to reduce its risk of migration.

Two approaches are commonly utilized when abandoning a lead. The first approach is the Cap and Bury method. Here, the lead end is covered using a specialized cap, which helps to prevent exposure of the conductive elements and reduces the risk of fluid or debris entering the lead, aiming to reduce infection risk. It is essential to ensure the cap is secured with at least one suture, ideally fixated to the muscular plane to avoid migration to the surface. This approach keeps the inner conductor clean (minimizing debris accumulation) and therefore allows the leads to be extracted readily if needed in the future. There may be an advantage of this approach if future extraction is required due to longer available lead length.

The second approach is to use the insulation pull-back approach. In this approach, the lead is cut and the insulation of the lead is pulled over the proximal end aiming to cover the exposed conductor. For greater protection and to avoid entry of fluid into the lead lumen (which may compromise future use of locking stylets for extraction), the lead end is knuckled (bent back upon itself) and firmly tied using non-resorbable suture material before being fixated to the muscular plane. This approach reduces the volume of the pocket; it is important though to try to leave sufficient lead length for potential future extraction should this need arise.$^{98,134}$

Consideration 6: infection prevention and antibiotic envelopes (Table 7)

Pre-procedural antibiotic use particularly ensuring cover against Staphylococcus aureus has been shown to reduce infection rates in all device procedures and has now become an absolute requirement.$^{38,98}$ Chlorhexidine-alcohol skin preparation has been shown to reduce

24

D. Keene et al.

|Intervention|Study design|n|Main results| |---|---|---|---| |Pre-procedural antibiotics$^{223}$|RCT|649|Trial stopped early by the safety committee due to a significant reduction in the risk of CIED-related infection (0.64% in the cefazolin group and 3.28% in the placebo group; P = 0.016)| |Chlorhexidine vs. iodine-povidone for skin $preparation^{195}$|RCT|849|Chlorhexidine resulted in a lower risk of surgical site infection, particularly superficial (4.2% vs. 8.6%, P = 0.008) and deep incisional infections (1% vs. 3%, P = 0.05) and deep incision infections in patients undergoing clean-contaminated surgeries Although not directly tested in CIED-related procedures, chlorhexidine may be the preferred choice in CIED upgrades| |Antibacterial envelope$^{203}$|RCT|6983|Compared to standard of care, the use of an antibacterial envelope resulted in a significant reduction in CIED-related infections (hazard ratio, 0.60; 95% CI, 0.36–0.98; P = 0.04)| |Adherent iodine drape$^{196}$|OBS|14 225|Use of an adherent skin drape was associated with a lower risk of CIED-related infection (OR: 0.32; 95% CI 0.154–0.665; P = 0.002) Possible differences in the baseline characteristics between groups (i.e. drape vs. no drape) are not ascertained| |Chlorhexidine vs. normal saline lavage$^{198}$|OBS|1504|Chlorhexidine lavage was associated with a significant reduction in CIED-related infection (hazard ratio 0.138; 95% CI 0.04–0.45; P = 0.001)|

surgical site infections over iodine-based products in non-CIED surgery and should therefore be a preferred skin cleansing approach for device upgrade or downgrade procedures.$^{195}$

Adherent drapes (with or without iodine coating) have been evaluated in 14 225 patients in an observational study and were observed to be associated with lower infection rates compared to no drapes (odds ratio 0.32; 95% CI 0.154–0.665; P = 0.002).$^{196}$ However, there was no separate reporting of results for the subgroup of patients with upgrades, or of efficacy of iodinated vs. non-iodinated drapes. The results of an RCT assessing the effectiveness of using a barrier-adhesive-draping on reducing the end-of-procedure pocket swab culture are pending.$^{197}$ Pocket lavage has been associated with reduced infection rates with a large observational study showing that chlorhexidine has even been associated with lower infection rates compared to normal saline lavage.$^{197,198}$ Short post-operative antibiotic regimes have failed to demonstrate incremental benefit, whereas longer regimes showed possible potential incremental benefit, but only limited supporting data are available.$^{199,200}$ An additional dose of intraprocedural antibiotics may be appropriate if the duration from the time of antibiotic prophylaxis administration plus procedure time is >240 min or if the duration of the procedure exceeds 2 half-lives of the drug.$^{201,202}$

Antibiotic envelopes

The addition of antibiotic envelopes to the current pre-operative intravenous antibiotic prophylaxis regimen can be useful in lowering the rate of pocket infection. There are currently two commercially available envelopes—TYRX™ (Medtronic, USA) and the CanGaroo® (Elutia Inc., USA).

The TYRX™ (Medtronic, USA) antibiotic envelope is an antibiotic-impregnated mesh sleeve that releases rifampicin and mino-cycline over a minimum of 7 days and is fully absorbed within 9 weeks. The Worldwide Randomized Antibiotic Envelope Infection Prevention trial (WRAP-IT) is the only randomized clinical trial of antibiotic

envelopes to date.$^{203}$ The study demonstrated a 40% reduction in major infections (HR 0.60, 95% CI 0.36–0.98), (mainly pocket infections) within 12 months. The rate of infection was, however, low (1.2% vs. 0.7% events), thus the number needed to treat is high at 200 patients, although this number may be lower in patients at higher risk, such as those undergoing device upgrade procedures. The benefit of TYRX™ (Medtronic, USA) was sustained beyond the first 12 months (mean 21 ± 8.3 months) with a reduction in CIED infections of 1.9% vs. 1.3% (HR 0.64 95% CI 0.41–0.99).$^{109}$ These results have been emphasized in a recent meta-analysis and other retrospective and observational studies, which all present a significant risk reduction in CIED reoperations with TYRX™ (Medtronic, USA).$^{204,205}$

The 1010 patient ENVELOPE RCT demonstrated no incremental benefit of antibiotic pocket irrigation and post-operative oral antibiotics beyond prophylactic measures of chlorhexidine skin preparation, pre-operative intravenous antibiotics, and an antibiotic envelope in reducing CIED infection in high-risk patients (mean PADIT score 7.4).$^{206}$

Clinical risk prediction tools such as the PADIT and BLISTER scores have been developed to guide infection prevention strategies. Both incorporate key patient-level risk factors, such as renal dysfunction, prior procedures, and device type, but differ in the inclusion of variables, like immunosuppression (BLISTER) and timing of prior procedures (PADIT).$^{199,207,208}$ However, neither score accounts for procedure duration nor complexity, such as difficulty with vascular access. These factors may also influence infection risk and inform the decision to use an antibiotic envelope and indeed in a European survey of 301 physicians, only 49% used a risk score to support their decision whether to use an antibiotic envelope or not.$^{209,210}$

The utilization of TYRX™ (Medtronic, USA) comes with an incremental cost and is currently not suggested in all CIED reoperations. However, it seems reasonable to consider the use of the antibiotic envelope in:

(1) Patients with the highest risk of CIED infection over time. Several studies have associated young age with a higher risk of infections,

Consensus statement on CIED upgrades and downgrades

25

probably in part due to multiple reoperations and a lower competing risk of dying than in the elderly population.

(2) Patients who are at the highest risk for adverse outcomes from CIED infection, such as those with CRT in whom re-implantation of an LV lead may be very challenging or technically unfeasible following extraction, individuals with prosthetic valves who are vulnerable to prosthetic valve endocarditis secondary to systemic infection, or patients in whom lead extraction carries significant morbidity.$^{32-35,134}$

(3) Patients in whom the incremental cost-effectiveness ratio supports the usage of antibiotic envelopes based on the gained quality-adjusted life years compared to the standard of care.$^{211}$

(4) Patients undergoing CRT-reinterventions particularly with previous interventions in the same pocket.$^{43,204}$

In the cost-effective analysis of the WRAP-IT study, high cost-effectiveness was demonstrated in immunocompromised patients, those with high-power devices, two or more previous procedures, as well as in revision or upgrade of low-power devices.$^{212}$ In an additional analysis of WRAP-IT, related to the context of European countries, the absorbable antibacterial envelope was associated with cost-effectiveness ratios below European benchmarks in selected patients at increased risk of infection, suggesting economic value for the European healthcare systems when used in these patients.$^{213}$

The CanGaroo® (Elutia Inc., USA) is a biologic envelope that is hydrated in a solution of gentamicin before implantation, and ensures an early peak concentration that prevails for up to a week.$^{214}$ In a pre-clinical setting, it has demonstrated promising findings in reducing bacterial burden.$^{215}$ In the 1025 patient, single-arm SECURE study (NCT 02530970; results available on clinicaltrials.gov) 1.2% had a major pocket infection over an average follow-up period of 235 days. These data are limited though due to the lack of a control group.

Compared to this extracellular matrix envelope, pockets with TYRX™ showed less inflammation, more rapid provisional matrix formation, faster absorption, and thinner capsules.$^{216}$ The extracellular matrix envelope does not seem to reduce the incidence of device skin erosion.$^{217}$ Consequently, there is not enough information available to advise using the CanGaroo® (Elutia Inc., USA) extracellular matrix envelope currently.

Taurolidine has been shown to destroy pathogens, impede surface adhesion, neutralize endotoxin and exotoxins, and promote wound healing.$^{218-221}$ It is a derivative of the non-essential amino acid taurine. Taurolidine is unstable in aqueous solution and breaks down into derivatives, which are responsible for its biological activity.$^{220}$ N-methylol groups released during the process chemically react with the amino and hydroxyl groups of susceptible molecules in the cell wall of pathogens and certain toxins, denaturing the endotoxins and polysaccharide-lipopolysaccharide components in the cell wall of pathogens thus deactivating susceptible exotoxins.$^{222}$

The use of tauroidline solution (TauroPace™ Tauropharm, Germany) as an intraoperative antimicrobial solution adjunct (AMSA) for combatting CIED infection was studied in an observational study.$^{222}$ The incidence of infection using tauroidine solution was compared to historical controls using hydrogen peroxide (H₂O₂) (1205 CIED

Table of Advice: Surgical considerations including infection risk reduction

|Surgical considerations including infection risk reduction|Strength| |---|---| |Advised to do|| |Utilization of an antibiotic envelope is advised in|High quality, large observational studies|

Table of Advice: Continued

|Surgical considerations including infection risk reduction|Strength| |---|---| |patients at higher risk of infection$^{109,203,205,212}$|| |May be appropriate to do|| |Utilization of low-temperature electrocautery techniques may be appropriate over regular diathermy and blunt dissection where available|>70% of writing group agree| |The cap & bury approach may be an appropriate management approach for abandoned leads. The insulation pull-back approach may be an appropriate management approach for abandoned leads|>90% of writing group agree >70% of writing group agree| |The insulation pull-back approach may be an appropriate management approach for abandoned leads|>90% of writing group agree| |It may be appropriate to coil and secure abandoned leads within the pocket to reduce risk of interaction with other pacing components and erosion|>90% of writing group agree| |Advised not to do|| |Capsleutomy is not routinely advised due to risk of possible resultant haematoma|>70% of writing group agree| |Areas of uncertainty|| |It is unknown whether during an upgrade procedure (without a need for generator replacement), if there is a benefit to obtaining vascular access prior to skin incision and opening the pocket, or whether alternatively, making a skin incision and opening the pocket is a preferable first procedural step|>90% of writing group agree|

procedures). Tauroline solution was observed to be associated with lower rate of acute device infection than H₂O₂ (0% vs. 1.1% respectively, P = 0.0075).$^{22}$ However, these data were observational in nature, and randomized controlled clinical trials with long follow-up are warranted to confirm these results. Until then, it cannot be advised to use tauroidine for CIED upgrade procedures.

Patient perspective

When considering a CIED upgrade from the patient's perspective, several key aspects must be addressed to ensure clarity and manage expectations.

Patients may feel overwhelmed and anxious when informed that their current device is no longer sufficient for their clinical needs and that an upgrade might be necessary. This anxiety can often be offset by explaining technological advances have occurred, facilitating these recommendations, which possibly may not have been available at the time of initial device implant.

It is crucial to discuss the potential for vascular access issues with pa-tients upfront, along with alternate solutions and their advantages and lim-itations. When preparing for the procedure, it is important to have an open dialogue about the timing of venography. Conducting venography in advance of the procedure date can offer clarity by helping delineate with greater certainty expectations and solutions, which can then be

26

D. Keene et al.

discussed with the patient. This proactive approach helps to manage expectations and ensures patients feel informed about their options. Although this upfront approach enables better planning for the medical team and clarity for the patient, logistically, it is not always an available option with many patients having their venogram only on the day of the procedure.

It is crucial to discuss with each patient available options for delivery of additional leads. It is important to consider a patient's occupation and physical activities as contralateral implantation may impact upon their dominant side, which could affect their physical activities or raise cosmetic concerns due to multiple chest wall scars.

Lead extraction (where considered as an option) should be explored in detail with the patient. Patients need to be aware of and appraise the risks and benefits associated with lead removal as opposed to lead abandonment or alternative approaches. Individual patients may have different perspectives on acceptable risk.

Considerations for individualized risk-benefit analysis

As with any clinical decision, clinicians must analyse the risk-benefit profile of the patient with respect to any proposed device upgrade or downgrade.$^{13,59,111,134}$

The risks will vary depending on multiple factors (patient-, procedural-, and system-related); these must be carefully calibrated to ensure that risks are minimized and decisions are made correctly and appropriately. These are summarized in Table 8 and Figure 11.

A generator change or end of battery life offers a key opportunity for clinicians to re-evaluate the suitability of a device and assess whether it is still aligned with a patient's current clinical requirements and prognosis as well as their individual preferences.$^{13,111,134}$

When a patient's medical needs have changed and an up- or downgrade of their device is suggested, it is important to acknowledge that the patient's personal circumstances, preferences, expectations, risk perception, etc. may have changed too. These

Table 8 Outlines the various factors to consider when evaluating the risks and benefits of a procedure

|Patient Factors|System Factors|Procedural Factors| |---|---|---| |- Age and Frailty|- Team & Centre experience|- Procedure type/complexity| |- Comorbid state|- Operator experience|- Existing and proposed device type| |(Heart Failure, Renal failure, Diabetes, Bleeding diathesis, Malignancy, Cognitive state)|- Availability of tools and kit|- Number and type of leads already implanted| |- Prior device history||- Number and type of leads to implant| |- Infection history and risk||- Techniques available to use| |- Need for device procedure e.g. ERI|||

![graph TD\n A1[LVEF] --> A2[Heart failure symptoms Decline in LVEF High burden RV pacing Arrhythmic substrate/events Number of leads in situ Anticipation of future procedures MRI conditionality Patient and/or caregiver preference]\n A2 --> A3[Assess upgrade risk/benefit]\n A3 --> A4[PPM to CSP/CRT-P OR PPM to ICD/CRT-D]\n A5[Patient at the time of generator replacement or with signs of disease progression] --> A6[Seek patient participation Explore and compare treatment options Discuss patient's values and preferences Reach a shared decision with the patient]\n A5 --> A7[Absence of ICD therapies Clinical futility of shock defibrillation Short life expectancy Major comorbidities Competing risks for mortality Arrhythmic substrate/events Quality-of-life indicators Risk of redundant leads Patient and/or caregiver preference]\n A7 --> A8[CRT-D to CRT-P ICD to PPM OR Any CIED with pacing indication]\n A8 --> A9[Assess downgrade risk/benefit]\n A9 --> A10[Age frailty Comorbidities Discussion around opportunities for deactivation at later date]\n](1/parseimages/figuref9270e0f105e4ef38770b79bd3013014_0_12.png) Consensus statement on CIED upgrades and downgrades

must be assessed again as part of the shared decision-making process.$^{224-229}$

A change of device type might be prompted by a deterioration in the patient's cardiac condition. This information alone may be overwhelming for a patient, which might preclude immediate device upgrade or downgrade discussions. Repeat consultations may therefore be required perhaps with a family member or other advocate present to progress such discussions. At all clinic appointments, signposting possible device upgrades or downgrades as a potential future management strategy may better prepare patients for such scenarios. This can help manage patient expectations.

Suggested framework

(1) Seek the patient's participation. Explain why the implanted device no longer meets the needs of the patient. Encourage the patient to play an active role in the decision to keep, up- or downgrade the device. Suggest inviting caregivers to participate in the process. (2) Help the patient explore and compare treatment options. Provide an overview of all treatment options, including refusal, in a way that can be understood by everyone. Discuss the benefits and risks of each option. Provide patient educational materials and, if possible, decision-making aids to support the consultation. Give the patient enough time to consider the options. (3) Assess the patient's values and preferences. It is important to acknowledge that as their disease progressed or improved, a patient's personal circumstances, preferences, expectations, risk perception, treatment goals, etc. may have changed as well. (4) Reach a decision together with the patient.

(4) Reach a decision together with the patient.

Role of the multi-disciplinary team and ethical framework

Shared decision-making should be at the cornerstone of this process throughout.13,224-229

In the cohort of CIED downgrade, there will be some patients unable to input into the decision-making process (due to either temporary or permanent loss of mental capacity, for example in patients with dementia). In these circumstances, these discussions should be held in conjunction with those healthcare professionals who are most familiar with the patient's care and in the absence of any advance directives, also include any nominated persons (family, other next of kin, or legal representative if available) who will be able to advocate for the patient's best interests. The process of determining what is in the best interests of the patient must consider the patient's previously expressed wishes, beliefs, and values and should seek views from those who knew the patient best, to understand their priorities, where possible. If loss of capacity is temporary, the decision should be delayed, where possible, until capacity has recovered. Any best interests process should also be carried out in accordance with national guidance.

Situations where referral to a court of law or an Ethics committee are outlined in the 'EHRA consensus statement of management of CIED in patients nearing end of life or requesting withdrawal of therapy', and may be equally applicable in circumstances where a downgrade of a device is considered.$^{227}$

In many countries, withdrawal of pacing therapy (for bradycardia indications) in pacemaker-dependent patients is contrary to the law, however, even where this is permissible from a legal standpoint, it is uncommon in practice, and the impact of continued pacemaker therapy in prolonging an inevitable dying process is unknown.$^{227}$

The multi-disciplinary team should re-evaluate the indication for device therapy and any specific requirements. Repeat assessment of a patient's heart rhythm, need for CRT, or reliance on pacing should be assessed. Complete assessment may even require a short period of device reprogramming to a passive or backup only mode to enable information gathering i.e. whether a patient deteriorates rapidly without

delivery of CRT therapy when considering device downgrade or discontinuation of therapy. The multi-disciplinary team should also evaluate whether programming changes (i.e. utilization of pacing avoidance algorithms) or medication-based strategies can be better utilized. It is unknown whether HF GDMT can be used as a sole alternative to device upgrade in the scenario of genuine PICM or when a strong indication for upgrade to CRT or ICD has developed.

Risk at generator change vs. any other time

The risk of infections or pocket-related complications of recurrent CIED replacements is incrementally elevated with each procedure.$^{43}$ Early intervention prior to ERI of the original device may result in a patient having an additional device procedure in their lifetime with a potential subsequently elevated risk. However, cardiac resynchronization (if high burden of RV pacing/broad LBB and LV impairment) or ICD (for secondary prevention of arrhythmia or if severe LV systolic dysfunction) upgrades should be performed in a timely manner to potentially avoid irreversible decline and to reduce morbidity and mortality.$^{13,111,134}$ This should not be delayed until generator replacement or a later date to improve patient outcomes.$^{27,112}$

In cases, however, where the indication is currently more nuanced (for example, patients with mid-range LVEF and only modest pacing burden), discretion should be used as this is currently an area of uncertainty and waiting till ERI might be more appropriate.

Specific situations

As already stated, risk-benefit evaluations for device upgrades and downgrades need to be individualized, with certain conditions warranting particular attention.$^{13,11,134,230}$

In patients with a degree of frailty and/or comorbidity, assessment should be tailored to the individual, but should always include: frailty, functional status, care requirements, and patient support network (par-ticularly relevant for recovery), comorbidity burden including polyphar-macy, and cognitive status, including capacity to consent for a procedure.$^{230}$

The multi-disciplinary team should seek to integrate these elements of the patient assessment, as well as the views of the patient and their advocates and those of other stakeholders, such as primary care providers and allied health professionals, to make an informed decision.

Certain conditions may need particular consideration:

Advanced frailty: By definition, a condition of impaired reserve, the physiological stress of device upgrade in someone with advanced frailty may represent an unacceptable risk to their already declining health, offsetting any potential benefit of device upgrade. Input from a geriatrican or other frailty practitioner can support the patient and the multi-disciplinary team in reaching a decision. Frailty scores may be a useful adjunct as an objective tool, particularly if serial measurements are available, but do not replace a complete clinical assessment.

Frailty may be associated with sarcopenia and/or low body-mass index (BMI), which may necessitate suturing of the device in place or intramuscular placement of the device. Sarcopenia may also cause greater than expected levels of pain related to the procedure and may increase the risk of device/lead erosion. The risk of erosion or infection can be elevated with the use of caps for abandoned leads. This strategy is, however, frequently preferred in this patient cohort in lieu of extraction. The careful placing of any abandoned leads therefore is advised.

Whilst in principle frail and older patients may be at higher risk of procedural complications, data from high-volume centres have shown that efficacy and complication rates are similar to those in younger, non-frail populations, provided additional care is taken. Therefore, referring such patients to experienced centres is crucial.$^{230}$

28

D. Keene et al.

CIED downgrade/deactivation

Considerations for:

ICD/CRT-D:

* Patient informed preferences

* Comorbidities and frailty

$\bullet$ Burden of future therapies (physical discomfort/psychological distress).

* Impact on QOL (e.g. even appropriate therapies can cause stress, pain, anxiety).

$\bullet$ Evolving nature of patient's illness i.e. sepsis, electrolyte disturbances and heart failure in terminal illness which may predpose to inappropriate shocks.

* Risk of withholding potentially life-saving therapies in future

* Appropriateness of therapies (ATP/shocks)

To be balanced against:

$\bullet$ Lack of clinical prediction regarding the need for ICD/CRT-D therapies e.g. no risk score exists

* Ethical and legal framework for withdrawing life-sustaining therapies i.e. withdrawal in pacing-dependent patients is often prohibited by the law.

$\bullet$ Need for prospective data

Any CIED with pacing therapy:

$\bullet$ Active shared decision to offer continued pacing therapy

$\bullet$ Alignment with goals of care even in frailty or palliation i.e. pacing not usually perceptible to the patient and provides symptomatic relief.

$\bullet$ Distinct approaches for those patients with prognostic pacing indication e.g. AV block to those where low pacing burden (current and/or anticipated) i.e. sinus node disease, atrial arrhythmia with bradycardia etc.

Procedural considerations

Impact of device choice on risks at generator change i.e. increased infection risk in frail patient with using large defibrillator devices at generator change

Risk of leaving redundant leads in situ at generator change (venous obstruction, infection) vs. potentially complex downgrade requiring extraction of leads in frail/comorbid patient

Figure 12 Framework for assessing CIED upgrade or downgrade decisions. ATP, anti-tachycardia pacing; QOL, quality of life; AV, atrioventricular; ICD, implantable cardioverter defibrillator; CRT, cardiac resynchronization therapy.

Long-term corticosteroid use can be associated with skin atrophy, impaired wound healing, and elevate the risk of infections. Such patients are more susceptible to bacteraemia, longer hospitalization, and higher mortality rate. If procedures can be performed without com- contamst steroid use, this is preferred.

COPD or other severe pulmonary disease may be associated with a higher risk of pneumothorax. Additionally, in those cases, when lead removal or extraction is planned, the risk of prolonged use of respiratory therapy is higher.

Patients with diabetes are at higher risk of device infection and prolonged wound healing; therefore, antibiotic envelope use may have potential benefit.$^{42}$

Patients with kidney disease are also at higher risk of device infection and prolonged wound healing; therefore, antibiotic envelope use may have use may have potential benefit.

When patients are treated with regular dialysis, venous access can be limited, and upgrades here can be associated with a 4-times higher risk of infection. Therefore, the use of leadless pacing and subcutaneous ICD is a potential option if clinically recommended.$^{13,111,134,145}$

Device deactivation

Evidence-based information on the topic of downgrade is limited. Observational data suggest reduced benefit of ICD therapy with increased age (likely due to competing risk of death from non-arrhythmic causes) and meta-analysis data suggest a non-minor procedural risk from ICD replacement (median rate of 4% for major complications).$^{40}$ Ideally, any discussions around deactivation should be part of an advanced care planning process, which should involve caregivers, primary care, palliative care, and other multi-disciplinary team members, including geriatricians and specialist nurses (pro-posed framework outlined in Figure 12). It is important that the patient is given the opportunity to articulate their anxieties, which may

Table of Advice: Individualized risk-benefit analysis

|Individualized risk-benefit analysis|Strength| |---|---| |Advised to do|| |Individual patient risk-benefit analysis before the upgrade/downgrade procedure is advised|>90% of writing group agree| |In patients considered for CIED up- or downgrade, it is advised that risks and benefits of the procedure, the patient's preferences, and goals of care are explored in a shared decision-making process|>90% of writing group agree| |It is advised that frailty, comorbidity burden, cognitive status as well as functional status, care requirements, and support network are assessed by the multi-disciplinary team with the involvement of caregivers to help guide device upgrade or downgrade decisions|>90% of writing group agree| |If a device is deactivated/made non-functional, it is advised that the procedural risks of device removal and the risks associated with leaving the device in situ are assessed individually|>90% of writing group agree| |May be appropriate to do|| |It may be appropriate to prepare patients for potential decisions relating to their device early on in their device lifetime. This may help patients prepare for decisions relating to future upgrade, downgrade, or deactivation of their CIED|>90% of writing group agree|

Consensus statement on CIED upgrades and downgrades

|Individualized risk-benefit analysis|Strength| |---|---| |Areas of uncertainty|| |It is unknown whether GDMT should be optimized first in the treatment of isolated pacing-induced cardiomyopathy (before proceeding to device upgrade) or whether earlier device upgrade should be performed potentially reducing the need for medication utilization|>90% of writing group agree|

include thoughts such as 'are you giving up on me?' and 'I was told the device is needed to keep me alive. If you switch it off, will I die?'. Exploration of these and other concerns, in a sensitive and supportive way, using clear, unambiguous, jargon-free language, is central to the patient's experience at a time when they are likely to be feeling vulnerable. It should also be noted that the option of not replacing the ICD generator is worthwhile to broach at the time of original implant where appropriate and the decision to continue offering ICD therapy should be re-evaluated at follow-up. Outside of the end-of-life scenario, up to 15% of patients would consider electing not to proceed with ICD generator replacement when they are engaged in the decision-making process.$^{226}$

Distinct case scenario discussions

This section explores commonly encountered scenarios and provides an overview of available upgrade options. Figure 13 provides a graphical summary overview of common CIED upgrade scenarios.

Table 9 outlines upgrade options and management strategies for patients with an already implanted transvenous pacemaker or ICD. In patients with a transvenous pacemaker, an upgrade may be appropriate if a patient develops a defibrillator indication. In both patients with transvenous pacemakers or ICDs, an upgrade may be appropriate to deliver additional pacing capabilities; for example an atrial lead for AV synchrony (if only a ventricular lead was initially implanted), or addition of a LV or CSP lead if a physiological pacing strategy is desired (due to either high burden of pacing or the development of an indication for resynchronization therapy for example LBBB and heart failure).

Upgrade options include implantation of additional transvenous pacing or defibrillator leads if defibrillator capability is needed. Infrequently, leads outside of the vasculature can be utilized as part of the upgrade strategy [i.e. S-ICD (Boston, Scientific, USA) for defibrulator capability or leadless or surgical lead placement for pacing needs].

Table 10 outlines upgrade options and management strategies for patients with a single-chamber leadless pacemaker. Upgrades may be appropriate to provide better atrioventricular synchrony and/or

|CIED upgrade|Strength| |---|---| |Advised to do|| |It is advised that when transvenous or leadless pacemakers are combined with subcutaneous or extravascular defibrillators, possible interactions between the systems are considered. Evaluation before and during implant should aim to reduce|>90% of writing group agree|

Continued

|CIED upgrade|Strength| |---|---| |risk of inappropriate defibrillation delivery or inappropriate inhibition of pacing|| |When a CIED with an RV pacemaker lead is upgraded to a system with a transvenous defibrillator lead or to a conduction system pacing lead, considering the pros and cons of extraction over abandonment of the original RV lead is advised|>90% of writing group agree| |It is advised when upgrading an existing pacemaker to a CSP system (no defibrillator) to avoid abandoning leads, and the already implanted RV lead can be incorporated into a CRT-P generator|>90% of writing group agree| |May be appropriate to do|| |It may be useful where a defibrillator lead is to be added, and there is a short lead dwell time of the original RV lead and/or younger age of the patient, that a lead extraction strategy is preferred over lead abandonment|>90% of writing group agree|

biventricular synchrony, which is not provided by the existing device. Upgrade may also be needed if the patient develops an ICD indication. Upgrade options include: (1) a leaded system (either a pacemaker, CRT, or defibrillator), (2) an S-ICD or EV-ICD, or (3) a dual-chamber leadless system.

Table 11 outlines upgrade management and strategies for patients with an existing S-ICD (Boston Scientific, USA) or EV-ICD (Medtronic, USA). Upgrades may be appropriate if a patient develops a pacing indication, or if it is felt the individual patient would benefit significantly from ATP, or if the patient develops a resynchronization (i.e. LBBB and severe LV impairment) indication. Upgrade options include (1) a leadless pacemaker, (2) a leaded CIED (with removal of the S-ICD/EV-ICD if a TV-ICD is implanted), or (3) switch to EV-ICD from S-ICD or vice versa.

Device downgrade considerations

The decision to downgrade from a device capable of defibrillation has several considerations. Decision-making should begin with a careful evaluation of patient fraity, comorbidity burden, life expectancy, and adequately informed personal preferences. In patients with significant competing risks from non-arrhythmic causes of death or limited functional reserve, the benefits of continued defibrillator therapy may be outweighed by its potential burdens. As emphasized in the recent EHRA consensus document,$^{146}$ shared decision-making is critical in these scenarios, ensuring that choices align with the patient's overall goals of care and values.

Beyond these core considerations, other factors may provide additional context for decision-making at the time of generator replacement. These include recovery of LVEF, the burden of ventricular arrhythmias, the presence or absence of an arrhythmogenic substrate (such as myocardial fibrosis or genetic abnormalities), and a patient's history of ICD therapies. While such elements can support clinical judgment in selected cases, they are largely informed by retrospective analyses and observational data, and no validated risk stratification tool currently exists to guide these decisions with confidence. As a result, the predictive value of these markers remains uncertain and should

30

D. Keene et al.

![Upgrade OptionsInitial Device+TV pacing+LLP+TVICD+ NON-TVICD665cePcP iAV VPyAVe A c666Single Chambe+RA VICSP lRA& CSP & Dofb le+EV.orS-ICDPV(AP&AV eony+DotbleadWISELDCSPSldyLdBivenricula+EVorS.ICDDoihled(atSP & DofbCD(a VSP66/NV0D(ARA& CSP leads(AP& AALOyRA&LVIA* Not shown in illustrationDual Chamb● Epicardial LV lead placement is not shown(AV CSP leads(Avroy)in figure but could be utilized to add an LVleadSubcutoneou● Only Lead Abandonment scenarios are shownexamples do NOT show Extraction optionsLLERV+EV orS-ICDFigure 13Overview of common upgrade scenarios\n](1/parseimages/figure882ac6cf6ece4948bffb2c00d2ba65b6_0_2.png) be regarded as adjunctive rather than definitive. The decision to continue or withdraw defibrillator therapy requires individualized, patient-centred assessment.

In a meta-analysis aiming to determine the effect of LVEF recovery following CRT implantation on the incidence of appropriate ICD therapy, Chatterjee et al.$^{258}$ reported that LVEF recovery to ≥45% had a lower estimated incidence of ICD therapy compared to those who did not recover (2.3 vs. 8.2 per 100 patient years, P < 0.001). Therefore, although the risk of appropriate ICD therapy was not nil, it was below the estimated arrhythmic risk threshold of 3 per 100 patient years for ICD benefit, leading

to a number needed to treat of 50 patients. If a blanking period for evaluating ventricular arrhythmias was applied between CRT implantation and diagnosis of LVEF recovery, the estimated arrhythmic risk was even lower at 1.7 per 100 patient years. Using an LVEF recovery cutoff of ≥35%, the estimated incidence of ICD therapy was 5.4 per 100 patient years (i.e. above the estimated threshold for benefit).

It is well established that patients with a secondary prevention indication are more likely to receive appropriate ICD therapy than those with a primary indication during the course of follow-up.$^{259}$ However, it is unknown to what extent patients with secondary

Consensus statement on CIED upgrades and downgrades

Table 9 Upgrade options and management strategies for patients with existing transvenous CIED

|Existing CIED|Upgrade|Management strategies|Key points| |---|---|---|---| |Leaded TV CIED|Single-chamber to dual-chamber PPM (upgrade AAI or VVI PPM to dual-chamber PPM)|If VVI system add a transvenous RA lead and use the existing RV lead for a DDD pacemaker If VDD system abandon the RA sensing port and add a transvenous RA lead. This may impact MRI access in some centres If AAI system add a transvenous RV lead If single-chamber PPM, dual-chamber leadless pacemaker (Aveir DR system, Abbott, USA), or Aveir single-chamber device to co-exist with leaded CIED to provide dual-chamber system or AV Micra® (Medtronic, USA) can all be utilized if vascular access issues preclude implantation of new atrial or ventricular lead|If initial lead is a VDD lead, 231-233 could consider lead extraction and replacement with a standard RV bipolar lead for example to achieve formal MRI-conditionality, but this would not be a common first-line approach in non-infected systems| |Leaded TV CIED|Upgrade from single or dual chamber PPM to CRT-P/CSP pacing strategy|Addition of either an LV lead to an RV pacing system or addition of a CSP lead to an RV pacing system Where a CSP lead is added, consideration can be given whether to incorporate the original RV lead within the new pacing approach or whether to abandon the original RV lead. If original system is VVI, it could incorporate CSP upgrade within a dual-chamber device (utilizing the atrial port) or a CRT-P device with a plugged atrial port|Upgrading to a physiological pacing strategy may be relevant in patients with pacing-induced cardiomyopathy or for those with broad LBBB If upgrade to CSP is performed, both the existing RV lead and CSP can be incorporated into the new device (typically CRT-P generator). Whilst a CRT-P costs a little more than a DR PPM, this cost is frequently not prohibitive. This may be useful rather than abandoning leads The decision to abandon leads is patient-specific and will take into account individual considerations, including the risk of extraction and the patient's vasculature In the case of pacing-induced cardiomyopathy resulting in severe LV impairment (no other cause for LV deterioration identified), the added net value of a defibrillator is not known and is an area of uncertainty. The LV function may well improve with improvement in cardiac activation from change in pacing approach. Retrospective cohort studies suggest >70% of patients with PICM will improve within 12 months to have an LVEF > 35%. 234 In patients with EF <35% and with significant comorbidity and increased frailty, upgrade to a pacing only solution may be more appropriate than upgrade to CRT-D Considerations include the need for ATP, likelihood of lead-induced or worsening of TR and its impact, need for physiological pacing, and patient factors (age, comorbidity, etc.) A key consideration in this scenario is often whether to abandon the existing RV lead or extract it. When adding two leads, the RV pacing lead can be abandoned or extracted, but in younger patients, particularly where existing leads have only been in a short duration, it may be preferable to consider extraction of the RV pacing lead prior to ICD/CRT-D upgrade to reduce the| |Leaded TV CIED|Upgrade from PPM/CRT-P to ICD or to CRT-D|Addition of a standard RV-ICD lead is the usual solution, but it will result in at least two leads traversing the tricuspid valve that could induce or worsen TR If a physiological pacing strategy is needed, addition of two leads will likely be required to a standard pacemaker: either a lead in the coronary sinus or at the conduction system plus the standard RV defibrillation lead In younger patients (particularly with pacing leads that have only been in for a short duration), it may be preferable to consider extraction of the RV pacing lead prior to ICD/CRT-D upgrade to reduce the|Considerations include the need for ATP, likelihood of lead-induced or worsening of TR and its impact, need for physiological pacing, and patient factors (age, comorbidity, etc.) A key consideration in this scenario is often whether to abandon the existing RV lead or extract it. When adding two leads, the RV pacing lead can be abandoned or extracted, but in younger patients, particularly where existing leads have only been in a short time, greater consideration should be given to extraction. 238 Dedicated ICD leads that are able to provide CSP are being trialled and may be an option to reduce|

32

D. Keene et al.

|Existing CIED|Upgrade|Management strategies|Key points| |---|---|---|---| |||number of leads in the heart as part of a lead management strategy If vascular access is challenging, an S-ICD can be used in conjunction with the existing pacing device, particularly in those who do not need ATP. This is a reasonable solution for primary prevention indications. There is evidence of efficacy and no increase in inappropriate shock.$^{235}$ EV-ICD is a possible alternative if ATP is required, but long-term data remain limited.$^{149}$ In both cases, the patients will need two devices, and MRI non-compatibility may be a concern. ICD need only: If there is patent venous access, implant RV bipolar ICD lead at alternative site within the ventricle. Subsequently, abandon or extract the existing RV lead. This will be the most common strategy. Implant a subcutaneous ICD as a co-implant if anticipated low ATP requirements$^{235}$ Add an EV-ICD with ATP potential$^{149,236,237}$ ICD + CRT (BiV or CSP) need Implant a standard CRT-D, either abandon or extract the RV pacing lead (an RV defibrillator lead and LV/ CSP lead are added)|additional lead requirements contingent on ongoing trial results. Additionally, the role of repurposing existing ICD leads for deployment at the left bundle location is being investigated too.$^{239}$ If an ICD is added without extracting the RV pacing lead, it may be worth considering whether the new ICD lead should be a dedicated bipolar lead rather than an integrated bipolar lead to avoid potential problems with sensing|

|Existing CIED|Upgrade to|Management strategies|Key points| |---|---|---|---| |Leadless pacemaker (LLP)|Leaded CIED (Pacemaker or Defibrillator)|Implant of transvenous ICD/pacemaker (single, dual, or CRT system) and deactivation of leadless pacemaker Implant of transvenous ICD/pacemaker (single, dual, or CRT system) and removal of the leadless pacemaker|Removing a leadless pacemaker when converting to a leaded system reduces the implanted material inside the heart—the benefit of this is, however, unknown. Removing LLPs poses challenges related to procedural complexity, risks, and costs. If the LLP is a Micra device (Medtronic, USA) and there is only one leadless device in the heart, it is often preferred in clinical practice to abandon the device (particularly, if longer dwell time and provided there is no evidence of LLP-related infection). Micra devices (Medtronic, USA) do not have a dedicated mechanism for retrieval, but can still be removed with a snare with high success rates.$^{240,241}$ However, should the device be an Abbott NanoStim or Aveir (Abbott, USA) leadless pacemaker, these devices have a dedicated docking button for retrieval, and studies report a ~90% retrieval success rate.$^{242,243}$ Evidence of infection would necessitate explant/extraction of the device|

Consensus statement on CIED upgrades and downgrades

Table 10 Continued

|Existing CIED|Upgrade to|Management strategies|Key points| |---|---|---|---| |Existing CIED|LLP + S-ICD or EV-ICD|Implant of S-ICD or EV-ICD system in addition to ongoing pacing provision from leadless $device^{145,244,245}$|Consider the possible interactions between LLP and ICD. Broad RV paced QRS morphology can potentially lead to double counting of signals and therefore possible inappropriate shocks Intraoperative testing for interactions (ICD waveform screening during ventricular pacing and sensing with possible lead placement revision to ensure appropriate sensing) should be taken into account. For EV-ICD, caution is required if the LLP is an atrial pacing device, as atrial pacing may lead to inappropriate sensing, which needs to be evaluated at implant if being considered. VF induction is advised in these cases to confirm satisfactory function of both devices| |Existing CIED|Dual-chamber leadless|Addition of leadless atrial device as part of Dual-Chamber AVEIR® system (Abbott, USA)$^{151}$|Upgrading with an atrial leadless pacemaker may alleviate symptoms from leadless pacemaker syndrome. Success rates for deployment reported in a 36-patient series were 91.4% with complication rates reported to be 13.9%.$^{246-248}$|

|Existing CIED|Upgrade|Management strategies|Key points| |---|---|---|---| |S-ICD (Boston Scientific, USA)|S-ICD + LLP|For pacing needs, combining an S-ICD with an LLP may be preferred, particularly if elevated risk of CIED infection. This combination is more frequently encountered in complex congenital heart disease and/or if vascular access is limited/challenging$^{244,249-252}$|During LLP deployment, S-ICD waveform screening should be performed for reasons detailed above. Consider repositioning of the LLP in case of S-ICD sensing concerns$^{244,245}$ Confirm appropriate S-ICD sensing during VOO at high pacing output.$^{244,245}$ Consider modular cardiac rhythm management (EMPOWER), when commercially available, as this will provide ATP functionality| |S-ICD + leaded CIED||For pacing indications, conventional transvenous leads are commonly used. However, if there are vascular access issues, an epicardial pacing system is a potential option. If ATP functionality is desired, a transvenous ICD lead can be implanted. In this scenario, the S-ICD should be removed If no ATP requirement but only Brady or CRT requirements, a pacemaker or CRT-P could be implanted and the S-ICD left in situ. If vascular issues are not present, it may be appropriate to implant an entire TV system and remove the S-ICD. Combining a TV CIED and an S-ICD may be used when vascular access is not sufficient for multiple lead implantation|Intraoperative S-ICD waveform screening during maximal voltage output and inhibition should be performed for reasons detailed above.$^{253}$ Set the pacemaker in VOO mode at maximum output to watch for oversensing and also for VF under-sensing during defibrillation threshold testing in patients (i.e. test the 'worst-case scenario')$^{254}$ Consider programming the pacemaker upper rate to <50% of the S-ICD tachycardia zone (avoids inappropriate therapy due to double detection) if oversensing appears to be a concern Obligate bipolar pacing output to avoid inappropriate double detection of pacing stimulus and evoked response consider pacing approaches that deliver narrower QRS morphologies to reduce double counting risk$^{255}$| ||Replace with EV-ICD|This could be an option if ATP functionality is desired, no anti-bradycardia pacing indication and there is no venous access for a TV-CIED|Technology is in relative infancy. The pivotal study (n = 356 enrolled) has reported high defibrillation success rates and low complication rates.$^{149}$ More data are needed on this technology| |EV-ICD|EV-ICD + LLP|For pacing needs combining an EV-ICD with an LLP may be preferred, particularly if elevated risk of CIED infection or vascular is limited/challenging|Technology is in relative infancy. Atrial pacing may lead to inappropriate sensing and interaction will need to be tested at implant if being considered. Experience is limited to date$^{148,256}$|

34

D. Keene et al.

|Existing CIED|Upgrade|Management strategies|Key points| |---|---|---|---| ||EV-ICD + leaded CIED|For pacing indications, conventional transvenous leads are commonly used. However, if there are vascular access issues, an epicardial pacing system is a potential option. If only Brady or CRT requirements, a pacemaker or CRT-P could be implanted, and the EV-ICD left in situ. If vascular issues are not present, it may be appropriate to implant an entire TV system and remove the EV-ICD. Combining a TV CIED and an EV-ICD may be used when vascular access is not sufficient for multiple lead implantation|Technology is in relative infancy. Atrial pacing may lead to inappropriate sensing and interaction will need to be tested at implant if being considered. Experience is limited to $date^{148,257}$| ||Replace with S-ICD|This could be an option if vascular access is limited/challenging and an alternative defibrillation strategy is desired|Maintains a non-transvenous and extravascular system|

prevention are still at risk for sudden death after having recovered their LVEF with CRT.

Despite recovery of LVEF, there may be underlying substrates that maintain patients' arrhythmic risk, such as genetic abnormalities (e.g. laminopathy$^{260}$) or substrate detected by MRI, such as grey-zone myocardial fibrosis in ischaemic heart disease (an admixture of fibrosis and viable tissue),$^{261}$ or mid-wall fibrosis in non-ischaemic cardiomyopathy.$^{95}$

Downgrade of cardiac resynchronization therapy defibrillator to cardiac resynchronization therapy pacemaker

The evidence for downgrading CRT-D to CRT-P is currently limited to two series totalling 73 patients with a primary prevention indication, no appropriate IC therapy and LVEF improvement to >45%$^{262}$ or >50%.$^{263}$ There were no significant ventricular arrhythmic events over a mean follow-up of ~4 years. Despite the arguments for down-grading from CRT-D to CRT-P, respondents to an EHRA survey reported that they performed these downgrades in <10% of patients.$^{264}$ Furthermore, we still lack data from RCTs on this topic.

Practical considerations

Unlike the DF-1 ICD connection standard, where downgrade from CRT-D to CRT-P can be easily achieved by simply abandoning the DF-1 pin(s) and connecting the IS-1 pin to the pacemaker generator, the DF-4 standard is more challenging due to the low- and high-voltage components all being integrated together on the same pin (with additional limitations, as covered elsewhere.$^{1}$ The issue could be solved by a DF-4IS-1 adaptor, which is unfortunately currently unavailable. The following options are possible though:

(1) The simplest approach involves a like-for-like generator replacement and simply deactivating the ICD capabilities of the new device at the time of generator replacement. Unfortunately, this is not a particularly cost-efficient upfront solution for these patients but has minimal additional risk.

(3) Extracting the DF-4 lead and replacing with a ventricular pacing lead. This strategy carries the inherent risks of lead extraction and the additional risk of damaging or dislodging the CS lead in case of lead-lead adhesions. The cost of extraction and its possible complications, especially in frail patients, likely outweighs the cost of replacement with a defibrillator device.

(4) Conversion of a DF-4 ICD lead to a unipolar IS-1 lead has recently been described in four patients in abstract-form only. This approach uses an Oscar M/IS-10 5 mm to IS1 adaptor and Abbott 4033A lead cap. Experience with this is limited and the long-term durability of this modified lead is unknown$^{265}$

(5) Simple lead switch on the CRT-P header in patients with a DF-4 lead and a bipolar CS lead.$^{266}$ It is possible to connect the DF-4 pin to the IS-4 port (of the CS pacing lead), but not vice versa—see Figure 14. This is a safety feature to prevent shocks being delivered via an inadvertently switched pacing lead in a DF-4/IS-4 CRT-D device. However, as the great majority of CS leads currently are of the IS-4 standard, this is rarely an option for CRT-D cases.

(6) Utilization of an Abbott CRT-P, which allows LV sensing. It is possible to cap the DF-4 component, which is not required, and simply connect an IS-4 quadripolar LV lead to the LV port, which will enable delivery of LV-only pacing. Retaining MRI-conditionality in this scenario is a labelling issue.

Downgrade of implantable cardioverter defibrillator to pacemaker

In this scenario, patients have a pacing indication (otherwise, a CIED would no longer be indicated) with either infrequent requirement for pacing or relatively preserved ejection fraction (otherwise, CRT would be preferable). Using a cutoff for LVEF of >40-45% is prudent, bearing in mind the variability of LVEF measurement (even if most indications for ICDs are LVEF <35%).$^{258}$ Programmed ventricular stimulation may be performed in selected cases (via the transvenous ICD), aiming to improve risk stratification of sudden death and continued ICD indication, although this strategy has never been tested.

Practical considerations

As described above, the DF-4 lead may be connected to the IS-4 port of a CRT-P, with plugging of the RV port. However, currently, only Abbott (Sylmar, USA) CRT-Ps are enabled with bipolar sensing from the LV channel with resetting of standard timing cycles. Boston Scientific (Marborough, USA) and Biotronik (Berlin, Germany) CRT-Ps are also capable of sensing from the LV channel, but this is only intended to prevent LV pacing in the vulnerable period after a left-sided ventricular extrasystole and does not impact right-sided timing cycles, which makes them unsuitable for downgrades in instances without an RV pacing lead. A case series of three patients with this strategy has been published.$^{267}$ MRI-conditionality is lost when the DF-4 lead is plugged into the LV port, but this configuration is unlikely to put patients at risk for MRI scans (although no data are available).

Consensus statement on CIED upgrades and downgrades

![|Lead Type|Connector Design|\n|:---:|:---:|\n|IS-4 LLLL|No contact|\n|DF-4 LLHH|High, High|\n](1/parseimages/figureb9555125d90c4d48a4f856c4b4bb3d98_0_1.png) ![graph TD\n A1[ICD/CRT-D requiring generator replacement] --> A2[Primary prevention indication? No appropriate therapy or monitored sustained ventricular arrhythmias? a LVEF ≥ 45%?]\n A2 --> A3[No]\n A2 --> A4[Yes]\n A3 --> A5[Replace ICD b]\n A4 --> A6[Additional considerations Underlying arrhythmic substrate? Comorbidities (competing risks for death)? Advanced age and/or frailty? Cognitive dysfunction/dementia? Risk of harm from ICD therapy (inappropriate shocks)? Requirement for implanting an additional pacing lead? Patient's preference?]\n A6 --> A7[Device]\n A7 --> A8[ICD]\n A7 --> A9[CRTD]\n A8 --> A10[Indication for pacing]\n A9 --> A12[Consider downgrading to CRT-P]\n A10 --> A11[No]\n A10 --> A13[Yes]\n A11 --> A14[Consider explanting generator (with or without lead extraction) or abandon CIED]\n A13 --> A15[Consider downgrading to pacemaker b]\n](1/parseimages/figure90af1ec858354e2ab636042beccc6fe8_0_2.png) 36

D. Keene et al.

A decision tree for evaluating options at ICD generator change is outlined in Figure 15 and highlights when downgrade may be a suitable management approach.

Downgrade from dual-chamber to single-chamber device

In patients with dual-chamber devices who develop permanent atrial fibrillation, the atrial lead becomes redundant. At generator change, the following options are available:

(1) Replace with a dual-chamber generator. This avoids abandoning the atrial leads, thus retaining MRI-conditionality, which may be the preferred strategy in most patients, particularly those requiring MRI.

(2) Replace with a single-chamber device and abandon the atrial lead. MRI-conditionality is thereby lost. In the same manner that the ESC pacing guidelines allow MRIs to be performed in selected pacemaker patients with abandoned leads, an EHRA consensus document similarly allows MRI in ICDs with abandoned leads.$^{268}$ Furthermore, it is unknown if abandoned atrial leads are associated with the same potential risks as abandoned ventricular leads (as induced atrial arrhythmia is not an issue in patients with permanent atrial fibrillation).

(3) Replace with a single-chamber device and extract the atrial lead. This has the advantage of avoiding any abandoned material but exposes the patient to the risks of lead extraction.

When a cardiac implantable electronic device is no longer indicated

Different options are available for patients in whom a CIED is no longer indicated:

(1) Abandon the CIED. The end-of-life behaviour of CIEDs is device-specific and may vary from erratic behaviour (with potential pro-arrhythmia) to safe inactivation. Inactivating and abandoning the CIED may be the preferred option in frail and elderly patients.$^{13}$

(2) Explant the generator and leave the lead(s) in situ. This avoids consequences of end-of-life erratic CIED behaviour but precludes MRI-conditionality and additionally carries a degree of procedural risk (namely infection) from re-intervention on the generator pocket. However, generator explant has an additional advantage in particularly frail, sarcopenic, or cachectic patients, where a redundant generator may excessively impact upon skin integrity

|Patient scenario|Management strategies and options|Key points| |---|---|---| |A 63-year-old man with ischaemic cardiomyopathy and a dual-chamber ICD implanted 10 years ago for primary prevention (LVEF 30%). He develops VT storm and is treated with Amiodarone, which increases his ventricular pacing percentage to >20%. He presents with worsening heart failure and a deterioration of his LVEF to 20%. He is referred for an upgrade to a CRT-D. His venogram reveals his axillary vein is occluded|Given the high prevalence of venous stenosis in patients with existing CIED leads, a peripheral venogram was performed, which confirmed axillary vein occlusion. Management strategies were discussed, including medial subclavian puncture, attempts at re-establishing vascular patency via ipsilateral venoplasty or lead extraction, vs. lead abandonment and re-implant of a new CRT-D on the contralateral side, or implantation of upgrade lead on contralateral side and tunnelling to existing pocket The patient opted to attempt re-establishing vascular patency first by venoplasty, and if that failed, lead extraction and re-implantation would have been used to facilitate completion of the upgrade procedure at the same sitting to avoid multiple procedures increasing risk of infection A successful upgrade procedure was performed after venoplasty with addition of an IS4 LV lead|A consideration in this case is his relatively young age and possibility of risks related to lead abandonment if an entire de novo system were to be implanted on the contralateral side attempts at re-establishing vascular patency with venoplasty are relatively low risk and frequently successful even when veins appear occluded. Regaining vascular access via lead extraction may have a higher risk than venoplasty, but it is an effective alternative if venoplasty fails. Lead abandonment and potential future risk of lead extraction, which might increase with higher lead burden and dwell time (for example in the scenario of if abandoned leads become involved in a later infected device)| |An 82-year-old woman with permanent atrial fibrillation, renal impairment, and diabetes. VDD pacemaker implanted 9 years ago as part of an ablate and pace strategy. Inhibition of right ventricular pacing was noted due to the oversensing of myopotentials with ipsilateral arm movements Her venogram revealed chronic ipsilateral left subclavian and left brachiocephalic vein occlusion. No evidence|Management options discussed included attempting venoplasty to gain access for a new lead, extraction and replacement of ventricular lead, abandonment and addition of a new contralateral ventricular lead and device, or abandonment and implantation of a leadless pacemaker Ipsilateral venous occlusion would require contralateral lead placement, ipsilateral vein|Key considerations were lack of evidence of PICM, preserved LV function, advanced age and patient preferences A leadless pacemaker reduces the likelihood of future CIED infections Only one abandoned pacing lead. Likelihood of future CIED interventions (the patient may not require further pacemaker replacements or interventions due to her life|

Consensus statement on CIED upgrades and downgrades

|Patient scenario|Management strategies and options|Key points| |---|---|---| |of LV impairment or PICM was apparent even with her high burden of RV pacing|venoplasty, or lead extraction and re-implant of a new lead through the occluded vein. Considerations included the patient's preference for a simpler procedure with lowest expected complications. The decision was to abandon the malfunctioning lead and to implant a VVVIR leadless pacemaker. The decision also included to remove the old pacemaker generator to avoid future risks from interactions with the new leadless pacemaker, for example, automatic programming in VOO at ERI or EOL|expectancy, thus decreasing further the possibility of CIED infection) It is important to note that the whole transvenous system could have been abandoned, potentially reducing procedural complication, but this would be more relevant if the original device could be programmed 'OOO' thus mitigating risk of device interaction| |An 81-year-old woman with non-ischaemic cardiomyopathy had an ICD implanted 13 years ago (primary prevention). Dual coil and passive fixation ICD lead. No pacing indication During follow-up and on guideline-recommended heart failure medical treatment, the LVEF increased to 50%. No ventricular arrhythmias were observed during all follow-up visits. Frequent episodes of myopotential oversensing were observed during remote monitoring|Management options discussed included abandoning lead and generator, removing generator and abandoning the ICD lead, or extracting the ICD lead and removing the generator Reappraise ICD indication (primary prevention, no ICD therapies, LVEF improvement) Considerations included the patient's preference not to have a prolonged hospitalization/ recovery period and not wanting a redundant generator in the pocket The patient did not want to undergo extraction. At her request, the decision was made to remove the generator and abandon the lead. No new ICD was implanted|Currently, no indication for ICD implantation. Considerations are: Age of the patient The type of ICD lead (dual-coil and passive fixation), as well as long dwell time, increases the risk of lead extractionAbandoned leads are only a relative contraindication for MRIThe patient remains at ongoing risk for lead infection, which, if occurred, would require a higher-risk extraction in the futureOpening the pocket to remove the generator may have exposed the patient to a risk of infectionHowever, leaving the ICD in situ could leave the patient at risk (likely low) of erratic EOL behaviourConsideration could have been given to abandoning the whole system, reducing procedural complication, and preserving MRI labelling| |A 79-year-old woman with dual-chamber PPM for intermittent complete heart block. Recent generator replacement (leads 12 years old). Recent mitral valve replacement and tricuspid valve annuloplasty for severe MR and TR. Post-operative LBBB development. TR remains moderate to severe despite intervention and in part this is due to splinting of valve by existing RV pacing lead. LVEF has reduced to 23% post-surgery|Decision regarding upgrade of device to CRT-P vs. D and decision regarding upgrade to biventricular pacing vs. CSP Despite a recent generator replacement, it was appropriate to proceed quickly to device upgrade and not defer. Given her age, comorbidity, non-ischaemic aetiology, absence of any ventricular arrhythmia history and her elicited values with shared decision-making, upgrade to CRT-P with an LV lead only using venoplasty to gain adequate venous access was chosen|To avoid additional leads traversing the tricuspid valve the decision was made to not implant an ICD lead nor for a CSP lead. In view of her low BMI, sarcopenia and frailty an ICD was not favoured to avoid a large generator. This was the favoured approach particularly as ventricular arrhythmia risk was felt to be low. Evidence of benefit of LV leads alone in patients with non-ischaemic cardiomyopathy and broad LBBB is significant At the time of cardiac surgery, epicardial LV lead implantation could have been a potential alternative|

(3) CIED extraction. This may be the preferred option in young patients as it avoids all future device-related complications but comes with the highest initial risk.

Individual risk-benefit analysis should be performed in each case with shared decision-making with the patient and caregivers. Some considerations in such risk-benefit analysis are currently unknown including for

example, the value of programmed ventricular stimulation via the ICD before generator replacement in patients with LVEF > 35%, to stratify risk of sudden death and evaluate requirement for continued ICD therapy or the risks with MRI in the setting of abandoned atrial leads (without abandoned ventricular leads), abandoned DF-1 components of high-voltage leads, DF-4 leads plugged in IS-4 ports and abandoned leads which are capped vs. those which are sectioned.

38

D. Keene et al.

|CIED downgrade|Strength| |---|---| |Advised to do|| |At the time of elective generator replacement, it is advised to undertake an evaluation of the ongoing need for defibrillator therapy. Consideration to potential device downgrade—such as from CRT-D to CRT-P, from ICD to a pacemaker, or deactivation or explant of the ICD may be appropriate. Factors to prioritize to guide the decision include patient frailty, comorbidities, life expectancy, and informed patient preference. All considerations should be integrated into a shared decision-making process tailored to the patient's overall clinical context and goals of care$^{262,263}$|>90% of writing group agree| |When pacing is no longer indicated, it is advised that the management strategy is based on an individual risk—benefit analysis in a shared decision-making process together with the patient and caregivers|>90% of writing group agree| |May be appropriate to do|| |At generator change in a patient with a dual-chamber pacemaker or ICD who has developed permanent atrial fibrillation, replacement with a dual-chamber device to retain MRI-conditionality may be appropriate (rather than utilizing a single-chamber device and abandoning the atrial lead)|>90% of writing group agree| |Areas of Uncertainty|| |In patients with primary prevention ICDs who have not previously had device therapies, the optimal approach to determine the arrhythmic risk at the time of generator replacement is uncertain. This risk stratification might inform who can be safely downgraded.|>70% of writing group agree| |The safety of downgrading CRT-D to CRT-P in patients with secondary prevention indications with recovered LVEF (>40-45%), without appropriate ICD therapy, remains uncertain|>90% of writing group agree| |The prognostic significance of ventricular arrhythmias occurring only early after CRT-D implantation, before recovery of LVEF has resulted (which may constitute a blanking period) remains uncertain|>90% of writing group agree|

be accounted for when determining device upgrade or downgrade suitability, including, importantly, patient frailty. We provide an overview of available tools and techniques that can be used to safely and efficiently perform procedures, with a particular focus on how to achieve vascular access with discussion around the relative risks and benefits of lead abandonment and extraction approaches. We finally provide a multi-disciplinary framework for device downgrades that can be utilized in clinical practice. Knowledge gaps remain for definitive assessment as to whether any of the procedural techniques lead to superior patient outcomes over another technique. Furthermore, randomized evidence is required to help better refine which patient groups benefit most from the armamentarium of device upgrade options.

Acknowledgements

The authors thank the EHRA Scientific Document Committee: Prof. Katja Zeppenfeld (Chairperson), Prof. Jens Cosedis Nielsen (Co-Chairperson), Dr Maria Frausing, Dr Estelle Gandjbakh, Prof. Isabelle C. van Gelder, Dr Georgios Kollias, Prof. Michael Kühne, Prof. Radoslaw Lenarczyk, Assoc. Prof. Petr Peichl, Dr Archana Rao, Dr Avi Sabbag, Prof. Frederic Sacher, Dr Michelle Samuel, Dr Hadrian Wijnmaalen.

Conflict of interest: D.K. receives honoraria and research support from Medtronic and Abbott. J.C.N. is an executive editor for Europace. H.B. reports institutional fellowship and research support, speaker's bureau, and advisory boards from Abbott, Biotronik, Boston Scientific, Medtronic, and Microport. J.-C.D. speaker honoraria or travel grants from Medtronic, Boston Scientific, Abbott, Micropoty and Phillips. J.E.I. Abbott: consulting, steering committee, speaking honoraria, Biotronik: consulting, Boston Scientific: data safety monitoring committee, speaking honoraria, Medtronic: Consulting. CWI Advisory board: Medtronic, Honoraria for presentations and travel: Abbott, Biotronik, Boston Scientific, Medtronic, Microport, S.L. Board of the British Geriatrics Society, Cardiovascular Group. J.MT. medical advisory committees and speaker honoraria Medtronic. L.M. speaker fee from Biotronik. M.P.-L. speaker honoraria from Medtronic. A.R. Speaker honoraria from Medtronic, Boston Scientific, Abbott and Phillips. J.M.T. Consulting fees Boston Scientific, Medtronic, Abbott and Microport. All remaining authors have declared no conflicts of interest.

Data availability

No new data were generated or analysed in support of this research.

References

1. Stichering C, Ellenbogen KA, Burri H. Stepping back for good reasons: a reappraisal of the DF-1 connector for definitator leads. Europace 2024;26:eue057.

2. Ellenbogen KA, Auricchio A, Burri H, Gold MR, Leclercq C, Leyva F et al. The evolving state of cardiac resynchronization therapy and conduction system pacing: 25 years of research at EP Europace journal. Europace 2023;25:euad168.

3. Defaye P, Biffi M, El-Chami M, Boveda S, Glikson M, Piccini J et al. Cardiac pacing and lead devices management: 25 years of research at EP Europace journal. Europace 2023; 25:euad202.

4. Glikson M, Burri H, Abdin A, Cano O, Curila K, De Pooter J et al. European Society of Cardiology (ESC) clinical consensus statement on indications for conduction system pacing, with special contribution of the European Heart Rhythm Association of the ESC and endorsed by the Asia Pacific Heart Rhythm Society, the Canadian Heart Rhythm Society, the Heart Rhythm Society, and the Latin American Heart Rhythm Society. Europace 2025;27:euaf050.

5. Kiel EL, Makki T, Kumar R, Gumber D, Kwon DH, Rickard JW et al. Incidence and predictors of right ventricular pacing-induced cardiomyopathy in patients with complete atrioventricular block and preserved left ventricular systolic function. Heart Rhythm 2016;13:2272-8.

6. Khurshid S, Epstein AE, Verdino RJ, Lin D, Goldberg LR, Marchlinski FE et al. Incidence and predictors of right ventricular pacing-induced cardiomyopathy. Heart Rhythm 2014;11:1619-25.

7. Curtis AB, Worley SJ, Adamson PB, Chung ES, Nazi I, Sherfesee L et al. Biventricular pacing for atrioventricular block and systolic dysfunction. N Engl J Med 2013;368: 1585–93.

8. Fruelund PZ, Sommer A, Frekjær JB, Lundbye-Christensen S, Zaremba T, Søgaard P et al. Risk of pacing-induced cardiomyopathy in patients with high-degree

Consensus statement on CIED upgrades and downgrades

atrioventricular block—impact of right ventricular lead position confirmed by computed tomography. J Clin Med 2022;11:7228. 9. Kleiman J, Varrias D, Varkey A, Young A, Wolf E, Gasparis C et al. Incidence and predictors of cardiomyopathy after implantation of leadless pacemakers: a comparative analysis with patients with transvenous systems. Heart Rhythm O2 2024;5:597-600.

10. Heart Rhythm Society. 2012 HRS/ACCF Expert Consensus Statement on Pacemaker Device and Mode Selection. https://www.hhronline.org/guical/clinical-resources/2012-hrsacc-expert-consensus-statement-pacemaker-device-mode-selection (22 August 2024, date last accessed). 11. Brignole M, Auricin A, Baron-Esquivas G, Bordachar P, Boriani G, Brethardt OA et al. 2013 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: The Task Force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA). Eur Heart J 2013;3:42:2281-329. 12. Parkash R, Philipson F, Shanks M, Thibault B, Cox J, Low A et al. Canadian Cardiovascular Society guidelines on the use of cardiac resynchronization therapy: implementation. Can J Cardiol 2013;29:1346-60. 13. Glikson M, Nielsen JC, Kronborg MB, Michowitz Y, Auricin A, Barbash M et al. 2021. ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: developed by the Task Force on cardiac pacing and cardiac resynchronization therapy of the European Society of Cardiology (ESC) with the special contribution of the European Heart Rhythm Association (EHRA). Eur Heart J 2021;42:3427-520. 14. Heart Rhythm Society. 2023 HRS/ARPHSLAHRS Guideline on Cardiac Physiologic Pacing for the Avoidance and Mitigation of Heart Failure. https://www.hhronline/guidance/clinical-resources/2023-hrsarphslahrs-guideline-cardiac-physiologic-pacing-avoidance-and-mitigation-heart-failure (22 August 2024, date last accessed). 15. Burri H, Jastrzebski M, Cano Čurila K, de Pooter J, Huang W et al. EHRA clinical consensus statement on conduction system pacing implantation: endorsed by the Asia Pacific Heart Rhythm Society (APHRS), Canadian Heart Rhythm Society (CHRS), and Latin American Heart Rhythm Society (LAHRS). Europe 2023;25:12:3086. 16. Kaza N, Htun V, Miyazawa A, Simader F, Porter B, Howard J et al. Upgrading right ventricular pacemakers to biventricular pacing or conduction system pacing: a systematic review and meta-analysis. Europace 2022;25:1077-86. 17. Deloyn PP, Otterwanger J, Vos DH. Elvan A. Misser AR, Beukema WP & et al. Upgrading to biventricular pacing guided by pressure-volume loop analysis during implantation. J Cardiovasc Electrophysiol 2011;22:677-83. 18. Ledercq C, Cazeau S. Lellouche D, Fossati F, Anselme F, Davy J M et al. Upgrading from single chamber right ventricular to biventricular pacing in permanently paced patients with worsening heart failure: the RD-CHF Study. Pacing Clin Electrophysiol 2007;30 Suppl 1:52:30. 19. van Geldorp IE, Vernooy K, Delhaas T, Prins MH, Crijns HJ, Prinzen FW et al. Beneficial effects of biventricular pacing in chronically right ventricular paced patients with mid cardiomyopathy. OH 2020;12:223-9. 20. Gierula J, Cubbon RM, Jamil HA, Byrom R, Baxter PD, Pavitt S et al. Cardiac resynchronization therapy in pacemaker-dependent patients with left ventricular dysfunction. Europace 2013;15:1609-14. 21. Barbieri F, Adukauskaite A, Heidbreder a Brandauer E, Bergmann M, Stefani A et al. Central sleep apas and pacing-induced cardiomyopathy. Am J Cardol 2021;139-704. 22. Höijer CJ, Meurling C, Brandt J. Upgrade to biventricular pacing in patients with conventional pacemakers and heart failure: a double-blind, randomized crossover study. Europace 2006;8:51-5. 23. Bogale N, Witte K, Priori S, Cleland J, Auricchio A, Gadler F et al. The European Cardiac Resynchronization Therapy Survey: comparison of outcomes between de novo cardiac resynchronization therapy, operating and upgrades. Eur J Heart Fail 2011;33:974-83. 24. Nemer DM, Patel DR, Madden RA, Wilkoff BL, Rickd JW, Taraki KG et al. Comparative analysis of procedural outcomes and complications between de novo and upgraded cardiac resynchronization therapy. JACC Clin Electrophysiol 2021;7:62-72. 25. Rickard J, Gorodeski EZ, Patel D, Hussein A, Moennich LA, Cantillon D et al. Long term outcomes in patients with chronic right ventricular pacing uparded to cardiac resynchronization therapy. J Cardiovasc Electrophysiol 2019;30:1979-83. 26. New England Journal of Medicine. The Effect of Cardac Resynchronization on Morbidity and Mortality in Heart Failure. https://www.nejm.org/doi/full/10.1056/NEJMa050496 (8 June 2025, date last accessed). 27. Merkely B, Hatala R, Wranicz JK, Durya G, Difcolé, C. Som Z et al. Upgrade of right ventricular pacing to cardia resynchronization therapy in heart failure: a randomized trial. Eur Heart J 2023;44:4259-69. 28. Merkely B, Kosztin A, Roka A, Geller Z, Lima E, Kovacs A et al. Rationale and design of the BUDAPEST-CRT Upgrade Study: a prospective, randomized, multicentre clinical trial. Europace 2017;19:1549-55. 29. Chung MK, Patton KK, Lau CP, Dal Forno ARJ, AlKhat-SM, Arora V et al. 2023 HRS/APHRS/LAHRS guideline on cardiac physiologic pacing for the avoidance and mitigation of heart failure. Heert Rhythm 2023;30:17:91-91.

30. Vijayaraman P, Herweg B, Dandamudi G, Mittal S, Bhatt AG, Marcantoni L et al. Outcomes of His-bundle pacing upgrade after long-term right ventricular pacing and/or pacing-induced cardiomyopathy: insights into disease progression. Heart Rilyhtn 2019:16:1554-61. 31. Hosseini SM, Moazzani K, Rozen G, Vaid J, Saleh A, Kevin Heist E et al. Utilization and in-hospital complications of cardiac resynchronization therapy: trends in the United States from 2003 to 2013. Eur Heart J 2017:38:122-8. 32. Azari A, Kristjänsdöttir I, Gatti P, Berge A, Gadler F. Long- and short-term outcomes after transvenous lead extraction in a large single-centre patient cohort using the clinical frailty scale as a risk assessment tool. Indian Pacing Electrophysiol J 2024:24:263-70. 33. Mehta VS O'Brien H, Elliott MK, Wijesuriya N, Auricchio A, Ayis S et al. Machine learning-derived major adverse event prediction of patients undergoing transvenous lead extraction: using the ESC EHRA ERP European lead extraction ConTRrolled ELECTRAz Hityth 2022;19:885-93. 34. Naruddi RUC Muso, E. Nurchis MC, Domenico P, Scacciavillani R, Bencardino G et al. Mortality after transvenous lead extraction: a risk prediction model for sustainable care delivery. Eur J Clin Invest 2023;53:e13969. 35. Sidhu BS, Ayis S, Gould J, Elliott MK, Mehta V, Kennergren C et al. Risk stratification of patients undergoing transvenous lead extraction with the ELECTRα Registry Outcome Score (EROS): an ESC EUR AORP European lead extraction ConTRrolled ELECTRAz Hity, 2021;23:1462-71. 36. Zeterf EP, Wang Y, Dharmarajan K, Anstrom KJ, Peterson ED, Daubert JP et al. Outcomes 1 year after implantable cardioverter-defibrillator lead abandonment versus explanation for unused or malfunctioning leads. Circ Arrhythm Electrophysiol 2016;9e03953. 37. Pecha S, Burger H, Chung DU, Möller V, Madej T, Maali A et al. The GermAn laser lead extraction� Ritter:Y. GALLERY. Europace 2022:24:1627-35. 38. Blomström-Lundqvist C, Thaykov V, Erba PA, Burri H, Nielsen JC, Bongiorni MG et al. European Heart Rhythm Association (EHRA) international consensus document on how to prevent, diagnose, and treat cardiac implantable electronic device infections —endorsed by the Heart Rhythm Society (HRS), the Asia Pacific Heart Rhythm Society (APHRS), the Latin American Heart Rhythm Society (LAHRS), International Society for Cardiovascular Infectious Diseases (ISCVID) and the European Society of Clinical Microbiology and Infectious Diseases (ESCMID) in collaboration with the Association of Heart Cardio-Thoracic Surgery (EACTS). Europace 2019:22:515-49. 39. Bongiorni MG, Kennergren C, Butter C, Deharo JC, Kutarski A, Rinaldi CA et al. The European Lead Extraction ConTRrolled (ELECTRa) study: a European Heart Rhythm Association (EHRA) registry of transvenous lead extraction outcomes. Eur Heart J 2017:38:2995-3005. 40. Poie JLE, GEha M, It, Chung MK, Uslan DZ, Borge R et al. Complication rates associated with pacemaker or implantable cardioverter-defibrillator generator replacements and upgrade procedures. Circulation 2010;122:1553-61. 41. Kosztin A, Vamos M, Aradi D, Schwertner WR, Kovacs A, Nagy KV et al. De novo implantation vs. Upgrade cardiac resynchronization therapy: a systematic review and meta-analysis. Heart Hold (2018:23:15-26). 42. Olsen T, Jorgensen OD, Nielsen JC, Thøgersen AM, Philbert BT, Frausing MHJP et al. Risk factors for cardiac implantable electronic device infections: a nationwide Danish study. Eur Heart J 2022:43:4946-56. 43. Olsen T, Jorgensen OD, Nielsen JC, Thøgersen AM, Philbert BT, Johansen JB. Incidence of device-related infection in 97 750 patients: clinical data from the complete Danish device-cohort (1982-2018). Eur Heart J 2019:40:1862-9. 44. Mei DA, Imberti JF, Vitolo M, Bonini N, Serafini K, Mantovani M et al. Systematic review and meta-analysis on the impact on outcomes of device algorithms for minimizing right ventricular pacing. Europace 2024;26:eae212. 45. Keene D, Shun-Shin MJ, Arnold AD, March K, Qureshi N, Ng FS et al. Within-patient comparison of His-bundle pacing, right ventricular pacing, and right ventricular pacing and avoidance algorithms with PR prolongation: acute hemodynamic study. J Cardiovasc Electrophysiol 2020;31:2964-74. 46. Borani G, Pieragnoli P, Botto GL, Puerefellner H, Mont L, Ziacchi M et al. Effect of PR interval and pacing mode on persistent atrial fibrillation incidence in dual chamber pacemaker patients: a sub-study of the international randomized MINERVA trial. Europace 2019;21:636-44. 47. Biffi M, Claudio C, Spadotto A, Bartoli L, Sorrentino S, Minguzzi A et al. Pacing devices to treat bradycarda: current status and future perspectives. Expert Rev Med Devices 2021;18:161-77. 48. Ye Y, Zhang Z, Sheng X, Vang B, Chen S, Pan Y et al. Upgrade to his bundle pacing in pacing-dependent patients referred for pulse generator change: feasibility and intermediate term follow up. Int J Cardiol 2018:26:08:82-98. 49. Yang YH, Wang KX, Ma PP, Zhang RF, Waleed KB, Yin X et al. His purkinje system pacing upgrade improvement in the heart performances in patients suffering from pacing-induced cardiomyopathy with or without permanent atrial fibrillation. Int J Cardiol 2021:336-47-51.

D. Keene et al.

50. Zegard A, Naneishvili T, Viyapurapu R, Desai P, White S, Patel PA et al. Diagnostic yield of a heart failure referral pathway using N-terminal pro-brain natriuretic peptide. Open Heart 2023;10:e002469. 51. Verheul M, Groeneveld S, SA Stoks J, Hoeseka MF, Clivtmans MJ, Postema PG et al. The Dutch Idiopathic Ventricular Fibrillation Registry: progress report on the quest to identify the unidentifiable. Neth Heart J 2024;32:238-44. 52. Auricchio A, Demarchi A, Odarski A, Ozartal T, Camparale D, Caputo M, di Valentino M et al. Role of genetic testing in young patients with idiopathic atrioventricular conduction disease. Europace 2023;25:643-50. 53. Blom LJ, Visser M, Christiaans I, Scholten MF, Bootsta M, van den Berg MP et al. Incidence and predictors of implantable cardioverter-defibrillator therapy and its complications in dipatic and in diopatic and in diopatic and in diopatic and in diopatic and in diopatic and in diopatic and in diopatic and in diopatic and in diopatic and in diopatic and in diopatic and in diopatic and in diopatic and in diopatic and in diopatic and in diopati

73. Nowak B, Tasche K, Barnewold L, Heller G, Schmidt B, Bordignon S et al. Association between hospital procedure volume and early complications after pacemaker implantation: results from a large, unselected, contemporary cohort of the German nationwide obligatory external quality assurance programme. Europace 2015;17:787-93. 74. Deharo JC, Dreyfus J, Bongorni MG, Burri H, Defaye P, Glikson M et al. Management of patients with transvalvular right ventricular leads undergoing transcatheter tricuspid valve interventions: a scientific statement of the European Heart Rhythm Association and the European Association of Percutaneous Cardiovascular Interventions of the ESC endorsed by the Heart Rhythm Society, the Asian Pacific Heart Rhythm Society and the Canadian Heart Rhythm Society. Europace 2025;27:euaf061. 75. Becker M, Hoffmann R, Schmitz F, Hundemer A, Kühl H, Schauerte P et al. Relation of optimal lead positioning as defined by three-dimensional echocardiography to long-term benefit of cardiac resynchronization. Am J Cardiol 2007;10:1671-61. 76. Eldadah ZA, Rosen B, Hay I, Edvardsen T, Jayam V, Dickfeld T et al. The benefit of upgrading chronically right ventricle-paced heart failure patients to resynchronization therapy demonstrated by strain rate imaging. Heart Rhythm 2006;3:435-42. 77. Delgado-Montero A, Tayal B, Goda A, Ryo K, Marek JJ, Sugahara M et al. Additive prognostic value of echocardiographic global longitudinal and global circumferential strain to electrocardiographic criteria in patients with heart failure undergoing cardiac resynchronization therapy. Circ Cardiovasc Imaging 2016;9:e004241. 78. Hunt SA, Abraham WT, Chin MH, Feldman AM, Francis GS, Ganiats TG et al. ACC/AHA 2005 guideline update for the diagnosis and management of chronic heart failure in the adult: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Update the 2001 Guidelines for the Evaluation and Management of Heart Failure): developed in collaboration with the American College of Chest Physicians and the International Society for Heart and Lung Transplantation: endorsed by the Heart Rhythm Society. Circulation 2005;112:e154-235. 79. Stanovic L, Prinz C, Clarka A, Daraban AM, Mo Y, Aarones M et al. Long-term outcome after CRT in the presence of mechanical dyssynchrony seen with chronic RV pacing or intrinisc LGBB. LBC Cardiovasc Imaging 2017;10:1091-9. 80. Gorcas J, Lumens J. Rocking and flashing with RV pacing. JACC Cardiovasc Imaging 2017;10:1100-2. 81. Andreas M, Burri H, Praz F, Soliman O, Badano L, Barreiro M et al. Tricuspid valve disease and cardiac implantable electronic devices. Eur Heart J 2024;45:346-65. 82. Addetia K, Harb SC, Hahn RT, Kapadia S, Lang RM. Cardiac implantable electronic device lead-induced tricuspid regurgitation. JACC Cardiovasc Imaging 2019;12:622-36. 83. Tatum R, Maynes EJ, Wood CT, Deb AK, Austin MA, O'Malley TJ et al. Tricuspid re-gurgitation associated with implantable electrical device insertion: a systematic review and meta-analysis. Pacing Clin Electrophysiol 2021;44:1297-302. 84. SdiK 9, Sallih S, Hatti DI, Tricuspid regurgitation in His bundle pacing: a systematic review. Ann Noninvasive Electrocardiol 2022;27:e12986. 85. Ip JE, Krishnan U, Girardi LN, Lerman BB. A hybrid endocardial-epicardial biventricular implantable cardioverter-defibrillator to circumvent the tricuspid valve. Pacing Clin Electrophysiol 2021;44:399-401. 86. Boersma LV, El-Chami M, Steinwender C, Lambiase P, Murgatroyd F, Mela T et al. Practical considerations, indications, and future perspectives for leadless and extravas-cular cardiac implantable electronic devices: a position paper by EHRA/HRS/LAHS/APHRS. Europace 2022;24:1691-708. 87. Beurskens NEG, Tjong FVY, de Bruin-Bon RHA, Dasselaar KJ, Kuijt WJ, Wilde AAM et al. Impact of leadless pacemaker therapy on cardiac and atrioventricular valve function through 12 months of follow-up. Circ Arrhythm Electrophysiol 2019;12:e007144. 88. Sommer A, Kronborg MB, Norgaard BL, Poulsen SH, Bouchelouche K, Böttcher M et al. Multimodality imaging-guided left ventricular lead placement in cardiac resynchronization therapy a randomized controlled trial. Eur Heart Fail 2016;18:1362-74. 89. Borgustil R, Carlsson M, Markstad H, Werther-Evaldsson A, Ostenfeld E, Roijer A et al. Cardiac resynchronization therapy guided by echocardiography, MRI, and CT imaging a randomized controlled study. JACC Clin Electrophysiol 2020;6:1300-9. 90. Khan FZ, Virdee MS, PRAM, CUR, Pugh PJ, O'Halloran D, Elsik M et al. Targeted left ventricular lead placement to guide cardiac resynchronization therapy: the TARGET study: a randomized controlled trial. J Am Coll Cardiol 2021;59:1509-18. 91. Saba S, Marek J, Schwartzman D, Jain S, Adelstein E, White P et al. Echocardiography-guided left ventricular lead placement for cardiac resynchronization therapy. Circ Heart Fail 2013:6:427-34. 92. Singh JP, Rinaldi CA, Sanders P, Kubo SH, James S, Niazi IK et al. Leadless ultrasound-based cardac resynchronization system in heart failure. JAMA Cardiol 2024;9:871-9. 93. Leya F. The role of cardiovascular magnetic resonance in cardiac resynchronization therapy. Heart Fail 2017;13:63-77. 94. Fybeno DB, Kronborg MB, Nielsen JC. Go for the right left ventricular lead position at initial implantation of a cardiac resynchronization therapy device. Heart Rhythm 02:2023;4:64-5 10. Leya F. Zegard A, Acquaye E, Gubran C, Taylor R, Foley PWX et al. Outcomes of cardiac resynchronization therapy with or without defibrillation in patients with nonischemic cardiomyopathy. J Am Coll Cardiol 2017;70:1216-27.

Consensus statement on CIED upgrades and downgrades

96. Gerrits W, Wouters P, Chiu CSL, Guglielmo M, Cramer MJ, van der Harst P et al. Optimizing CRT lead placement accuracy with CMR-guided on-screen targeting a randomized controlled trial (ADVISE-CRT III). JACC Clin Electrophysiol 2025;11:1293-305.87. Kontempli L.Aboehassan M, Cerini M, Salghetti F, Fabbricore D, Maiolo V et al. Technical considerations for CRT-D implantation in different varieties of persistent left superior vena cava. J Interv Card Electrophysiol 2021;61:517-24.98. BurriH, Starck-C Auricchio A Biffi M, Burri M, D-Avila A et al EHRA expert consensus statement and practical guide on optimal implantation technique for conventional pacemakers and implantable cardioverter-defibrillators: endorsed by the Heart Rhythm Society (HRS), the Asia Pacific Heart Rhythm Society (APHRS), and the Latin-American Heart Rhythm Society (LAHRS). Europace 2021;23:983-1008.99. Starkovic L. VoigtV JU, HurriH, Muraru D, Sade LE Hauga KH et al imaging in patients with cardiovascular implantable electronic devices: part 2-imaging after device implantation. A clinical consensus statement of the European Association of Cardiovascular Imaging (EACVI) and the European Heart Rhythm Association (EHRA) of the ESC-Eur Heart Jcardov Imaging 2023;25:33-54.100.78. Worley SJ, Gohn DC, Pulliam RW, Raifsnider MA, Ebersole Bl, Tuzi J. Subclavian venoplasty by the implanting physicians in 373 patients over 11 years. Heart Rhythm 2011:8:526-33.10 101. Vitali F, Malagü M, Bianchi N, De Refalle M, Manfrini M, Gibgi F et al. Ultrasound-guided venous axillary access versus standard fluoroscopic technique for cardiac lead implantation. JACC Clin Electrophysiol 2024;10:554-65.102.101.75. Zuin F, Zuin M, Charles P, Jiménez-Diaz J, Sheldon SH, Tagli AP et al. Ultrasound-guided vs. fluoro-guided axillary venous access for cardiac implantable electronic devices: a patient-based meta-analysis. Europace 2024;26:eau274.103.109. DGiles Stiles PK. Stewart JT, Armartng GP. Ultrasound-guided venous access for permanent pacemaker leads. Pacing Clin Electrophysiol 2006;29:852-7.104.103.106:47.25.124.103.109:46.151.102.106:47.25.124.103.106:47.25.103.105.106:47.25.103.105.106:47.25.103.105.106:47.25.103.105.106:47.25.103.105.106:47.25.103.105.106:47.25.103.105.106:47.25.103.105.106:47.25.103.105.106:47.25.103.105.106:47.25.103.105.106:47.25.103.105.106:47.25.103.105.106:47.25.103.105.106:47.25.103.105.106:47.25.103.105.106:47.25.103.105.106:47.25.103.105.106:47.25.103.105.106:47.25.103.105.106:47.25.103.105.106:47.25.103.105.106:47.25.103.105.106:47.25.103.105.106:47.25.103.105.106:47.25.103.105.106

119. Tilney NL, Griffiths HJG, Edwards EA. Natural history of major venous thrombosis of the upper extremity. Arch Surg 1970;101:792–6. 120. Demeter SL, Pritchard JS, Piedad OH, Cordasco EM, Taherj S. Upper extremity thrombosis: etology and prognosis. Angiology 1982:33:7343–55. 121. Sethi GK, Bhayana JN, Scot SM. Innominate venous thrombosis: a rare complication of transvenous pacemaker electrodes. Am Heart J 1974;87:770–2. 122. Blackburn T, Dunn M. Pacemaker-induced superior vena cava syndrome: consideration of management. Am Heart J 1988;116:893–6. 123. Pauletti M, Pingitore R, Contini C. Superior vena cava stenosis at site of intersection of two pacing electrodes. Br Heart J 1979;42:487–9. 124. Szakjowski M, Polewczyk A, Jacheć W, Kosior J, Nowosielecka D, Tulecki Ł. Multilevel venous obstruction in patients with cardiac implantable electronic devices. Medicina (Kauas) 2024:60:336. 125. Abu-El-Haija B, Bhave PD, Campbell DN, Mazur A, Hodgson-Zingman DM, Cotarlan V et al. Venous stenosis after transvenous lead placement: a study of outcomes and risk factors in 212 consecutive patients. J Am Heart Assoc 2015;4:e001878. 126. Knorr D, Bandorski D, Bogossian H, Iliodromitis K, Schiedat F, Karosiene Z et al. Cephalic vein cutdown is superior to subclavian puncture as venous access for patients with cardiac implantable devices after long-term follow-up. J Clin Med 2024;13:1044. 127. Atti V, Turagam MK, Garg J, Koerber S, Angirekula A, Gopinathannair R et al. Subclavian and axillary vein access versus cephalic vein cutdown for cardiac implantable electronic device implantation: a meta-analysis. JACC Clin Electrophysiol 2020;6:66:61–71. 128. Chan NY, Kwong NP, Cheong AP. Venous access and long-term pacemaker lead failure: comparing contrast-guided axillary vein puncture with subclavian puncture and cephalic cutdown. Europace 2017;19:1193–7. 129. Marcial JM, Worley SJ, Verous system interventions for device implantation. Card Electrophysiol Clin 2018;10:163–77. 130. Sohal M, Williams S, Ahhtar M, Shah A, Chen Z, Wright M et al. Laser lead extraction to facilitate cardiac implantable electronic device upgrade and revision in the presence of central venous obstruction. Europace 2014;16:81–7. 131. Worley SJ, Gohn DC, Pullam RW. Excimer laser to open refractory subclavian occlusion in 12 consecutive patients. Heart Rhythm 2010;7:634–8. 132. Gula LJ, Ames A, Woodburn A, Matkins J, McCormick M, Bell J et al. Central venous occlusion is not an obstacle to device upgrade with the assistance of laser extraction. Pacing Clin Electrophysiol 2005;28:661–6. 133. Lüthje L, Zabel M, Seegers J, Zenker D, Vollmann D. Acute and long-term feasibility of contralateral transvenous lead placement with subcutaneous, pre-sternal tunnelling in patients with chronically implanted rhythm devices. Europace 2011;13:1004–8. 134. Kusumoto FM, Schoenfeld MH, Wilkoff BL, Berul CI, Birgersdotter-Green UM, Carrillo R et al. 2017 HRS expert consensus statement on cardiovascular implantable electronic device lead management and extraction. Heart Rhythm 2017;14:e503–51. 135. Zipes DP, Roberts D. Results of the international study of the implantable pacemaker cardioverter-defibrillator. Circulation 1995;92:59–65. 136. Lawrie GM, Seale JP, Morris GC, Howell JF, Whisennand HH, DeBakey ME. Results of epicardial pacing by the left subcostal approach. Ann Thorac Surg 1979;28:561–7. 137. Molina JE. Surgical options for endocardial lead placement when upper veins are obstructed or nonusable. J interv Card Electrophysiol 2004;11:149–54. 138. Hejel L, Németh M, Mel�zer L, Kónyi A. Cardiac resynchronization therapy with intraoperative epicardial mapping via minithoracotomy: 10 years' experience. Pacing Clin Electrophysiol 2021;44:101–9. 139. Reynolds D, Duray GZ, Omar R, Soejima K, Neuzil P, Zhang S et al. A leadless intracardia; transcatheter pacing system. N Engl J Med 2016;374:533–41. 140. Reddy VY, Exner DV, Cantillon DJ, Doshi R, Bunch TJ, Tomassoni GF et al. Percutaneous implantation of an entirely intracardia leadless pacemaker. N Engl J Med 2015;373:1125–35. 141. Knops RE, Reddy VY, Ip JE, Doshi R, Exner DV, Defaye P et al. A dual-chamber leadless pacemaker. N Engl J Med 2023;388:2360–70. 142. Ip je Rastimah M, Enerv DV, Reddy VY, Doshi R, Badie N et al. Atrioventricular synchrony delivered by a dual-chamber leadless pacemaker system. Circulation 2024;150:439–50. 143. Bardy GH, Smith WNP, Hood MA, Crozier IG, Melton IC, Jordaens L et al. An entirely subcutaneous implantable cardioverter-defibrillator. N Engl J Med 2010;363:36–44. 144. Knops RE, Olde Nordkamp LRA, Delnoy PPHM, Boersma LVA, Kuschyk J, El-Chami MF et al. Subcutaneous on transvenous defibrillator therapy. N Engl J Med 2020;383:526–35. 145. Ip JE (Completely wireless: first concomitant dual-chamber leadless pacemaker and subcutaneous defibrillator implant. Heart Rhythm 2024;22:403–4. 146. Knops RE, Lloyd MS, Roberts PR, Wright DJ, Boersma LVA, Doshi R et al. A modular communicative leadless pacing-defibrillator system. N Engl J Med 2024;391:1402–12. 147. Erath JW, Sirat AS, Vamos M, Hohnloser SH. Epicardial CRT-P- and S-ICD implantation in a young patient with persistent left superior vena cava. Herzschrittmacherther Elektrophysiol 2016;27:39–8.8. 148. Römers H, Liebregts M, van Dijk V, Boersma L. The substernal implantable cardioverter-defibrillator: first experience combining an extravascular-implantable

D. Keene et al.

cardioverter-defibrillator with a dual-chamber transvenous pacemaker. HeartRhythm Case Rep 2023:10:132–6. 149. Friedman P, Murgatroyd F, Boersma LVA, Manlucu J, O'Donnell D, Knight BP et al. Efficacy and safety of an extravascular implantable cardioverter-defibrillator. N Engl J Med 2022:387:1292–302. 150. Osswald BR. Minimal touch surgical implantable cardioverter defibrillator implantations. Interact Cardioverz. Ther Surg 2006;5343–4. 151. Burger H, Pecha S, Hakmi S, Opalka B, Schoenburg M, Ziegelhoeffer T. Five-year follow-up of transvenous and epicardial left ventricular leads: experience with more than 1000 leads. Interact Cardioverz. Ther Surg 2020;30:74–80. 152. van Dijk VF, Fanggiday J, Balt JC, Wijfels MCEF, Daeter EJ, Kelder JC et al. Effects of epicardial versus transvenous left ventricular lead placement on left ventricular function and cardiac perfusion in cardiac resynchronization therapy: a randomized clinical trial. J Cardiovasc Electrophysiol 2017;28:917–23. 153. Friedman P, Murgstroyd F, Boersma LVA, Manlucu J, Knight BP, Clémenty N et al. Performance and safety of the extravascular implantable cardioverter defibrillator through long-term follow-up: final results from the pivotal study. Circulation. 2025; 151:322–32. 144. Crozier I, Murgatroyd F, Amin A, Clementy N, Duncker D, Kotschet E et al. Perprocedural outcomes from the postmarket study of the extravascular implantable cardioverter defibrillator: preliminary Lighten study results and meta-analysis. Heart Rhythm 2025:5271(25)001274-4. doi:10.1016/j.hthm.2025.02.012. Epub ahead of print. 155. Reddy VY. Miller MA, NeuziP. Segard P, Butter C. Seifert M et al. Cardac resynchronization therapy with wireless left ventricular endocardial pacing: the SELECT-LV study. J Am Coll Cardiol 2017;69:2119–29. 156. Okate T, Hummel JD, Bank AJ, Naziz IK, McGrew FA, Kindvater S et al. Leadless left ventricular stimulation with WISE-CRT system—initial experience and results from phase I of SOLVE-CRT study (nandomrandized, roll-in phase). Heart Rhythm 2022; 19:22–9. 157. Elay CS, Allen CL, Leung S, Lusher S, Morales GX, Wiisanen M et al. Inside-out access: a new method of lead placement for patients with central venous occlusions. Heart Rhythm 2011:8;851–7. 158. Mathur G, Stables RH, Heaven D, Ingram A, Sutton R. Permanent pacemaker implantation via the femoral vein: an alternative in cases with contraindications to the pectoral approach. Europace 2001:3;56–9. 159. Ching CK, Elay CS, Di Biase L, Barrett CD, Martin DO, Saliba Wl et al. Transiliac ICD implantation: defibrillation vector flexibility produces consistent success. Heart Rhythm 2009;6:978–83. 160. Eladse MH, French J. Iliac vein approach to permanent pacemaker implantation. Pacing Clin Electrophysiol 1989;12:1030–3. 161. Pandian NG, Kosowsky BD, Gurewich V. Transfemoral temporary pacing and deep vein thrombosis. Am Heart J 1980;100:847–51. 162. Erdogan O, Augustini R, Saliba W, JuratiN, Wilkoff BL. Transiliac permanent pacemaker implantation after extraction of infected pectoral pacemaker systems. Am J Cardiol 1999;84:474–5. 163. Brignole M, Deharo JC. Different techniques of venous access for CIEDs: advantages and disadvantages. Eur Heart J 2023;44-4859–61. 164. Su J, Kusumoto FM, Zhou X, Elayi CS. How to perform extrathoracic venous access for cardiac implantable electronic device placement: detailed description of techniques. Heart Rhythm 2022;19:1184–91. 165. Lee CS, Yoon SH, Lee SM, Lee IK, Kim JY, Cho HM et al. Micropurecture access set use during implantation of totally implantable venous access device may reduce upper extremity DVT incidence among patients undergoing chemotherapy for colorectal cancer. World J Surg 2020;44:1302–8. 166. Perna F, Flore T. Telesca A, Scacciavillani R, Bencardino G et al. Ultrasound-guided axillary vein puncture versus landmark-guided approach for cardiac implantable electronic device placement. Pacing Clin Electrophysiol 2025;48–9–20. 167. Lipitic E, Daniels F, Groenfeld H, Rienstra M, Mass AH. When and how to perform venoplasty for lead placement. Heart Rhythm 2024;21:1923–8. 168. Ji SY, Gundewar S, Palma EC. Subclavian venoplasty may reduce implant times and implant failures in the era of increasing device upgrades. Pacing Clin Electrophysiol 2012;35:444–8. 169. Ferro EG, Kramer DB, Li S, Locke AH, Misra S, Schmaa AA et al. Incidence treatment, and outcomes of symptomatic device lead-related venous obstruction. J Am Coll Cardiol 2023;81:2328–40. 170. Zimetbaum P, Carroll BJ, Locke AH, Secemsky E, ScherhermH M. Lead-related venous obstruction in patients with implanted cardiac devices. J Am Coll Cardiol 2022;79–299–308. 181. Nowad S, Vance AZ, Cobb RM, Mantell PP, Cohen R, Clark TVI, Radiofrequency guidewire-facilitated recanalization of chronic thoracic central venous occlusions in hemodialysis. CVIR Endovasc 2024;7:10. 196. Melyk W, Ibrahim R, Westerman S, Shah A, Bhatia NK, Merchant FM et al. Procedural and long-term outcomes of tunneled transvenous leads. Heart Rhythm 2024;22:1307–11.

173. Bracke FA, Rademakers LM, van Veghel D. Extraction of non-infected redundant pacing and defibrillator leads does not result in better patient outcomes. Neth Heart J 2023:31:327–9. 174. Lee JJZ, Tabel F, Tan MC, Srivathsan K, Sorajja D, Valverde A et al. Long-term outcomes with abandoning versus extracting sterile leads: a 10-year population-based study. Pacing Clin Electrophysiol 2024:47:626–34. 175. Wazni O, Epstein M, Carrillo RG, Love C, Adler SW, Riggio DW et al. Lead extraction in the contemporary program setting the LExICon study. J Am Coll Cardiol 2010;55:579–86. 176. Milman A, Leshem E, Massalla E, Jia K, Meitus A, Kariv S et al. Occluded vein as a pre-dector for complications in non-infectious transvenous lead extraction. Front Cardiovasc Med 2022;9:1016657. 177. Khan F, Sverin G, Birgersditter-Green U, Miller JP, Lalani G, Pollema T et al. Risk of collateral lead damage in percutaneous cardiac implantable electronic device extrac-tion. JACC Clin Electrophysiol 2018;4:193–200. 178. Barakat AF, Wazni OM, Taraki Kji, S Caliba WI, Nimri N, Rickard J et al. Transvenous lead extraction at the time of cardiac implantable electronic device upgrade: complexity, safety, and outcomes. Heart Rhythm 2017;14:1807–11. 179. Witte OA, Adiyaman A, van Bemmel MW, Smit JJJ, Ghani A, Misier ARR et al. Mechanical power sheath medecalization and lead implantation in patients with venous occlusion: technique and results. J Cardiovasc Electrophysiol 2018;29: 316–21. 180. Tonko JB, Blauth C, Rosenthal E, Rinaldi CA. Completely epicardial implantable cardi-overto/defrillator (IC) and CRT-D systems: a case series and systematic literature review. Pacing Clin Electrophysiol 2021;44:1616–30. 181. da Silva GL, Cortez-Dias N, de Sousa J, Nobre Ã, Pinto FJ. Surgical epicardial CRT-D implantation in a patient with complete obstruction of the superior vena cava. Arq Bras Cardiol 2018;110:490–2. 182. Boersma LVA, Amin A, Clénenty N, Duncker D, Engel G, Epstein L et al. Design of a post-market registry for the extravascular implantable cardioverter-defibrillator: the Lighten Study. Heart Rhythm 02 2024;6:64–9. 183. Carabelli A, Jabeur M, Jacon P, Rinaldi CA, Leclercq C, Rovaris G et al. European ex-perence with a first totally leadless cardiac resynchronization therapy pacemaker system. Europace 2021;23:740–7. 184. Voirol U, Burn H. Standard pacemaker implantation via femoral venous access. Cardiovasc Med 2017;20:101–3. 185. Kaya E, Totzeck M, Rassaf T. Pulsed electron avalanche knife (PEAK) PlasmaBlade™ in pacemaker and defibrillator procedures. Eur J Med Res 2017;22:49.

186. Kyta A, Blessberger H, Saleh K, Hónig S, Kammler J, Neeser K et al. An electrical plasma surgery tool for device replacement—retrospective evaluation of complications and economic evaluation of costs and resource use. Pacing Clin Electrophysiol 2015; 38:28–34. 187. Mittal S, Wilkoff BL, Olele JE, Kennergren C, Wright DJ, Berman BJ et al. Low-temperature electrocetary reduces adverse effects from secondary cardiac implantable electronic device procedures: insights from the WRAP-IT trial. Heart Rhythm 2011;18:1142–50. 188. Wasserlauf J, Esheim T, Jarett NM, Chan EKY, Schaller RD, Garcia FC et al. Avoiding damage to transvenous leads—a comparison of electrocautery techniques and two insulated electrocautery blades. Pacing Clin Electrophysiol 2018;41:1593–9. 189. Chiang C, Pivparth, S. Ayub MT, Trohman RG, Larsen TR, Huang HD et al. Comparison of electrocautery platforms for pulse generator replacement procedures. J Interv Card Electrophysiol 2022;64:557–8. 190. Kleemann T, Becker T, Strauss M, Dyck N, Weisse U, Saggau W et al. Prevalence of bacterial colonization of generator pockets in implantable cardioverter defibrillator patients without signs of infection undergoing generator replacement or lead revision. Europe 2010;12:58–63. 191. Essebag V, Verma A, Healey J, KSron A, Kalfon E, Coutu B et al. Clinically significant pocket hematoma increases long-term risk of device infection: BRUISE CONTROL INFECTION study. J Am Coll Cardiol 2016;67:1300–8. 192. Lakireddy D, Pillarsetti J, Atkins D, Biria M, Reddy M, Murray C et al. IMPact of pocKet eivision of the rate of Infectration and other CompLications in patients rEquiring pocket manAponulation for generator replacement and/or lead replacement or revision. (MAKE IT CLEAN): a prospective randomized study. Heart Rhythm 2015;12:9506. 193. Grazia Bongiorni M, Dagres N, Estner H, Pison L, Todd D, Blomstrom-Lundqvist C et al. Management of malfunctioning and recalled pacemaker and defibrillator leads to the European Heart Rhythm Association survey. Europace 2014;16:1674–8. 194. Pokoney SD, Mi X, Lewis RK, Greiner M, Epstein LM, Carrillo RG et al. Outcomes associated with extraction versus capping and abandoning pacing and defibrillator leads. Circulation 2017;136:1387–95. 195. Darouiche RO, Wall MJ, Itani KMF, Otterson MF, Webb AL, Carrick MM et al. Chlorhexidine-alcohol versus povidone-iodine for surgical-site antisepsis. N Engl J Med 2010;362:18–26. 196. Golan M, Berbetene NM, Odabashian R, Sadek MM, Corrales-Medina V, Aydin A et al. Adherent skin barrier drape is associated with a reduced risk of cardiac implantable device infection: results from a prospective study of 14,225 procedures. Heart Rhythm O2 2024;0234–103–12.

Consensus statement on CIED upgrades and downgrades

197. Aydin A, Golian M, Klein A, Redpath C, Davis DR, Ramirez DF et al. Do barrier dress-singe direct cardiac implantable device infection: protocol for a randomized controlled trial (BARRIER-PROTECT). Helyion 2023;9:e22229. 198. Diaz JC. Braunstein ED, Calas F, Duque M, Marin JE, Aristizabal J et al. Chlorhexidine gluconate pocket lavage to prevent cardiac implantable electronic device infection in high-risk procedures. Heart Rhythm 2023;20:1674-81. 199. Krain AD, Longtin Y, Philipson F, Birnie DH, Manuju, Anguran P et al. Prevention of arrhythmia device infection: the PADIT trial. J Am Coll Cardiol 2018;72:3098-109. 200. Malagú M, Vitali F, Brieda A, Cimaglia P, De Raffele M, Tazzari E et al. Antibiotic prophylaxis based on individual infective risk stratification in cardiac implantable electronic device: the PRACTICE study. Europace 2021;24:413-20. 201. Braterd DW, Delinger EP, Olsen K.P, Perl TH, Auwaarter P, Golon MK et al. Clinical practice guidelines for antimicrobial prophylaxis in surgery. Am J Health Syst Pharm 2013;70:195-283. 202. Baddour LM, Epstein AE, Erickson CC, Knight BP, Levison ME, Lockhart PB et al. Update on cardiovascular implantable electronic device infections and their management: a scientific statement from the American Heart Association. Circulation 2010;121:458-77. 203. Tarakji KG, Mittal S, Kennergren C, Corey R, Poole JE, Schloss E et al. Antibacterial en-velope to prevent cardiac implantable device infection. N Engl / Med 2019;38:1895-905. 204. Frausing MHJP, Nielsen JC, Johansen JB, Jørgensen OD, Gerdes C, Olsen T et al. Rate of device-related infections using an antibacterial envelope in patients undergoing cardiac resynchronization therapy reoperations. Europace 2022;24:421-9. 205. Kumar A, Doshi R, Shariff M. Role of antibiotic envelopes in preventing cardiac implantable electronic device infection: a meta-analysis of 14 859 procedures. J Arrhythm 2020;36:176-9. 206. Ellis CR, Greenspan AJ, Andriulli JA, Gould PA, Carillo RG, Kolek MJ et al. Randomized trial of stand-alone use of the antimicrobial envelope in high-risk cardiac device patients. Circ Arrhythm Electrophysiol 2023;16:e011740. 207. Birnhe DH, Wang J. Alings M, Philipson F, Parkash R, Manucu J et al. Risk factors for infections involving cardiac implantable electronic devices. J Am Coll Cardiol 2019;74:2845-54. 208. Maclean E, Mahtari K, Honarbakshh S, Butcher C, Ahluwala N, Dennis ASC et al. The BLISTER score: a novel, externally validated tool for predicting cardiac implantable electronic device infections, and its cost-utility implications for antimicrobial envelope use. Circ Arphyth Electrophysiol 2024;17:012444. 209. Biffi M, Götte A, Wright J, Trucco E, Klug D, Turley A et al. Cardiac implantable electronic device infection awareness—a European survey amongst implanting physicians. Int J Cardiol 2024;415:132454. 210. Ravendren A, Kaza N, Naraen A, Geneste A, Zuhair M, Porter B et al. Results from a national observational survey of UK cardiologists on antibiotic envelope use during cardiac electronic device implantation. BMJ Open 2025;15:e096157. 211. Frausing MHJP, Johansen JB, Afonso D, Jørgensen OD, Olsen T, Gerdes C et al. Cost-effectiveness of an antibacterial envelope for infection prevention in patients undergoing cardiac resynchronization therapy reoperations in Denmark. Europace 2023;59:459. 212. Wilkoff BL, Boriani G, Mittal S, Poole JE, Kennergren C, Corey GR et al. Cost-effectiveness of an antibacterial envelope for cardiac implantable electronic device infection in the US healthcare system from the WRAP-IT trial. Circ Arrhythm Electrophysiol 2020;13:e008503. 213. Boriani G, Vitolo M, Wright DJ, Biffi M, Brown B, Taraki KG et al. Infections associated with cardiac electronic implantable devices: economic perspectives and impact of the TYRX™ antibacterial envelope. Europace 2021;23:iv33-44. 214. Xiang K, Catarzano JN, Esiorgin Rahantix-EL-oluting envelopes to prevent cardiac-implantable electronic device infection: past, present, and future. Cureus 2021;13:e13088. 215. Sohal MR, Esquer Garrigos Z, ElayS, Ciang K, Catanzaro JN. Precinal evaluation of efficacy and pharmacokinetics of gentaminic containing extracellular-matrix envelope. Pacing Clin Electrophysiol 2020;43:341-9. 216. Virrman 1, Ripponson F, Mittal S, Finn A, Kudik D, Kirchhof N et al. Effects of envelopes on cardiac implantable electronic device pocket healing: ha head-to-head pre clinical evaluation. Heatr Hrytin 2024;24:132-35. 217. Ip JE, Xu L, Lerman BB. Differences between cardiac implantable electronic device envelopes evaluated in an animal model. J Cardiovasc Electrophysiol 2021;32:1346-54. 218. Pirrmann RW, Leslie GB. The ant-endotoxin activity of turoalin in experimental animals. J Appl Bacterial 1979;46:97-102. 219. Blenkharn Jl. Sustained anti-adherence activity of taurolidine (Taurolin) and noxythiolin (Noxyflex S) solutions. J Pharm Pharmacol 1988:409-511. 220. Caruso F, Darnowski JW, Opazo C, Goldberg A, Kishore N, Agoston ES et al. Taurolidine antiadine properties on interaction with E. coli its transformation in biological environment and interaction with bacteria cell. PLoS One 2010;5:e8927. 231. Dincer S, Babül A, Erdoğan D, Ozoğul C, Dinger SL. Effect of taurine on wound healing. Amino Acids 1996;10:59-71.

222. Borov S, Baldauf B, Henke J, Pavaci H, Perani A, Zrenner B et al. Use of a taurolidine containing antimicrobial wash to reduce cardiac implantable electronic device infection. Europace 2023;25:eud306. 23. de Oliveira JC, Martinelli M, Nishioka SAD, Varejão T, Uipe D, Pedrosa AAA et al. Efficacy of antibiotic prophylaxis before the implantation of pacemakers and cardioverter-defibrillators. Circ Arrhythm Electrophysiol 2009;2:29-34. 24. Smantrakels EN, Nakou ES, Vardas PE. Upgrading or downgrading a cardiac resynchroization therapy device (CRT) device? Gaps and dilemmas in current clinical practice. Europace 2018;20:217-8. 25. Chung MK, Fagerlin A, Wang PJ, Ajayi TB, Allen LA, Baykaner T et al. Shared decision making in cardiac electrophysiology procedures and arrhythmia management. Circ Arhythm Electrophysiol 2021;14:e007958. 26. Lewis KB, Nery PB, Birne DH. Decision making at the time of ICD generator change: patients' perspectives. JAWA Intern Med 2014;174:1508-11. 27. Padleletti L, Arrar DO, Boncelliin L, Brachman J, Camm JA, Daubert JC et al. EHRA Expert Consensus Statement on the management of cardiovascular implantable electronic devices in patients nearing end of life or requesting withdrawal of therapy. Europace 2010;12:1480-9. 28. Ting HH, Brito JP, Montori VM. Shared decision making. Circ Cardiovasc Qual Outcomes 2014;7:233-7. 29. Nakou ES, Simantirakis E, Kallergis EM, Nakos KS, Vardas PE. Cardiac resynchronization therapy (CRT) device replacement considerations: upgrade or downgrade? A complex decision in the current clinical setting. Europace 2017;19:705-11. 29. Savelieva I, Fumagalli S, Kenny RA, Anker S, Benetos A, Boriani G et al. EHRA expert consensus document on the management of arrhythmias in frailty syndrome, endorsed by the Heart Rhythm Society (HRS), Asia Pacific Heart Rhythm Society (APHRS), Latin America Heart Rhythm Society (LAHRS), and Cardiac Arrhythmia Society of Southern Africa (CASSA). Europace 2023;25:1249-76. 23. Gausz FD, Lena KNM, Geedon PE, Miklos M, Benak A, Bencsik G. Arrhythmia detection in atrioventricular, single-lead, floating atrial dipole ICD systems compared with conventional single-and dual-chamber defibrillators. J Cardiovasc Dev Dis 2024;11:386. 23. Biffi M, lori M, De Maria E, Bolognesi MG, Placci A, Calvi V et al. The role of atrial sensing for new-onset atrial arrhythmias diagnosis and management in single-chamber implantable cardioverter-defibrillators recipients: results from the THINGS registry. J Cardiovasc Electrophysiol 2020;31:846-53. 23. Hindkins G, The Daurs, Bar-Lev D, Anguera I, Ayala Paredes FA, Arnold M et al. Ability to remotely monitor atrial high-rate episodes using a single-chamber implantable cardioverter-defibrillators with a cardioverter-defibrillator with a floating atrial sensing dipole. Europace 2023;25:40261.14:8677. 23. Khurshid S, Obeng-Gyimah E, Supple GE, Schaller R, Lin D, Owens AT et al. Reversal of pacing-induced cardiomyopathy following cardiac resynchronization therapy. JACC Clin Electrophysiol 2018;416:168-77. 235. Okazaki M, Sahashi Y, Nagase T, Inoue K, Sekiguchi Y, Nitta J et al. Inappropriate shock incidence in patients with subcutaneous implantable cardioverter-defibrillators with concomitant cardiac implantable electronic devices: a single-center cohort study. Pacing Clin Electrophysiol 2024;47:131-8. 236. Weipert KF, Kostic S, Gokyldrim T, Johnson V, Chasan R, Gemein C et al. Safety and performance of the subcutaneous implantable cardioverter defibrillator detection algorithm INSIGHTTM in pacemaker patients. J Clin Med 2023;13:129. 237. Kajiyama T, Kondo Y, Nakano M, Chiba T, Kobayashi Y. Concomitant use and interaction of a subcutaneous implantable defibrillator and epicardial pacemaker. J Arrhythmia 2022;38:155-6. 238. Mei DA, Imberti JF, Vitolo M, Bonini N, Gerra L, Romiti GF et al. Single-lead VDD pa-cing: a literature review on short-term and long-term performance. Expert Rev Med Devices 2023;20:187-97. 239. Imnadez G, Fink T, Etzec i Zhocchi N, Artonyan L, Harutyunyan K et al. Standard defibrillators for left bundle branch area pacing. JACC Clin Electrophysiol 2024;10:2263-8. 240. Afal MR, Jamal SM, Son JH, Chung JH, Gabriels J, Okabe T et al. Tips and tricks for safe retrieval of the based leadless pacemakers. J Innov Card Rhythm Manog 2021;12:4562-8. 241. Dar T, Akella K, Murtaza G, Sharma S, Afzal MR, Gopinathannair R et al. Comparison of the safety and efficacy of Nanostim and Micra transcatheter leadless pacemaker (LP) extractions: a multicenter experience. J Interv Card Electrophysiol 2020;57:133-40. 242. Neuzil P, Exer DV, Knops RE, Cantillon DJ, Defaye P, Banker R et al. Worldwide chronic retrieval experience of helix-fixation leadless cardiac pacemakers. J Am Coll Cardiol 2025;85:1111-20.

D. Keene et al.

246. Bongiorni MG, Reddy VY, Ip JE, Doshi R, Exner DV, Defaye P et al. Upgrading a single-chamber leadless pacemaker to a dual-chamber leadless pacemaker system. Europace 2024;26:eae102.392. 247. Ip JE Modular leadless pacing: a case series of leadless pacemaker up. JACC Clin Electrophysiol 2025;11:854-64. 248. Ip JE. Exertional intolerance after a leadless pacemaker implant: what is the mechanism? Pacing Clin Electrophysiol 2025;4:394-6. 249. Ito R, Kondo Y, Winter J, Hayashi T, Nakano M, Kajiyama T et al. Combination of a leadless pacemaker and subcutaneous implantable cardioverter defibrillator therapy for a Japanese patient with prosthetic valve endocarditis. J Arrhythmia 2019;35:311-35. 250. Ljungström E, Brandt J, Mörtsell D, Borgquist R, Wang L. Combination of a leadless pacemaker and subcutaneous defibrillator with nine effective shock treatments during follow-up of 18 months. J Electrocardiol 2019;56:1-3. 251. Montgomery JA, Orton JM, Ellis CR. Feasibility of defibrillation and pacing without transvenous leads in a combined MICRA and S-ICD system following lead extraction. J Cardiovasc Electrophysiol 2017;28:233-4. 252. Ng JB, Chua K, Teo WS. Simultaneous leadless pacemaker and subcutaneous implantable cardioverter-defibrillator implantation—when vascular options have run out. J Arrhythmia 2018;35:136-8. 253. Lüker J Sultan A, Sreeram N, Brocknie M, Steven D. Implantation of a subcutaneous implantable cardioverter defibrillator with right parasternal electrode position in a patient with D-transposition of the great arteries and concomitant AAl pacemaker: a case report. Eur Heart J Case Rep 2018;27:09-09. 254. Kossidas K, Kalman R, Folis WP, Cooper JM. Managing cross talk between a subcutaneous implantable cardioverter-defibrillator and a dual-chamber unipolar pacemaker system. HeartRhythm Case Rep 2017;3:579-83. 255. Ip JE, Wu MS, Kennel PJ, Thomas G, Liu CF, Cheung JW et al. Eligibility of pacemaker patients for subcutaneous implantable cardioverter defibrillators. J Cardiovasc Electrophysiol 2017;28:244-8. 256. Sterliński M, Świerzyńska-Wodarska E, Zakrzewska-Koperska J, Guzek K, Drohomirecka A, Szunowski L. First experience in simultaneous use of the extravascular implantable cardioverter defibrillator and the leadless atrioventricular pacemaker. Kardiol Pol 2024;82:1295-6. 257. Michalak M, Ozeralski K, Grabowski M. Simultaneous use of extravascular implantable cardioverter defibrillator and left ventricular pacing through the coronary sinus in a patient with history of transcatheter edge-to-edge tricuspid valve repair. Kardiol Pol 2025;83:1098-100. doi: 10.33963/v.pnj.106329.

258. Chatterjee NA, Roka A, Lubitz SA, Gold MR, Daubert C, Linde C et al. Reduced appropriate implantable cardioverter-defibrillator therapy after cardiac resynchronization therapy-induced left ventricular function recovery: a meta-analysis and systematic review. Eur Heart J 2015;36:2780–9. 259. Manfredi JA, Al-Khatib SM, Shaw LK, Thomas L, Fogel RI, Padanilam B et al. Association between left ventricular ejection fraction post-cardiac resynchronization treatment and subsequent implantable cardioverter defibrillator therapy for sustained ventricular tachyrythmias. Circ Arrhyth Electrophysiol 2013;6:257–64. 260. Sidhu K, Castrini Al, Parikh V, Reza N, Owens A, Tremblay-Gravel M et al. The response to cardiac resynchronization therapy in LMNA cardiomyopathy. Eur J Heart Fail 2022;24:685–93. 261. Zegard A, Okafor O, Deboo J, Kalla M, Lencioni M, Marshall H et al. Grayzone myocardial fibrosis and ventricular arrhythmias in patients with a left ventricular ejection fraction greater than 35%. Europace 2021;23:euab116.363. 262. Ogano M, Iwasaki YK, Tsubol i, Kawanaka H, Tajiri M, Takagi H et al. Mid-term feasibility and safety of downgrade procedure from defibrillator to pacemaker with cardiac resynchronization therapy. JIC Heart Vasc 2019;22:78–81. 263. Frey SM, Brenner R, Theuns DA, Al-Shoaibi N, Crawley RJ, Ammann P et al. Follow-up of CRT-D patients downgraded to CRT-P at the time of generator exchange. Front Cardiovasc Med 2023;10:1217523. 264. Tit R, Boveda S, Deharo JC, Dobreanu D, Haugaa KH, Dagles N. Replacement of implantable cardioverter defibrillators and cardiac resynchronization therapy devices: results of the European Heart Rhythm Association survey. Europace 2016;18:945–9. 265. Woo GW, Shedd OL, Bouhussein N, Turek RD, Bergen KM. PO-06-036 A novel technique for downgrading A DF-4/IS-4 quadripolar left ventricular lead cardiac resynchronization defibrillator system to a quadripolar cardiac resynchronization pacemaker. Heart Rhythm 2025;22:5647–8. 266. Giedrimas A, Sisson B, Casavant D. A novel method to enable biventricular defibrillator to biventricular pacemaker downgrade involving DF4 defibrillator lead. HeartRhythm Case Rep 2018;4:598–600. 267. Monhouse C, Elliott J, Collinson J, Hunter R, Lambiase P, Ahsan S et al. DF-4 defibrillator downgrade to pacemaker, a novel method of device downgrade for implantable cardioverter-defibrillator patients. HeartRhythm Case Rep 2023;9:587–9. 268. Stühlinger M, Burri H, Vernooy K, García R, Lenarczyk R, Sultan A et al. EHRA consensus on prevention and management of interference due to medical procedures in patients with cardiac implantable electronic devices: for the European Heart Rhythm Association (EHRA). Heart Rhythm Society (HRS), Latin America Heart Rhythm Society (LAHRS), Asian Pacific Heart Rhythm Society (APHRS). Europace 2022;24:1512–37.

表 1
Introduction4
Overview of clinical data and guidelines4
Summary of potential procedural risk and benefit6
Effective workflow and approach to practice6
Optimizing initial choice of cardiac implantable electronic device6
Screening for eligible patients6
Timing of device upgrades or downgrades8
Shared decision-making8
Hardware considerations during decision-making9
Operator experience and need for specialist skills10
Pre-procedural steps—importance of procedural planning and contingency planning10
Value of pre-procedural cardiac imaging10
Determining the need for upgrade-assessment of left ventricular function10
Assessment of tricuspid regurgitation11
Additional utilization of imaging11
Cardiac computerized tomography11
Cardiac magnetic resonance imaging11
Upfront superior vascular anatomy assessment: the venogram11
Enaesthetic considerations11
Medication considerations11
Medication considerations12
表 2
Value of pre-procedural cardiac imaging10
Determining the need for upgrade-assessment of left ventricular function11
Assessment of tricuspid regurgitation11
Additional utilization of imaging11
Cardiac computerized tomography11
Cardiac magnetic resonance imaging11
Upfront superior vascular anatomy assessment: the venogram11
Anaesthetic considerations11
Medication considerations12
表 3
Anticoagulation and antiplatelets12
Equipment and lab set-up considerations12
Vascular superior access considerations14
Vascular access issues14
Techniques to manage venous stenoses/obstruction14
Vascular superior access considerations15
Preparation15
Puncture17
Access through an occluded vessel18
Serial dilatation18
Balloon venoplasty18
Infrequently used methods18
Contralateral new lead implantation and tunnelling/contralateral new system implant and abandonment of original system18
Lead extraction to create venous access or manage redundant leads19
Alternative (non-superior access) delivery approaches20
Surgical/epicardial approach20
Leadless, endocardial, and other non-transvenous approaches20
Femoral approach for lead placement21
Surgical considerations22
Consideration 1: timing of skin incision and pocket opening in relation to vascular access for cardiac implantable electronic device lead placement22
Consideration 2: location of incision23
Consideration 3: surgical techniques; blunt dissection vs. diathermy vs. low-thermal-injury dissection device23
表 4
Consideration 4: capulectomy23
Consideration 5: management of abandoned leads23
Consideration 6: infection prevention and antibiotic envelopes (Table 7)23
Antibiotic envelopes24
Patient perspective25
Considerations for individualized risk-benefit analysis26
Suggested framework27
Role of the multi-disciplinary team and ethical framework27
Risk at generator change vs. any other time27
Specific situations27
Device deactivation28
Distinct case scenario discussions29
Device downgrade considerations29
Downgrade of cardiac resynchronization therapy defibrillator to cardiac resynchronization therapy pacemaker34
Practical considerations34
Downgrade of implantable cardioverter defibrillator to pacemaker34
Practical considerations34
Downgrade from dual-chamber to single-chamber device36
When a cardiac implantable electronic device is no longer indicated36
Distinct patient scenarios38
Conclusion38
表 5
DefinitionCategories of advice
Evidence or general agreement that a given measure is clinically useful and appropriateAdvice TO DO
Evidence or general agreement that a given measure may be clinically useful and appropriateMay be appropriate TO DO
Evidence or general agreement that a given measure is not appropriate or harmfulAdvice NOT TO DO
No advice can be given because of lack of data or inconsistency of data. The topic is important to be addressedAreas of uncertainty
表 6
Type of supporting evidenceStrength of evidence
Published $data^{a}$>1 high-quality RCT Meta-analysis of high-quality RCT High-quality RCT > 1 moderate quality RCT Meta-analysis of moderate quality RCT High quality, large observational studies
Expert $opinion^{b,c}$Strong consensus >90% of writing group (WG) supports advice Consensus >70% of WG supports advice
表 7
Definition and delineation of the targetStrength of evidence
Advised TO DOAdvised TO DO
It is advised to, it is appropriate, it is useful>1 high-quality RCT Meta-analysis of high-quality RCT
It is advised to, it is appropriate, it is usefulHigh-quality RCT > 1 moderate quality RCT Meta-analysis of moderate quality RCT
It is advised to, it is appropriate, it is useful>90% of writing group agree
It is advised to, it is appropriate, it is useful>70% of writing group agree
May be appropriate TO DO
It may be appropriate, it may be usefulHigh quality, large observational studies
It may be appropriate, it may be useful>90% of writing group agree
Advised NOT TO DOAdvised NOT TO DO
It is not advised, it is not appropriate, it is harmful>90% of writing group agree
Areas of uncertaintyAreas of uncertainty
It is unknown, lack of data, inconsistency of data>90% of writing group agree
It is uncertain, lack of data, inconsistency of data>70% of writing group agree
表 8
Prior guidelineCurrent guideline
ESC (EHRA) Guidelines2013 Recommendation for BiV upgrade if LVEF < 35% despite OMT and significant proportion of RVP The ESC 2021 pacing guidelines highlight a knowledge gap in this area.$^{13}$ The ESC 2021 pacing guidelines may be updated prior to Budapest-CRT publications and recent meta-analyses of available data No comment is made on upgrade to conduction system pacing devices in the guideline, however, upgrade considerations are included in the 2025 ESC/EHRA consensus on CSP indications$^{4}$2021 Recommendation for BiV upgrade if LVEF < 35% despite OMT and significant proportion of RVP The ESC 2021 pacing guidelines may be updated prior to Budapest-CRT publications and recent meta-analyses of available data No comment is made on upgrade to conduction system pacing devices in the guideline, however, upgrade considerations are included in the 2025 ESC/EHRA consensus on CSP indications$^{4}$
HRS/LAHRS/ APHRS Guidelines2012 No formal recommendation regarding device upgrade. However, the guideline 'recognized the ability to improve LV dysfunction in the setting of chronic RV pacing with an upgrade to a biventricular device'$^{10}$2023 Recommendation for device upgrade to biventricular pacing. Recommendation for upgrade to CSP.$^{14}$ Relevant patients are defined as those with substantial RVP and any decline in left ventricular function or symptoms noted, suggesting that device upgrade should not necessarily be restricted to only those with severely impaired ejection fraction
Canadian HRS Guidelines2014 Weak Recommendation CRT may be considered for patients with chronic RV pacing or who are likely to be chronically paced, have signs or symptoms of heart failure and a LVEF ≤ 35%(No guideline update since 2014)
表 10
General considerationsStrength
Advised to do
In patients who have received a conventional pacemaker or an ICD and who develop symptomatic HF with LVEF ≤ 35% despite OMT and who have a significant (>20%) proportion of RV pacing, upgrade to CRT is advised$^{16-19,22,27,57}$>1 moderate quality RCT or Meta-analysis of moderate quality RCT
In patients who have received a conventional pacemaker and who develop an ICD indication, ICD upgrade is advised$^{27}$>1 moderate quality RCT or Meta-analysis of moderate quality RCT
An integrated care approach is advised in pacemaker and ICD patients when considering upgrades and downgrades to ensure a patient-centred approach with patient involvement and shared decision-making>90% of writing group agree
When HFrEF is detected in the presence of high RV pacing burden, it is advised to discuss an upgrade strategy utilizing a multi-disciplinary team approach, including consideration of prognosis and risk benefit analysis. Discussion should include discussion of upgrade>90% of writing group agree
to CRT with or without ICD capability
It is advised that device upgrade procedures are performed by experienced operators in experienced centres (at least 20 upgrade procedures/year)>90% of writing group agree
May be appropriate to do
In patients who have received a conventional pacemaker or an ICD, and who subsequently develop symptomatic HF with either LVEF 36-40%, or a significant (~ 10%) decrease in LVEF from baseline despite OMT, and who have a significant (>20%) proportion of RV pacing, it may be appropriate to upgrade to CRT$^{16}$,$^{20}$High quality, large observational studies
Areas of uncertainty
It is uncertain whether patients who suffer from PICM alone, with no other substrate identified for LV impairment, benefit from upgrade to CRT plus a defibrillator, or rather just upgrade to CRT strategies>90% of writing group agree
表 11
Patient selection and assessment for upgrade indications and generic considerationsStrength
Advised to do
At all planned device follow-up visits (by physicians/cardiac physiologists/ technicians/device nurses), assessment as to whether a patient has the most appropriate device implanted is advised. This should include review of symptoms, hospitalizations, and pacemaker-derived data review. This assessment may trigger further initial investigations with echocardiography and/or NT-Pro-BNP>90% of writing group agree
When patients with a ventricular pacing burden >20% are approaching ERI (<12 months), evaluation of symptoms of heart failure and left ventricular function is advised>90% of writing group agree
When LVEF deterioration is confirmed, investigation of secondary causes, including arrhythmias, coronary atherosclerosis, lack of GDMT etc. and involvement of HF specialists or invasive cardiologists as needed is advised>90% of writing group agree
A shared decision-making process is crucial before the upgrade procedure considering patients' preferences and taking into consideration frailty and comorbidity burden. The multi-disciplinary team may play a key role in this>90% of writing group agree
When lead extraction is being considered, it is advised to involve a multi-disciplinary team including the CIED specialist, extraction specialist, anaesthesiologists, cardiac surgeons, geriatricians, and infectious diseases specialists where appropriate>90% of writing group agree
May be appropriate to do
It may be appropriate to perform device upgrade at the time patient symptoms or LV function deterioration is noted and not defer till time of elective generator replacement$^{27,112}$>1 moderate quality RCT or Meta-analysis of moderate quality RCT
表 13
Pre-procedural considerationsStrength
Advised to do
Before an upgrade procedure, it is advised that procedural steps and strategy are planned, including what steps (for example venous access strategies or extraction) would and would not be used for a given patient balancing net risk against net benefit>90% of writing group agree
Two-dimensional echocardiography is advised as the primary imaging technique to evaluate patients who may be candidates for an upgrade to CRT>90% of writing group agree
It is advisable to have a recent echocardiogram before replacing a device generator, to assess LVEF in those with>90% of writing group agree
ventricular pacing >20% and/or with heart failure symptoms (preferably done within 12 months of the proposed generator replacement)
Before a generator replacement, it is advised to evaluate patient's condition for signs and symptoms of heart failure and to check the device for ventricular pacing burden as well as atrial and ventricular arrhythmias>90% of writing group agree
Where the risk of bleeding significantly outweighs the risk of stroke or embolic complication, it is advised to hold anticoagulation for 24-48 h prior to device upgrade procedures and resume after 24-48 h.>90% of writing group agree
Where risk of bleeding outweighs the risk of coronary complications (i.e. stent thrombosis), it is advised to withhold DAPT for an appropriate duration prior to procedure (3 days for ticagrelor, 5 days for clopidogrel, and 7 days for prasugrel) whilst continuing on with aspirin and resume the second antiplatelet agent after 24-48 h>90% of writing group agree
Venography should be performed before skin incision. where possible, it is advised to undertake this in advance of a planned upgrade procedure date to enable procedural planning>90% of writing group agree
表 15
Vascular Access optionsAdvantageDisadvantageFor whom?
Superior vascular access options
Medial subclavian vein puncture$^{126-128}$Ipsilateral vascular access Preservation of contralateral vascular access Relatively easy techniqueIncrease in risk of pneumothorax and haemothorax Risk of long-term lead damage (subclavian crush) impacting on lead performance Success rate variable depending on the occlusion (length, complete occlusion)Favourable anatomy of the venous stenosis or occlusion (puncture distally) TLE and venoplasty not appropriate or available
'Wire under the insulation' technique$^{129}$ (figure demonstrating technique included in reference)Preserves vascular access Avoids repeated venous cannulation Minimally invasiveHighly challenging in patients with extensive venous thromboses or occlusion (therefore less useful in leads with longer dwell time) Success rate variable with risk of lead damagePatients with limited vascular access Relatively short lead dwell time Patients at high risk of complication from repeat venous cannulation (i.e. those with COPD or frailty)
Serial dilation or venous venoplasty$^{85,86}$Ipsilateral vascular access Preservation of contralateral venous accessProcedure duration potentially increased Potential damage to remaining leads Multi-disciplinary approach occasionally needed (interventional radiology support) but mostly can be done by cardiologist alone Training/experience requiredAll CIED upgrades with a suitable pre-operative venogram Upgrade in patients with <4 leads on one side and <5 through the SVC
Transvenous lead extraction (TLE)$^{130-132}$Ipsilateral vascular access Preservation of contralateral venous access Lead management in young patients i.e. minimises total number of leads MRI conditional systemProcedure duration potentially increased Additional risks including SVC tear, haemothorax, tricuspid valve damage, tamponade, death. Collateral damage (TLE of all leads instead of only a redundant lead due to inter-lead fibrosis), possible extraction of functional leads Availability of expertise and centre capability Femoral access requirement Cardiothoracic surgery backup generally required Availability of general anaesthesiaYoung age and long-term need for functional CIED therapy and desire to avoid an abandoned lead Upgrade in patients with venous occlusion + >1 dysfunctional lead or >1 functional lead to be replaced Single (non-ICD) lead for extraction, lead dwell time <10 years, no coronary sinus lead (risk of inadvertent extraction along with other leads) [lower risk] MRI conditional system required
Contralateral approach$^{133}$ Contralateral implantation of a new system Contralateral lead implantation plus tunnelling to initial siteNo complications associated with venous intervention or lead extraction Increased procedure duration Increased riskContralateral approach with both vascular access sides used Risk of bilateral subclavian vein occlusion Increased number of leads through the SVC (potentially higher risk of SVC syndrome) Future lead revisions are more challenging Potential risk of lead-lead abrasion or interaction Further compromise of venous access If total system extraction required secondary to infection, future device implantation options may be limited due to infection involving both right- and left-sided access sites True MRI-conditionality lost contralateral tunnelling carries risk of cutaneous lead perforationOlder patients Frail patients High-risk associated with TLE (low body-mass index (BMI) < 25 kg/m$^{2}$, severe LV dysfunction, renal dysfunction, complete venous occlusion, leads >10 y/o, high-risk leads)$^{134}$ Low total number of leads (right and left; ≤4, ≤1 ICD lead in total) Non-ICD lead (for contralateral tunnelling)
Epicardial/surgical lead $implantation^{135-137}$Lead can be readily placed if patient undergoing cardiac surgery for alternate reasonThoracotomy Additional risks of surgery Additional cost Suboptimal lead performance Difficult removal or revision Non-MRI conditional systemCRT upgrade after failed endocardial lead positioning and failed CSP For patients undergoing surgery for other reasons
Utilization of a non-transvenous approach (e.g. LLP, S-ICD, EV-ICD, WISE CRT)$^{138-156}$No lead associated problems in LLP Risk of CIED related endocarditis is lowerOnly available for selected patients depending on the planned up- or downgrade procedure Unclear future management of LLPs in patients <40 y/o (limited long-term data available at the current time) Additional cost CSP LLP not readily available at this time Only a limited option in CRTPatients at a high risk of CIED infection (e.g. haemodialysis, prior CIED infection, immunosuppression) ATP not required (S-ICD) Bilateral venous occlusion Patients at high risk for TLE
Other, rarely used strategies to overcome central venous occlusion (non-superior access)
Inside-out central venous access (IOCVA)$^{157}$Ipsilateral vascular access Classical pectoral placement of pacing and/ or defibrillation leads Applicable for chronic total SVC and subclavian vein occlusionsHigher potential risk of pneumothorax and haemothorax Limitations in patients with scarred or calcified subcutaneous tissue No experience with lead revision or TLE Unknown risk to existing leads Complex intervention for experienced operatorsBailout strategy for very few patients
Femoral (or trans-iliac) approach with abdominal or thigh pulse generator $placement^{158-162}$Pacing option in patients with central venous occlusion (predominately in the era prior to LLP)Unstable leads Risk of lead dislodgement, especially atrial lead Additional subcutaneous lead in defibrillator systems to overcome the abdominal pulse generator location Risk of deep vein thrombosis and pulmonary embolism Increased risk of procedural complications (infection, lead fracture, intestinal/vascular injury)Bailout strategy for very few patients, primarily for implantation of a new CIED in patients with complete central venous occlusion (particularly relevant now if LLP is not available/suitable)
表 17
Vascular access considerationsStrength
Advised to do
Superior (cephalic, subclavian, and axillary) vascular access strategies are advised over non-superior access strategies (i.e. femoral transvenous implantation)>90% of writing group agree
Good quality contrast venography from the ipsilateral arm is advised prior to skin incision to assess the patency of the vein to help guide the procedure>90% of writing group agree
Venous puncture is advised in an extra-thoracic location rather than using more medial punctures to avoid risk of pneumothorax and potential future lead failures$^{127}$High quality, large observational studies
May be appropriate to do
When possible, venography may be useful to be performed in advance of the planned upgrade date to help risk-stratify and plan an upgrade approach>90% of writing group agree
It may be appropriate that centres undertaking device upgrades have access to operators with skills to perform multiple techniques to cross stenoses, including venoplasty>90% of writing group agree
Venoplasty, where possible, may be appropriate to facilitate venous access and enable device upgrade in case of venous obstruction>90% of writing group agree
Lead extraction may be appropriate to obtain venous access and enable device upgrade in case of venous obstruction>90% of writing group agree
Where superior venous access is challenging, alternative approaches may be appropriate, including S-ICD/EV-ICD, leadless pacing, and epicardial/surgical systems>90% of writing group agree
Contralateral lead implantation and tunnelling may be an appropriate approach to add leads>90% of writing group agree
Advised not to do
Placement of more than 5 leads in the SVC and 4 from a single subclavian vein is not advised agree>90% of writing group agree
Ultrasound is not advised as a first choice/ optimal technique for assessment of venous patency to plan procedural vasculature access>90% of writing group agree
Areas of uncertainty
It is uncertain in which patients extraction of abandoned leads should be utilized when extraction is not required to facilitate vascular access>90% of writing group agree
表 19
InterventionStudy designnMain results
Pre-procedural antibiotics$^{223}$RCT649Trial stopped early by the safety committee due to a significant reduction in the risk of CIED-related infection (0.64% in the cefazolin group and 3.28% in the placebo group; P = 0.016)
Chlorhexidine vs. iodine-povidone for skin $preparation^{195}$RCT849Chlorhexidine resulted in a lower risk of surgical site infection, particularly superficial (4.2% vs. 8.6%, P = 0.008) and deep incisional infections (1% vs. 3%, P = 0.05) and deep incision infections in patients undergoing clean-contaminated surgeries Although not directly tested in CIED-related procedures, chlorhexidine may be the preferred choice in CIED upgrades
Antibacterial envelope$^{203}$RCT6983Compared to standard of care, the use of an antibacterial envelope resulted in a significant reduction in CIED-related infections (hazard ratio, 0.60; 95% CI, 0.36–0.98; P = 0.04)
Adherent iodine drape$^{196}$OBS14 225Use of an adherent skin drape was associated with a lower risk of CIED-related infection (OR: 0.32; 95% CI 0.154–0.665; P = 0.002) Possible differences in the baseline characteristics between groups (i.e. drape vs. no drape) are not ascertained
Chlorhexidine vs. normal saline lavage$^{198}$OBS1504Chlorhexidine lavage was associated with a significant reduction in CIED-related infection (hazard ratio 0.138; 95% CI 0.04–0.45; P = 0.001)
表 20
Surgical considerations including infection risk reductionStrength
Advised to do
Utilization of an antibiotic envelope is advised inHigh quality, large observational studies
patients at higher risk of infection$^{109,203,205,212}$
May be appropriate to do
Utilization of low-temperature electrocautery techniques may be appropriate over regular diathermy and blunt dissection where available>70% of writing group agree
The cap & bury approach may be an appropriate management approach for abandoned leads. The insulation pull-back approach may be an appropriate management approach for abandoned leads>90% of writing group agree >70% of writing group agree
The insulation pull-back approach may be an appropriate management approach for abandoned leads>90% of writing group agree
It may be appropriate to coil and secure abandoned leads within the pocket to reduce risk of interaction with other pacing components and erosion>90% of writing group agree
Advised not to do
Capsleutomy is not routinely advised due to risk of possible resultant haematoma>70% of writing group agree
Areas of uncertainty
It is unknown whether during an upgrade procedure (without a need for generator replacement), if there is a benefit to obtaining vascular access prior to skin incision and opening the pocket, or whether alternatively, making a skin incision and opening the pocket is a preferable first procedural step>90% of writing group agree
表 22
Patient FactorsSystem FactorsProcedural Factors
- Age and Frailty- Team & Centre experience- Procedure type/complexity
- Comorbid state- Operator experience- Existing and proposed device type
(Heart Failure, Renal failure, Diabetes, Bleeding diathesis, Malignancy, Cognitive state)- Availability of tools and kit- Number and type of leads already implanted
- Prior device history- Number and type of leads to implant
- Infection history and risk- Techniques available to use
- Need for device procedure e.g. ERI
表 23
Individualized risk-benefit analysisStrength
Advised to do
Individual patient risk-benefit analysis before the upgrade/downgrade procedure is advised>90% of writing group agree
In patients considered for CIED up- or downgrade, it is advised that risks and benefits of the procedure, the patient's preferences, and goals of care are explored in a shared decision-making process>90% of writing group agree
It is advised that frailty, comorbidity burden, cognitive status as well as functional status, care requirements, and support network are assessed by the multi-disciplinary team with the involvement of caregivers to help guide device upgrade or downgrade decisions>90% of writing group agree
If a device is deactivated/made non-functional, it is advised that the procedural risks of device removal and the risks associated with leaving the device in situ are assessed individually>90% of writing group agree
May be appropriate to do
It may be appropriate to prepare patients for potential decisions relating to their device early on in their device lifetime. This may help patients prepare for decisions relating to future upgrade, downgrade, or deactivation of their CIED>90% of writing group agree
Areas of uncertainty
It is unknown whether GDMT should be optimized first in the treatment of isolated pacing-induced cardiomyopathy (before proceeding to device upgrade) or whether earlier device upgrade should be performed potentially reducing the need for medication utilization>90% of writing group agree
表 25
CIED upgradeStrength
Advised to do
It is advised that when transvenous or leadless pacemakers are combined with subcutaneous or extravascular defibrillators, possible interactions between the systems are considered. Evaluation before and during implant should aim to reduce>90% of writing group agree
risk of inappropriate defibrillation delivery or inappropriate inhibition of pacing
When a CIED with an RV pacemaker lead is upgraded to a system with a transvenous defibrillator lead or to a conduction system pacing lead, considering the pros and cons of extraction over abandonment of the original RV lead is advised>90% of writing group agree
It is advised when upgrading an existing pacemaker to a CSP system (no defibrillator) to avoid abandoning leads, and the already implanted RV lead can be incorporated into a CRT-P generator>90% of writing group agree
May be appropriate to do
It may be useful where a defibrillator lead is to be added, and there is a short lead dwell time of the original RV lead and/or younger age of the patient, that a lead extraction strategy is preferred over lead abandonment>90% of writing group agree
表 27
Existing CIEDUpgradeManagement strategiesKey points
Leaded TV CIEDSingle-chamber to dual-chamber PPM (upgrade AAI or VVI PPM to dual-chamber PPM)If VVI system add a transvenous RA lead and use the existing RV lead for a DDD pacemaker If VDD system abandon the RA sensing port and add a transvenous RA lead. This may impact MRI access in some centres If AAI system add a transvenous RV lead If single-chamber PPM, dual-chamber leadless pacemaker (Aveir DR system, Abbott, USA), or Aveir single-chamber device to co-exist with leaded CIED to provide dual-chamber system or AV Micra® (Medtronic, USA) can all be utilized if vascular access issues preclude implantation of new atrial or ventricular leadIf initial lead is a VDD lead, 231-233 could consider lead extraction and replacement with a standard RV bipolar lead for example to achieve formal MRI-conditionality, but this would not be a common first-line approach in non-infected systems
Leaded TV CIEDUpgrade from single or dual chamber PPM to CRT-P/CSP pacing strategyAddition of either an LV lead to an RV pacing system or addition of a CSP lead to an RV pacing system Where a CSP lead is added, consideration can be given whether to incorporate the original RV lead within the new pacing approach or whether to abandon the original RV lead. If original system is VVI, it could incorporate CSP upgrade within a dual-chamber device (utilizing the atrial port) or a CRT-P device with a plugged atrial portUpgrading to a physiological pacing strategy may be relevant in patients with pacing-induced cardiomyopathy or for those with broad LBBB If upgrade to CSP is performed, both the existing RV lead and CSP can be incorporated into the new device (typically CRT-P generator). Whilst a CRT-P costs a little more than a DR PPM, this cost is frequently not prohibitive. This may be useful rather than abandoning leads The decision to abandon leads is patient-specific and will take into account individual considerations, including the risk of extraction and the patient's vasculature In the case of pacing-induced cardiomyopathy resulting in severe LV impairment (no other cause for LV deterioration identified), the added net value of a defibrillator is not known and is an area of uncertainty. The LV function may well improve with improvement in cardiac activation from change in pacing approach. Retrospective cohort studies suggest >70% of patients with PICM will improve within 12 months to have an LVEF > 35%. 234 In patients with EF <35% and with significant comorbidity and increased frailty, upgrade to a pacing only solution may be more appropriate than upgrade to CRT-D Considerations include the need for ATP, likelihood of lead-induced or worsening of TR and its impact, need for physiological pacing, and patient factors (age, comorbidity, etc.) A key consideration in this scenario is often whether to abandon the existing RV lead or extract it. When adding two leads, the RV pacing lead can be abandoned or extracted, but in younger patients, particularly where existing leads have only been in a short duration, it may be preferable to consider extraction of the RV pacing lead prior to ICD/CRT-D upgrade to reduce the
Leaded TV CIEDUpgrade from PPM/CRT-P to ICD or to CRT-DAddition of a standard RV-ICD lead is the usual solution, but it will result in at least two leads traversing the tricuspid valve that could induce or worsen TR If a physiological pacing strategy is needed, addition of two leads will likely be required to a standard pacemaker: either a lead in the coronary sinus or at the conduction system plus the standard RV defibrillation lead In younger patients (particularly with pacing leads that have only been in for a short duration), it may be preferable to consider extraction of the RV pacing lead prior to ICD/CRT-D upgrade to reduce theConsiderations include the need for ATP, likelihood of lead-induced or worsening of TR and its impact, need for physiological pacing, and patient factors (age, comorbidity, etc.) A key consideration in this scenario is often whether to abandon the existing RV lead or extract it. When adding two leads, the RV pacing lead can be abandoned or extracted, but in younger patients, particularly where existing leads have only been in a short time, greater consideration should be given to extraction. 238 Dedicated ICD leads that are able to provide CSP are being trialled and may be an option to reduce
number of leads in the heart as part of a lead management strategy If vascular access is challenging, an S-ICD can be used in conjunction with the existing pacing device, particularly in those who do not need ATP. This is a reasonable solution for primary prevention indications. There is evidence of efficacy and no increase in inappropriate shock.$^{235}$ EV-ICD is a possible alternative if ATP is required, but long-term data remain limited.$^{149}$ In both cases, the patients will need two devices, and MRI non-compatibility may be a concern. ICD need only: If there is patent venous access, implant RV bipolar ICD lead at alternative site within the ventricle. Subsequently, abandon or extract the existing RV lead. This will be the most common strategy. Implant a subcutaneous ICD as a co-implant if anticipated low ATP requirements$^{235}$ Add an EV-ICD with ATP potential$^{149,236,237}$ ICD + CRT (BiV or CSP) need Implant a standard CRT-D, either abandon or extract the RV pacing lead (an RV defibrillator lead and LV/ CSP lead are added)additional lead requirements contingent on ongoing trial results. Additionally, the role of repurposing existing ICD leads for deployment at the left bundle location is being investigated too.$^{239}$ If an ICD is added without extracting the RV pacing lead, it may be worth considering whether the new ICD lead should be a dedicated bipolar lead rather than an integrated bipolar lead to avoid potential problems with sensing
表 29
Existing CIEDUpgrade toManagement strategiesKey points
Leadless pacemaker (LLP)Leaded CIED (Pacemaker or Defibrillator)Implant of transvenous ICD/pacemaker (single, dual, or CRT system) and deactivation of leadless pacemaker Implant of transvenous ICD/pacemaker (single, dual, or CRT system) and removal of the leadless pacemakerRemoving a leadless pacemaker when converting to a leaded system reduces the implanted material inside the heart—the benefit of this is, however, unknown. Removing LLPs poses challenges related to procedural complexity, risks, and costs. If the LLP is a Micra device (Medtronic, USA) and there is only one leadless device in the heart, it is often preferred in clinical practice to abandon the device (particularly, if longer dwell time and provided there is no evidence of LLP-related infection). Micra devices (Medtronic, USA) do not have a dedicated mechanism for retrieval, but can still be removed with a snare with high success rates.$^{240,241}$ However, should the device be an Abbott NanoStim or Aveir (Abbott, USA) leadless pacemaker, these devices have a dedicated docking button for retrieval, and studies report a ~90% retrieval success rate.$^{242,243}$ Evidence of infection would necessitate explant/extraction of the device
Existing CIEDLLP + S-ICD or EV-ICDImplant of S-ICD or EV-ICD system in addition to ongoing pacing provision from leadless $device^{145,244,245}$Consider the possible interactions between LLP and ICD. Broad RV paced QRS morphology can potentially lead to double counting of signals and therefore possible inappropriate shocks Intraoperative testing for interactions (ICD waveform screening during ventricular pacing and sensing with possible lead placement revision to ensure appropriate sensing) should be taken into account. For EV-ICD, caution is required if the LLP is an atrial pacing device, as atrial pacing may lead to inappropriate sensing, which needs to be evaluated at implant if being considered. VF induction is advised in these cases to confirm satisfactory function of both devices
Existing CIEDDual-chamber leadlessAddition of leadless atrial device as part of Dual-Chamber AVEIR® system (Abbott, USA)$^{151}$Upgrading with an atrial leadless pacemaker may alleviate symptoms from leadless pacemaker syndrome. Success rates for deployment reported in a 36-patient series were 91.4% with complication rates reported to be 13.9%.$^{246-248}$
表 31
Existing CIEDUpgradeManagement strategiesKey points
S-ICD (Boston Scientific, USA)S-ICD + LLPFor pacing needs, combining an S-ICD with an LLP may be preferred, particularly if elevated risk of CIED infection. This combination is more frequently encountered in complex congenital heart disease and/or if vascular access is limited/challenging$^{244,249-252}$During LLP deployment, S-ICD waveform screening should be performed for reasons detailed above. Consider repositioning of the LLP in case of S-ICD sensing concerns$^{244,245}$ Confirm appropriate S-ICD sensing during VOO at high pacing output.$^{244,245}$ Consider modular cardiac rhythm management (EMPOWER), when commercially available, as this will provide ATP functionality
S-ICD + leaded CIEDFor pacing indications, conventional transvenous leads are commonly used. However, if there are vascular access issues, an epicardial pacing system is a potential option. If ATP functionality is desired, a transvenous ICD lead can be implanted. In this scenario, the S-ICD should be removed If no ATP requirement but only Brady or CRT requirements, a pacemaker or CRT-P could be implanted and the S-ICD left in situ. If vascular issues are not present, it may be appropriate to implant an entire TV system and remove the S-ICD. Combining a TV CIED and an S-ICD may be used when vascular access is not sufficient for multiple lead implantationIntraoperative S-ICD waveform screening during maximal voltage output and inhibition should be performed for reasons detailed above.$^{253}$ Set the pacemaker in VOO mode at maximum output to watch for oversensing and also for VF under-sensing during defibrillation threshold testing in patients (i.e. test the 'worst-case scenario')$^{254}$ Consider programming the pacemaker upper rate to <50% of the S-ICD tachycardia zone (avoids inappropriate therapy due to double detection) if oversensing appears to be a concern Obligate bipolar pacing output to avoid inappropriate double detection of pacing stimulus and evoked response consider pacing approaches that deliver narrower QRS morphologies to reduce double counting risk$^{255}$
Replace with EV-ICDThis could be an option if ATP functionality is desired, no anti-bradycardia pacing indication and there is no venous access for a TV-CIEDTechnology is in relative infancy. The pivotal study (n = 356 enrolled) has reported high defibrillation success rates and low complication rates.$^{149}$ More data are needed on this technology
EV-ICDEV-ICD + LLPFor pacing needs combining an EV-ICD with an LLP may be preferred, particularly if elevated risk of CIED infection or vascular is limited/challengingTechnology is in relative infancy. Atrial pacing may lead to inappropriate sensing and interaction will need to be tested at implant if being considered. Experience is limited to date$^{148,256}$
EV-ICD + leaded CIEDFor pacing indications, conventional transvenous leads are commonly used. However, if there are vascular access issues, an epicardial pacing system is a potential option. If only Brady or CRT requirements, a pacemaker or CRT-P could be implanted, and the EV-ICD left in situ. If vascular issues are not present, it may be appropriate to implant an entire TV system and remove the EV-ICD. Combining a TV CIED and an EV-ICD may be used when vascular access is not sufficient for multiple lead implantationTechnology is in relative infancy. Atrial pacing may lead to inappropriate sensing and interaction will need to be tested at implant if being considered. Experience is limited to $date^{148,257}$
Replace with S-ICDThis could be an option if vascular access is limited/challenging and an alternative defibrillation strategy is desiredMaintains a non-transvenous and extravascular system
表 33
Patient scenarioManagement strategies and optionsKey points
A 63-year-old man with ischaemic cardiomyopathy and a dual-chamber ICD implanted 10 years ago for primary prevention (LVEF 30%). He develops VT storm and is treated with Amiodarone, which increases his ventricular pacing percentage to >20%. He presents with worsening heart failure and a deterioration of his LVEF to 20%. He is referred for an upgrade to a CRT-D. His venogram reveals his axillary vein is occludedGiven the high prevalence of venous stenosis in patients with existing CIED leads, a peripheral venogram was performed, which confirmed axillary vein occlusion. Management strategies were discussed, including medial subclavian puncture, attempts at re-establishing vascular patency via ipsilateral venoplasty or lead extraction, vs. lead abandonment and re-implant of a new CRT-D on the contralateral side, or implantation of upgrade lead on contralateral side and tunnelling to existing pocket The patient opted to attempt re-establishing vascular patency first by venoplasty, and if that failed, lead extraction and re-implantation would have been used to facilitate completion of the upgrade procedure at the same sitting to avoid multiple procedures increasing risk of infection A successful upgrade procedure was performed after venoplasty with addition of an IS4 LV leadA consideration in this case is his relatively young age and possibility of risks related to lead abandonment if an entire de novo system were to be implanted on the contralateral side attempts at re-establishing vascular patency with venoplasty are relatively low risk and frequently successful even when veins appear occluded. Regaining vascular access via lead extraction may have a higher risk than venoplasty, but it is an effective alternative if venoplasty fails. Lead abandonment and potential future risk of lead extraction, which might increase with higher lead burden and dwell time (for example in the scenario of if abandoned leads become involved in a later infected device)
An 82-year-old woman with permanent atrial fibrillation, renal impairment, and diabetes. VDD pacemaker implanted 9 years ago as part of an ablate and pace strategy. Inhibition of right ventricular pacing was noted due to the oversensing of myopotentials with ipsilateral arm movements Her venogram revealed chronic ipsilateral left subclavian and left brachiocephalic vein occlusion. No evidenceManagement options discussed included attempting venoplasty to gain access for a new lead, extraction and replacement of ventricular lead, abandonment and addition of a new contralateral ventricular lead and device, or abandonment and implantation of a leadless pacemaker Ipsilateral venous occlusion would require contralateral lead placement, ipsilateral veinKey considerations were lack of evidence of PICM, preserved LV function, advanced age and patient preferences A leadless pacemaker reduces the likelihood of future CIED infections Only one abandoned pacing lead. Likelihood of future CIED interventions (the patient may not require further pacemaker replacements or interventions due to her life
of LV impairment or PICM was apparent even with her high burden of RV pacingvenoplasty, or lead extraction and re-implant of a new lead through the occluded vein. Considerations included the patient's preference for a simpler procedure with lowest expected complications. The decision was to abandon the malfunctioning lead and to implant a VVVIR leadless pacemaker. The decision also included to remove the old pacemaker generator to avoid future risks from interactions with the new leadless pacemaker, for example, automatic programming in VOO at ERI or EOLexpectancy, thus decreasing further the possibility of CIED infection) It is important to note that the whole transvenous system could have been abandoned, potentially reducing procedural complication, but this would be more relevant if the original device could be programmed 'OOO' thus mitigating risk of device interaction
An 81-year-old woman with non-ischaemic cardiomyopathy had an ICD implanted 13 years ago (primary prevention). Dual coil and passive fixation ICD lead. No pacing indication During follow-up and on guideline-recommended heart failure medical treatment, the LVEF increased to 50%. No ventricular arrhythmias were observed during all follow-up visits. Frequent episodes of myopotential oversensing were observed during remote monitoringManagement options discussed included abandoning lead and generator, removing generator and abandoning the ICD lead, or extracting the ICD lead and removing the generator Reappraise ICD indication (primary prevention, no ICD therapies, LVEF improvement) Considerations included the patient's preference not to have a prolonged hospitalization/ recovery period and not wanting a redundant generator in the pocket The patient did not want to undergo extraction. At her request, the decision was made to remove the generator and abandon the lead. No new ICD was implantedCurrently, no indication for ICD implantation. Considerations are: Age of the patient The type of ICD lead (dual-coil and passive fixation), as well as long dwell time, increases the risk of lead extractionAbandoned leads are only a relative contraindication for MRIThe patient remains at ongoing risk for lead infection, which, if occurred, would require a higher-risk extraction in the futureOpening the pocket to remove the generator may have exposed the patient to a risk of infectionHowever, leaving the ICD in situ could leave the patient at risk (likely low) of erratic EOL behaviourConsideration could have been given to abandoning the whole system, reducing procedural complication, and preserving MRI labelling
A 79-year-old woman with dual-chamber PPM for intermittent complete heart block. Recent generator replacement (leads 12 years old). Recent mitral valve replacement and tricuspid valve annuloplasty for severe MR and TR. Post-operative LBBB development. TR remains moderate to severe despite intervention and in part this is due to splinting of valve by existing RV pacing lead. LVEF has reduced to 23% post-surgeryDecision regarding upgrade of device to CRT-P vs. D and decision regarding upgrade to biventricular pacing vs. CSP Despite a recent generator replacement, it was appropriate to proceed quickly to device upgrade and not defer. Given her age, comorbidity, non-ischaemic aetiology, absence of any ventricular arrhythmia history and her elicited values with shared decision-making, upgrade to CRT-P with an LV lead only using venoplasty to gain adequate venous access was chosenTo avoid additional leads traversing the tricuspid valve the decision was made to not implant an ICD lead nor for a CSP lead. In view of her low BMI, sarcopenia and frailty an ICD was not favoured to avoid a large generator. This was the favoured approach particularly as ventricular arrhythmia risk was felt to be low. Evidence of benefit of LV leads alone in patients with non-ischaemic cardiomyopathy and broad LBBB is significant At the time of cardiac surgery, epicardial LV lead implantation could have been a potential alternative
表 35
CIED downgradeStrength
Advised to do
At the time of elective generator replacement, it is advised to undertake an evaluation of the ongoing need for defibrillator therapy. Consideration to potential device downgrade—such as from CRT-D to CRT-P, from ICD to a pacemaker, or deactivation or explant of the ICD may be appropriate. Factors to prioritize to guide the decision include patient frailty, comorbidities, life expectancy, and informed patient preference. All considerations should be integrated into a shared decision-making process tailored to the patient's overall clinical context and goals of care$^{262,263}$>90% of writing group agree
When pacing is no longer indicated, it is advised that the management strategy is based on an individual risk—benefit analysis in a shared decision-making process together with the patient and caregivers>90% of writing group agree
May be appropriate to do
At generator change in a patient with a dual-chamber pacemaker or ICD who has developed permanent atrial fibrillation, replacement with a dual-chamber device to retain MRI-conditionality may be appropriate (rather than utilizing a single-chamber device and abandoning the atrial lead)>90% of writing group agree
Areas of Uncertainty
In patients with primary prevention ICDs who have not previously had device therapies, the optimal approach to determine the arrhythmic risk at the time of generator replacement is uncertain. This risk stratification might inform who can be safely downgraded.>70% of writing group agree
The safety of downgrading CRT-D to CRT-P in patients with secondary prevention indications with recovered LVEF (>40-45%), without appropriate ICD therapy, remains uncertain>90% of writing group agree
The prognostic significance of ventricular arrhythmias occurring only early after CRT-D implantation, before recovery of LVEF has resulted (which may constitute a blanking period) remains uncertain>90% of writing group agree
图示:原图识别中
图示:Cardiac Implantable Electronic Device Upgrades and DowngradesUpgrades include adding ICD capability and/or delivering a distinct pacing strategyPatient's clinical stateand requirements canDowngrades are typically considered when ICD therapy is no longer appropriate.change over time~nMArrhythmia risk ?LBBB ?RV pacing (>20%)?LV Ejection Fraction√≤35%Echocardiography for LVEFIdentifying potentiaAssess utility oupgrade patients(√)36-40%RV pacing(NT-Pro BNPavoidance algorithmCan be used to rationalize imaging if required*>10% decrease from baselineEvaluate forMulti-disciplinary team approach for decision-making, involving where needed:Alternative causesElectrophysiologist / Device Specialist / Extraction Specialist / HF Specialist / Geriatrician / Heart Surgeon / Anesthetistfor LVEF reduction(& relevant therapy)● Patient frailty and comorbidityConsiderations indecision making● Forthcoming need for device procedure i.e elective generator replacemen● Upfront venogram and procedural planning with multi-disciplinary teamprocess● Risk / benefit of extraction vs lead abandonmen● Procedural risk relative to future harm from ongoing dyssynchrony or arrhythmia
图示:|Upgrades include:| | |:---:|:---:| |Dual-chamber PPM/ICD (+ A lead)|ICD (+ICD lead/S-ICD/EV-ICD)| |CRT-P (+ LV lead or CSP)|CRT-D (+ICD and LV or CSP lead)| |Leadless pacing (+ Atrial / Ventricular leadless pacemaker)| | |Additional Considerations:| | |Patient Frailty and Comorbidity| | |Risk / Benefit of Extraction vs. Lead abandonment| | |Forcoming need for device procedure i.e. generator replacement| |
图示:Possible time points forconsidering upgrade· Arrhythmia. Patient symptomsElective replacementChange in status ·Fluid accumulation / changeindicatedsymptoms/in thoracic Impedance.Increase In diureticPatient selection(ERI)NYHA classRe-evaluate· Heart failure events.Decline In activityAssess forArrhythmia risk ?LBBB ?RV pacing (>20%)?LV Ejection fraction≤35%0Echocardiography for LVEFAssess utility o(√)36-40%RV pacing avoidance algorithm(NT-Pro BNP)Can be used to rationalize imaging if required>10% decrease from baselineEvaluate forMulti-disciplinary team approach for decision-making, involving where neededAlternative causes for LVEFElectrophysiologist / Device specialist / Extraction specialist / HF specialist / Geriatrician / Heart surgeon / Anesthetistreduction (and relevant therapyDecisionNot appropriateUpgrade appropriatePossible strategies:If required• Biventricular pacing or conduction system pacingDevice upgrade. Resynchronization and/or defibrillator capabilitiesGeneratoCRT +/- Defibrillator· Lead abandonment vs. extraction of existing RV leadreplacement onlyif addition of defibrillator considered
图示:PreproceduralVenographyimagingRequired kit:(Low-thermal energy dissectionleads, sheaths, TLE and venoplastyUpgradetools etc.preproceduraInfectionplanningpreventionconsiderations- Personalization according to patientscharacteristics-Determine what procedural stepswould and would not be utilizedLead managementtricuspid regurgitation5.000assessmenMedication optimizationAvoid heparin bridging;Collaboration withmulti-disciplinary team
图示:~2/3 of stenoses occur in the peripheralcomponent of the venous tree(subclavian/distal innominate vein<1/10 of stenoses are central only(innominate/superior vena cava area~1/4 of patients with stenoses have bothperipheral and central lesionsFigure 5 Venous stenosis locations and their prevalence. The majority of venous stenoses occurs peripherally in the venous tree.114,12
图示:APreproceduraCollateral venous flowvenogramRetrograde fillingAntegrade filling
图示:Lesion/s crossedVascular access with micro punctureVenoplasty on stiffsystem and lesion crossed with KA2amplatz wirehydrophilic 5Fr catheter
图示:CVenoplasty back to pocketTight 'apple-core' lesionovercome with venoplasty aue
图示:DSuccessful CS cannulationand lead deployment
图示:D.99
图示:C
图示:D
图示:BACD
图示:0
图示:Upgrade OptionsInitial Device+TV pacing+LLP+TVICD+ NON-TVICD665cePcP iAV VPyAVe A c666Single Chambe+RA VICSP lRA& CSP & Dofb le+EV.orS-ICDPV(AP&AV eony+DotbleadWISELDCSPSldyLdBivenricula+EVorS.ICDDoihled(atSP & DofbCD(a VSP66/NV0D(ARA& CSP leads(AP& AALOyRA&LVIA* Not shown in illustrationDual Chamb● Epicardial LV lead placement is not shown(AV CSP leads(Avroy)in figure but could be utilized to add an LVleadSubcutoneou● Only Lead Abandonment scenarios are shownexamples do NOT show Extraction optionsLLERV+EV orS-ICDFigure 13Overview of common upgrade scenarios
图示:|Lead Type|Connector Design| |:---:|:---:| |IS-4 LLLL|No contact| |DF-4 LLHH|High, High|